Improved corrosion protection of a NiTi implant by an electrodeposited HAp-Nb2O5 composite layer

被引:22
作者
Safavi, Mir Saman [1 ,2 ]
Khalil-Allafi, Jafar [1 ]
Ahadzadeh, Iraj [3 ]
Walsh, Frank C. [4 ,5 ]
Visai, Livia [2 ,6 ]
机构
[1] Sahand Univ Technol, Fac Mat Engn, Res Ctr Adv Mat, POB 51335-1996, Tabriz, Iran
[2] Univ Pavia, Ctr Hlth Technol CHT, Mol Med Dept DMM, UdR INSTM, Via Taramelli 3-B, I-27100 Pavia, Italy
[3] Univ Tabriz, Fac Chem, Dept Phys Chem, Res Lab Electrochem Instrumentat & Energy Syst, Tabriz, Iran
[4] Univ Southampton, Fac Engn & Environm, Electrochem Engn Lab, Southampton SO17 1BJ, England
[5] Univ Southampton, Fac Engn & Environm, Natl Ctr Adv Tribol, Southampton SO17 1BJ, England
[6] IRCCS, UOR5 Lab Nanotecnol, Med Clin Specialist, ICS Maugeri, I-27100 Pavia, Italy
基金
美国国家科学基金会;
关键词
Biomaterials; Composite coating; Corrosion; NiTi; IN-VITRO BIOACTIVITY; BIOMEDICAL APPLICATIONS; NANO-HYDROXYAPATITE; COATINGS; DEPOSITION; ALLOYS; RESISTANCE; PHOSPHATE; BEHAVIOR; SURFACE;
D O I
10.1016/j.surfcoat.2023.129822
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The mechanical and biological characteristics of NiTi alloy provide opportunities in biomedical fields ranging from orthopedics to intravascular stents. However, Ni ion release when the alloy is exposed to corrosive body fluid can lead to long-term health-related concerns. Improved surface engineering via the application of biological layers on the NiTi can contribute to higher corrosion resistance. HAp-based layers were cathodically deposited from an aqueous electrolyte containing calcium nitrate, ammonium dihydrogen phosphate, and concentrations of suspended Nb2O5 particles from 0 to 1 g/L under pulsed current conditions. Surface morphology, topography, and surface chemistry of the electroplated layers were characterized by FESEM, AFM, and FTIR analysis, respectively. The corrosion behavior of the bare and coated samples in Ringer's solution was evaluated by EIS and potentiodynamic polarization; the corroded surfaces were imaged by FESEM. A more compact and smoother surface was formed in the presence of particulate Nb2O5 inclusions. The composite coating showed excellent anti-corrosion properties, increasing the polarization resistance of pure HAp films by >10 times. FESEM imaging of corroded surfaces showed the partial detachment of plate-like HAp crystals from the surface of the coated NiTi in contrast to the pitting of the bare NiTi surface. The surface modification of NiTi by electrodeposited HAp-Nb2O5 composite coatings might play an important role in the future use of NiTi in clinical applications.
引用
收藏
页数:14
相关论文
共 57 条
  • [1] Hydroxyapatite based and anodic Titania nanotube biocomposite coatings: Fabrication, characterization and electrochemical behavior
    Ahmadi, Shahab
    Mohammadi, Iman
    Sadrnezhaad, S. K.
    [J]. SURFACE & COATINGS TECHNOLOGY, 2016, 287 : 67 - 75
  • [2] Assessment of Ni ion release from TiTaHfNbZr high entropy alloy coated NiTi shape memory substrates in artificial saliva and gastric fluid
    Aksoy, C. B.
    Canadinc, D.
    Yagci, M. B.
    [J]. MATERIALS CHEMISTRY AND PHYSICS, 2019, 236
  • [3] A facile and cost-effective practical approach to develop clinical applications of NiTi: Fenton oxidation process
    Asghari, R.
    Safavi, M. S.
    Khalil-Allafi, J.
    [J]. TRANSACTIONS OF THE INSTITUTE OF METAL FINISHING, 2020, 98 (05): : 250 - 257
  • [4] In-vitro degradation behavior of Mg alloy coated by fluorine doped hydroxyapatite and calcium deficient hydroxyapatite
    Bakhsheshi-Rad, H. R.
    Hamzah, E.
    Daroonparvar, M.
    Yajid, M. A. M.
    Kasiri-Asgarani, M.
    Abdul-Kadir, M. R.
    Medraj, M.
    [J]. TRANSACTIONS OF NONFERROUS METALS SOCIETY OF CHINA, 2014, 24 (08) : 2516 - 2528
  • [5] THE ANALYSIS OF ELECTRODE IMPEDANCES COMPLICATED BY THE PRESENCE OF A CONSTANT PHASE ELEMENT
    BRUG, GJ
    VANDENEEDEN, ALG
    SLUYTERSREHBACH, M
    SLUYTERS, JH
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1984, 176 (1-2): : 275 - 295
  • [6] MWCNT reinforced bone like calcium phosphate-Hydroxyapatite composite coating developed through pulsed electrodeposition with varying amount of apatite phase and crystallinity to promote superior osteoconduction, cytocompatibility and corrosion protection performance compared to bare metallic implant surface
    Chakraborty, Rajib
    Seesala, Venkata S.
    Sen, Mainak
    Sengupta, Srijan
    Dhara, Santanu
    Saha, Partha
    Das, Karabi
    Das, Siddhartha
    [J]. SURFACE & COATINGS TECHNOLOGY, 2017, 325 : 496 - 514
  • [7] Synthesis of calcium hydrogen phosphate and hydroxyapatite coating on SS316 substrate through pulsed electrodeposition
    Chakraborty, Rajib
    Sengupta, Srijan
    Saha, Partha
    Das, Karabi
    Das, Siddhartha
    [J]. MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2016, 69 : 875 - 883
  • [8] Surface engineering and modification of biomaterials
    Chu, Paul K.
    [J]. THIN SOLID FILMS, 2013, 528 : 93 - 105
  • [9] Influence of deposition temperature on the properties of hydroxyapatite obtained by electrochemical assisted deposition
    Cotrut, Cosmin M.
    Vladescu, Alina
    Dinu, Mihaela
    Vranceanu, Diana M.
    [J]. CERAMICS INTERNATIONAL, 2018, 44 (01) : 669 - 677
  • [10] Lopes NID, 2017, MATER RES-IBERO-AM J, V20, P572, DOI [10.1590/1980-5373-MR-2016-0933, 10.1590/1980-5373-mr-2016-0933]