Optimized flip angle schemes for the split acquisition of fast spin-echo signals (SPLICE) sequence and application to diffusion-weighted imaging

被引:2
作者
Rahbek, Sofie [1 ]
Schakel, Tim [2 ]
Mahmood, Faisal [3 ,4 ]
Madsen, Kristoffer H. H. [5 ,6 ]
Philippens, Marielle E. P. [2 ]
Hanson, Lars G. G. [1 ,5 ,7 ]
机构
[1] Tech Univ Denmark, Dept Hlth Technol, Lyngby, Denmark
[2] Univ Med Ctr Utrecht, Dept Radiotherapy, Utrecht, Netherlands
[3] Odense Univ Hosp, Dept Oncol, Lab Radiat Phys, Odense, Denmark
[4] Univ Southern Denmark, Dept Clin Res, Odense, Denmark
[5] Copenhagen Univ Hosp, Danish Res Ctr Magnet Resonance, Ctr Funct & Diagnost Imaging & Res, Hvidovre, Denmark
[6] Tech Univ Denmark, Dept Appl Math & Comp Sci, Lyngby, Denmark
[7] Tech Univ Denmark, Bldg 349, DK-2800 Lyngby, Denmark
关键词
diffusion-weighted magnetic resonance imaging; point spread function; SNR; SPLICE; sequence optimization; variable flip angle; MRI; SENSITIVITY; MODULATION; HYPERECHO; TRAINS; BRAIN;
D O I
10.1002/mrm.29545
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe diffusion-weighted SPLICE (split acquisition of fast spin-echo signals) sequence employs split-echo rapid acquisition with relaxation enhancement (RARE) readout to provide images almost free of geometric distortions. However, due to the varying T2$$ {}_2 $$-weighting during k-space traversal, SPLICE suffers from blurring. This work extends a method for controlling the spatial point spread function (PSF) while optimizing the signal-to-noise ratio (SNR) achieved by adjusting the flip angles in the refocusing pulse train of SPLICE. MethodsAn algorithm based on extended phase graph (EPG) simulations optimizes the flip angles by maximizing SNR for a flexibly chosen predefined target PSF that describes the desired k-space density weighting and spatial resolution. An optimized flip angle scheme and a corresponding post-processing correction filter which together achieve the target PSF was tested by healthy subject brain imaging using a clinical 1.5 T scanner. ResultsBrain images showed a clear and consistent improvement over those obtained with a standard constant flip angle scheme. SNR was increased and apparent diffusion coefficient estimates were more accurate. For a modified Hann k-space weighting example, considerable benefits resulted from acquisition weighting by flip angle control. ConclusionThe presented flexible method for optimizing SPLICE flip angle schemes offers improved MR image quality of geometrically accurate diffusion-weighted images that makes the sequence a strong candidate for radiotherapy planning or stereotactic surgery.
引用
收藏
页码:1469 / 1480
页数:12
相关论文
共 49 条
  • [31] Accurate brain volumetry with diffusion-weighted spin-echo single-shot echo-planar-imaging and dual-clustering segmentation: Comparison with volumetry-validated quantitative magnetic resonance imaging
    Watanabe, Memi
    Sakai, Osamu
    Norbash, Alexander M.
    Jara, Hernan
    MEDICAL PHYSICS, 2010, 37 (03) : 1183 - 1190
  • [32] On the application of a non-CPMG single-shot fast spin-echo sequence to diffusion tensor MRI of the human brain
    Bastin, ME
    Le Roux, P
    MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) : 6 - 14
  • [33] Body diffusion-weighted imaging using magnetization prepared single-shot fast spin echo and extended parallel imaging signal averaging
    Gibbons, Eric K.
    Vasanawala, Shreyas S.
    Pauly, John M.
    Kerr, Adam B.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (06) : 3032 - 3044
  • [34] Optimising diffusion-weighted imaging in the abdomen and pelvis: comparison of image quality between monopolar and bipolar single-shot spin-echo echo-planar sequences
    Kyriazi, Stavroula
    Blackledge, Matthew
    Collins, David J.
    deSouza, Nandita M.
    EUROPEAN RADIOLOGY, 2010, 20 (10) : 2422 - 2431
  • [35] Locally advanced rectal cancer: 3D diffusion-prepared stimulated-echo turbo spin-echo versus 2D diffusion-weighted echo-planar imaging
    Zhang, Qinwei
    van Houdt, Petra J.
    Lambregts, Doenja M. J.
    van Triest, Baukelien
    Kop, Marnix P. M.
    Coolen, Bram F.
    Strijkers, Gustav J.
    van der Heide, Uulke A.
    Nederveen, Aart J.
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2020, 4 (01)
  • [36] Evaluation of a new Spin-echo diffusion-weighted sequence on a 0.35 T open magnetic resonance imaging (MRI)-system: first experiences within 3 h after acute stroke
    W. A. Wohlgemuth
    G. Schulte-Altedorneburg
    T. Becker
    L. Zha
    D. Kramer
    K. Kirchhof
    Neuroradiology, 2005, 47 : 532 - 538
  • [37] Evaluation of a new Spin-echo diffusion-weighted sequence on a 0.35 T open magnetic resonance imaging (MRI)-system: first experiences within 3 h after acute stroke
    Wohlgemuth, WA
    Schulte-Altedorneburg, G
    Becker, T
    Zha, L
    Kramer, D
    Kirchhof, K
    NEURORADIOLOGY, 2005, 47 (07) : 532 - 538
  • [38] Optimized T1-weighted contrast for single-slab 3D turbo spin-echo imaging with long echo trains:: Application to whole-brain imaging
    Park, Jaeseok
    Mugler, John P., III
    Horger, Wilhelm
    Kiefer, Berthold
    MAGNETIC RESONANCE IN MEDICINE, 2007, 58 (05) : 982 - 992
  • [39] 3D isotropic turbo spin-echo intermediate-weighted sequence with refocusing control in knee imaging: comparison study with 3D isotropic fast-field echo sequence
    Seo, Jung Min
    Yoon, Young Cheol
    Kwon, Jong Won
    ACTA RADIOLOGICA, 2011, 52 (10) : 1119 - 1124
  • [40] Value of three-dimensional T2-weighted turbo spin-echo imaging with tissue-specific variable refocusing flip angle for 3-T magnetic resonance imaging of prostate cancer: comparison with conventional two- and three-dimensional T2-weighted turbo spin-echo imaging
    Tanaka, Utaru
    Ueno, Yoshiko
    Morinaga, Yukiko
    Miyake, Hideaki
    Kyotani, Katsusuke
    Ueda, Yu
    Kitajima, Kazuhiro
    Sofue, Keitaro
    Suenaga, Yuko
    Sugimura, Kazuro
    Takahashi, Satoru
    JAPANESE JOURNAL OF RADIOLOGY, 2017, 35 (12) : 707 - 717