Hemodynamic monitoring and correlation between electrical cardiometry and esophageal Doppler in patients undergoing major abdominal surgery

被引:0
作者
Abdalla, Shady Rady [1 ]
Abdelazeem, Ahmed Salah [1 ]
Kaddah, Tarek Abdelhalium [1 ]
Elhadedy, Abla Salah [1 ]
Khafagy, Hanan Farouk [1 ]
Mohamed, Ahmed Abdalla [1 ]
Essam, Ahmed Mohamed [1 ]
机构
[1] Theodor Bilharz Res Inst, Intens Care & Pain Management Dept, Anesthesia, Giza, Egypt
来源
EGYPTIAN JOURNAL OF ANAESTHESIA | 2023年 / 39卷 / 01期
关键词
Hemodynamic monitoring; electrical cardiometry; esophageal Doppler; abdominal surgery; TRANSESOPHAGEAL DOPPLER; CARDIAC-OUTPUT; FLUID MANAGEMENT; ACCURACY; CHILDREN; TIME;
D O I
10.1080/11101849.2023.2209415
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Transesophageal Doppler (TED) is a minimally invasive monitor that allows continuous hemodynamic variables monitoring. Non-invasive electrical cardiometry (EC) and TED methods added an additional facility to monitor cardiac output (CO) continuously and guide fluid management. The aim of this study was to correlate hemodynamic monitoring between non-invasive EC and Esophageal Doppler (ED) in cases undergoing major abdominal surgery. Methods: This prospective observational study was carried out on 35 adult cases, American Society of Anesthesiologists (ASA) physical status II or III, undergoing major abdominal surgery. Esophageal Doppler and EC were attached to the same patient. Parameters measured were hemodynamic parameters. Results: Comparison of CO with ICON and ED showed that the ICON mean value ranged from 5.6 to 6.2 l/min, and the ED mean value always ranged from 5.7 to 7.6 l/min with the non-significant difference between the two methods. The precision for the ICON ranged from 15.19 to 17.99% and the precision for ED ranged from 13.39 to 17.08%. At a 15% change in ICON, the ED values' sensitivity was 72.6% and specificity was 30.9% with AUC 0.505. Conclusion: The agreement between CO measured by EC and ED is acceptable. Both monitored trend changes and guided fluid administration in the operation theater. The EC is as accurate as ED in measuring hemodynamics during major abdominal surgery.
引用
收藏
页码:418 / 425
页数:8
相关论文
共 50 条
  • [1] Comparison of stroke volume measurement between non-invasive bioreactance and esophageal Doppler in patients undergoing major abdominal–pelvic surgery
    Gennaro De Pascale
    Mervyn Singer
    David Brealey
    Journal of Anesthesia, 2017, 31 : 545 - 551
  • [2] Comparison of stroke volume measurement between non-invasive bioreactance and esophageal Doppler in patients undergoing major abdominal-pelvic surgery
    De Pascale, Gennaro
    Singer, Mervyn
    Brealey, David
    JOURNAL OF ANESTHESIA, 2017, 31 (04) : 545 - 551
  • [3] Intraoperative Esophageal Doppler Hemodynamic Monitoring in Free Perforator Flap Surgery
    Figus, Andrea
    Wade, Ryckie G.
    Oakey, Stephen
    Ramakrishnan, Venkat V.
    ANNALS OF PLASTIC SURGERY, 2013, 70 (03) : 301 - 307
  • [4] Electrical cardiometry for non-invasive cardiac output monitoring: a method comparison study in patients after coronary artery bypass graft surgery
    Greiwe, Gillis
    Saad, Rami
    Hapfelmeier, Alexander
    Neumann, Niklas
    Tariparast, Pischtaz
    Saugel, Bernd
    Flick, Moritz
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2024, : 371 - 376
  • [5] Intraoperative management and hemodynamic monitoring for major abdominal surgery : a narrative review
    Russo, A.
    Roman, B.
    ACTA ANAESTHESIOLOGICA BELGICA, 2021, 72 (02) : 63 - 71
  • [6] Effects of a restrictive fluid regimen in pediatric patients undergoing major abdominal surgery
    Mandee, Sahatsa
    Butmangkun, Wassana
    Aroonpruksakul, Naiyana
    Tantemsapya, Niramol
    von Bormann, Benno
    Suraseranivongse, Suwannee
    PEDIATRIC ANESTHESIA, 2015, 25 (05) : 530 - 537
  • [7] Esophageal Doppler (ODM II) improves intraoperative hemodynamic monitoring during laparoscopic surgery
    Koliopanos, A
    Zografos, G
    Skiathitis, S
    Stithos, D
    Voukena, V
    Karampinis, A
    Papastratis, G
    SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2005, 15 (06) : 332 - 338
  • [8] The Incidence of Perioperative Hypotension in Patients Undergoing Major Abdominal Surgery with the Use of Arterial Waveform Analysis and the Hypotension Prediction Index Hemodynamic Monitoring-A Retrospective Analysis
    Szrama, Jakub
    Gradys, Agata
    Bartkowiak, Tomasz
    Wozniak, Amadeusz
    Nowak, Zuzanna
    Zwolinski, Krzysztof
    Lohani, Ashish
    Jawien, Natalia
    Smuszkiewicz, Piotr
    Kusza, Krzysztof
    JOURNAL OF PERSONALIZED MEDICINE, 2024, 14 (02):
  • [9] Electrical cardiometry compared to transesophageal doppler for hemodynamics monitoring and fluid management in pediatrics undergoing Kasai operation. A randomized controlled trial
    Lotfy, M.
    Yassen, K.
    El Sharkawy, O.
    Elshoney, R.
    Moustafa, A.
    PEDIATRIC ANESTHESIA AND CRITICAL CARE JOURNAL, 2018, 6 (01): : 46 - 54
  • [10] Hemodynamic monitoring in patients with severe cardiovascular disease undergoing cardiac surgery
    Ivancan, V.
    NEUROLOGIA CROATICA, 2004, 53 : 57 - 66