Survival Analysis and Failure Modes of Total Hip Arthroplasty Using a Cemented Semi-Retentive Acetabular Cup

被引:0
作者
Stan, Gabriel [1 ]
Roman, Mihai Dan [2 ]
Orban, Horia [1 ]
Georgeanu, Vlad Alexandru [1 ]
Deculescu, Rares Stefan [1 ]
Brinduse, Lacramioara Aurelia [1 ]
Gheorghiu, Nicolae [1 ]
机构
[1] Carol Davila Univ Med & Pharm, Fac Med, Bucharest 050474, Romania
[2] Univ Lucian Blaga Sibiu, Fac Med, Sibiu 550024, Romania
关键词
constrained; dislocation; failure; survival; total hip arthroplasty; FEMORAL-HEAD SIZE; RETROSPECTIVE ANALYSIS; DISLOCATION; REVISION; SURVIVORSHIP; COMPONENTS; LINER; RISK;
D O I
10.3390/jcm12247506
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To reduce the incidence of total hip revisions, there have been continuous efforts to enhance prosthetic materials and designs to optimize implant survival. A primary implant with a constrained acetabular component is often used to minimize the risk of dislocations even though this approach has some drawbacks as reported in the literature. To address these concerns, this study aimed to assess the survivorship and dislocation rate of a semi-retentive cemented acetabular cup when used as a primary implant. The specific cemented cup that we studied was not present in any study that we consulted, so to fill this gap, we conducted a retrospective examination of 527 cemented hip prostheses that utilized the semi-retentive cup between the years 2005 and 2012. We employed Cox multiple regression models for our statistical analysis. The revision due to dislocation occurred in 12.8% of all cases, with a lower incidence of 5% (14 cases) in age groups >70 years than in age groups <70 years (14%-32 cases) (p < 0.001). The survival rates of the semi-retentive cemented acetabular cup were 98.6% (520 cases) at 5 years and 92.2% (487 cases) at 10 years. The survival rates were significantly lower in women than men, with 1.9% (7 cases) toward 0% at 5 years and 8.1% (30 cases) toward 5% (7 cases) at 10 years (p = 0.002). The difference in failure rates between age groups over 70 years (2.3%-10 cases) and age groups under 70 years (11.5%-34 cases) was also statistically significant (p < 0.001). Our study indicates that the semi-constrained design may cause frequent damage to the polyethylene liner due to impingement and wear, which are the primary factors for failure. Also, this implant has a similar risk of revision due to dislocation as reported in studies and may be beneficial as a primary implant in elderly patients with low-demanding lifestyles, muscular insufficiency, and low compliance regarding hip prosthetic behavior, without a major effect on survivorship.
引用
收藏
页数:10
相关论文
共 50 条
[21]   Constrained Acetabular Components Used in Revision Total Hip Arthroplasty: A Registry Analysis [J].
Lewis, Peter L. ;
Graves, Stephen E. ;
de Steiger, Richard N. ;
Cuthbert, Alana R. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (10) :3102-3107
[22]   Acetabular Component Positioning in Total Hip Arthroplasty: An Evidence-Based Analysis [J].
Moskal, Joseph T. ;
Capps, Susan G. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (08) :1432-1437
[23]   Novel Measure of Acetabular Cup Inclination and Anteversion During Total Hip Arthroplasty [J].
Walter, William L. ;
Baker, Neri A. ;
Marsden-Jones, Daniel ;
Sadeghpour, Ameneh .
MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2022, 15 :1-14
[24]   Cemented Dual Mobility Cup for Primary Total Hip Arthroplasty in Elder Patients with High-Risk Instability [J].
Maria Lamo-Espinosa, Jose ;
Gomez-Alvarez, Jorge ;
Gatica, Javier ;
Suarez, Alvaro ;
Moreno, Victoria ;
Diaz de Rada, Pablo ;
Valenti-Azcarate, Andres ;
Alfonso-Olmos, Matias ;
San-Julian, Mikel ;
Ramon Valenti-Nin, Juan .
GERIATRICS, 2021, 6 (01)
[25]   Is transverse acetabular ligament an anatomical landmark to reliably orient the cup in primary total hip arthroplasty? [J].
Viste, A. ;
Chouteau, J. ;
Testa, R. ;
Cheze, L. ;
Fessy, M. -H. ;
Moyen, B. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (03) :241-245
[26]   A laser guide technique: a novel method for accurate acetabular cup alignment in total hip arthroplasty [J].
Kohno, Yusuke ;
Nakamura, Tetsuro ;
Fujii, Masanori ;
Shin, Satoshi ;
Hara, Toshihiko .
SCIENTIFIC REPORTS, 2022, 12 (01)
[27]   Accuracy of Acetabular Cup Placement During Total Hip Arthroplasty in Supine Position Using a Pelvic Rotation Correction Device [J].
Nakasone, Satoshi ;
Takaesu, Mika ;
Ishihara, Masato ;
Onaga, Masamichi ;
Igei, Takahiro ;
Miyata, Yoshihide ;
Nishida, Kotaro .
ARTHROPLASTY TODAY, 2022, 16 :46-52
[28]   Relationship between use of screws and acetabular cup stability in total hip arthroplasty: a meta-analysis [J].
Fei, Chen ;
Wang, Peng-Fei ;
Wei, Wei ;
Qu, Shuang-Wei ;
Yang, Kun ;
Li, Zhi ;
Zhuang, Yan ;
Zhang, Bin-Fei ;
Zhang, Kun .
JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2020, 48 (02)
[29]   Long-term outcome of cementless total hip arthroplasty with threaded Tropic® acetabular cup and Corail® femoral stem [J].
Almeida, Fernando ;
Gallego, Silvia ;
Arguelles, Francisco ;
Silvestre, Antonio .
ACTA ORTHOPAEDICA BELGICA, 2021, 87 (03) :393-399
[30]   Use of Computer Navigation for Optimal Acetabular Cup Placement in Revision Total Hip Arthroplasty: Case Reports and Surgical Techniques [J].
Neitzke, Anna Colin C. ;
Chandi, Sonia K. ;
Gausden, Elizabeth B. ;
Debbi, Eytan M. ;
Sculco, Peter K. ;
Chalmers, Brian P. .
ARTHROPLASTY TODAY, 2024, 27