Remote Monitoring of Visual Function in Patients with Maculopathy: The Aphelion Study

被引:2
|
作者
Joseph, Anthony [1 ]
Bullimore, Mark [2 ]
Drawnel, Faye [3 ]
Miranda, Marco [4 ,7 ]
Morgan, Zoe [3 ]
Wang, Yi-Zhong [5 ,6 ]
机构
[1] Retina Associates, Tucson, AZ USA
[2] Univ Houston, Coll Optometry, Houston, TX USA
[3] F Hoffmann La Roche Pharmaceut AG, Basel, Switzerland
[4] Roche Prod Ltd, Welwyn Garden City, England
[5] Retina Fdn Southwest, 9600 N Cent Expressway,Suite 200, Dallas, TX 75321 USA
[6] UT SouthWestern Med Ctr, Dept Ophthalmol, Dallas, TX USA
[7] UCL, Inst Ophthalmol, London, England
关键词
Contrast sensitivity; Hyperacuity; Maculopathy; Remote monitoring; Visual acuity; Smartphone application; COMPACT REDUCED LOGMAR; MACULAR DEGENERATION; SHAPE-DISCRIMINATION; CONTRAST SENSITIVITY; ACUITY CHART; HYPERACUITY; PROGRESSION; VALIDATION; AGREEMENT; VALIDITY;
D O I
10.1007/s40123-023-00854-2
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
IntroductionRemote monitoring of vision, using tools such as the shape discrimination hyperacuity (SDH) test, can detect disease activity in patients with maculopathy. We determined the in-clinic accuracy and repeatability of three myVisionTrack expanded version (mVTx) tests for self-testing of visual acuity (VA) and contrast sensitivity.MethodsAphelion, a single-arm, prospective study conducted at two sites in the USA, included adults with any maculopathy and a baseline VA of 0.7 log of minimum angle of resolution (logMAR) (Snellen 20/100) or better. Participants completed the mVTx tests (tumbling E, Landolt C, contrast sensitivity, and SDH) and standard clinical tests (near and distance Early Treatment Diabetic Retinopathy Study [ETDRS] charts and the Pelli-Robson contrast sensitivity chart). Test-retest repeatability and agreement between the mVTx tests and the corresponding clinical test were assessed by Bland-Altman analyses. Participants also completed a usability survey.ResultsThe mean age of the 122 participants was 67 years. The most common diagnosis was age-related macular degeneration (42% of patients). The tumbling E test had a test-retest 95% limit of agreement (LoA) of +/- 0.18 logMAR; the Landolt C test, +/- 0.23 logMAR; the SDH test, +/- 0.24 logMAR; and the contrast sensitivity test, +/- 0.32 log contrast threshold (logCT). Compared with the distance ETDRS chart, the LoA was +/- 0.35 logMAR for the tumbling E test (mean difference, - 0.07 logMAR) and +/- 0.39 logMAR for the Landolt C test (mean difference, 0.03 logMAR). For the contrast sensitivity test, the LoA compared with the Pelli-Robson chart was +/- 0.30 logCT (mean difference, - 0.25 logCT). Most participants (85%) reported that they learned the tests quickly. The tumbling E test scored the highest on ease of use.ConclusionThe mVTx tests of VA are accurate and repeatable, supporting their potential use alongside the SDH test to detect disease progression remotely between clinic visits.
引用
收藏
页码:409 / 422
页数:14
相关论文
共 50 条
  • [21] A randomized study of remote monitoring and fluid monitoring for the management of patients with implanted cardiac arrhythmia devices
    Luethje, Lars
    Vollmann, Dirk
    Seegers, Joachim
    Sohns, Christian
    Hasenfuss, Gerd
    Zabel, Markus
    EUROPACE, 2015, 17 (08): : 1276 - 1281
  • [22] Exploring the Feasibility and Usability of Smartphones for Monitoring Physical Activity in Orthopedic Patients: Prospective Observational Study
    Ghaffari, Arash
    Lauritsen, Rikke Emilie Kildahl
    Christensen, Michael
    Thomsen, Trine Rolighed
    Mahapatra, Harshit
    Heck, Robert
    Kold, Soren
    Rahbek, Ole
    JMIR MHEALTH AND UHEALTH, 2023, 11
  • [23] Effects of cataract type and location on visual function: The Blue Mountains Eye Study
    B E Chua
    P Mitchell
    R G Cumming
    Eye, 2004, 18 : 765 - 772
  • [24] Telemedicine: Remote monitoring of cardiac patients
    de Azevedo, DFG
    de Moura, EP
    DeCastro, MCF
    DeCastro, FCC
    da Rocha, MF
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 1390 - 1393
  • [25] The Diabetes Visual Function Supplement Study (DiVFuSS)
    Chous, A. Paul
    Richer, Stuart P.
    Gerson, Jeffry D.
    Kowluru, Renu A.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (02) : 227 - 234
  • [26] A Remote Monitoring System for Diabetes Patients
    Pustozerov E.A.
    Yuldashev Z.M.
    Biomedical Engineering, 2014, 48 (2) : 74 - 77
  • [27] Effects of cataract type and location on visual function: The Blue Mountains Eye Study
    Chua, BE
    Mitchell, P
    Cumming, RG
    EYE, 2004, 18 (08) : 765 - 772
  • [28] RESEARCH ON VISUAL NAVIGATION AND REMOTE MONITORING TECHNOLOGY OF AGRICULTURAL ROBOT
    Wang, Yuanjie
    Yang, Fuzeng
    Wang, Tao
    Liu, Qiang
    Xu, Xiudong
    INTERNATIONAL JOURNAL ON SMART SENSING AND INTELLIGENT SYSTEMS, 2013, 6 (02): : 466 - 481
  • [29] Effect of posterior capsular opacification on visual function in patients with monofocal and multifocal intraocular lenses
    M A Elgohary
    A B Beckingsale
    Eye, 2008, 22 : 613 - 619
  • [30] Effect of posterior capsular opacification on visual function in patients with monofocal and multifocal intraocular lenses
    Elgohary, M. A.
    Beckingsale, A. B.
    EYE, 2008, 22 (05) : 613 - 619