Calibration Practices in Clinical Mass Spectrometry: Review and Recommendations

被引:18
作者
Cheng, Wan Ling [1 ]
Markus, Corey [2 ]
Lim, Chun Yee [3 ]
Tan, Rui Zhen [3 ]
Sethi, Sunil Kumar [1 ]
Loh, Tze Ping [1 ]
机构
[1] Natl Univ Singapore Hosp, Dept Lab Med, 5 Lower Kent,Ridge Rd, Singapore 119074, Singapore
[2] Flinders Univ S Australia, Flinders Univ Int Ctr Point Care Testing, Flinders Hlth & Med Res Inst, Adelaide, SA, Australia
[3] Singapore Inst Technol, Engn Cluster, Singapore, Singapore
关键词
Calibration; Mass spectrometry; Regression; Linearity; Statistics; LABELED INTERNAL STANDARD; SQUARES LINEAR-REGRESSION; PART II; VALIDATION; QUANTIFICATION; CURVES; ASSAYS; WATER; IONIZATION; STRATEGIES;
D O I
10.3343/alm.2023.43.1.5
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Calibration is a critical component for the reliability, accuracy, and precision of mass spectrometry measurements. Optimal practice in the construction, evaluation, and implementation of a new calibration curve is often underappreciated. This systematic review examined how calibration practices are applied to liquid chromatography-tandem mass spectrometry measurement procedures. Methods: The electronic database PubMed was searched from the date of database inception to April 1, 2022. The search terms used were "calibration," "mass spectrometry," and "regression." Twenty-one articles were identified and included in this review, following evaluation of the titles, abstracts, full text, and reference lists of the search results. Results: The use of matrix-matched calibrators and stable isotope-labeled internal standards helps to mitigate the impact of matrix effects. A higher number of calibration standards or replicate measurements improves the mapping of the detector response and hence the accuracy and precision of the regression model. Constructing a calibration curve with each analytical batch recharacterizes the instrument detector but does not reduce the actual variability. The analytical response and measurand concentrations should be considered when constructing a calibration curve, along with subsequent use of quality controls to confirm assay performance. It is important to assess the linearity of the calibration curve by using actual experimental data and appropriate statistics. The heteroscedasticity of the calibration data should be investigated, and appropriate weighting should be applied during regression modeling. Conclusions: This review provides an outline and guidance for optimal calibration practices in clinical mass spectrometry laboratories.
引用
收藏
页码:5 / 18
页数:14
相关论文
共 44 条
[1]   Effective validation of chromatographic analytical methods: The illustrative case of androgenic steroids [J].
Alladio, E. ;
Amante, E. ;
Bozzolino, C. ;
Seganti, F. ;
Salomone, A. ;
Vincenti, M. ;
Desharnais, B. .
TALANTA, 2020, 215
[2]  
Bonfiglio R, 1999, RAPID COMMUN MASS SP, V13, P1175, DOI 10.1002/(SICI)1097-0231(19990630)13:12<1175::AID-RCM639>3.3.CO
[3]  
2-S
[4]  
Boyd RK, 2011, TRACE QUANTITATIVE A, P373
[5]   A 24-Hour Extended Calibration Strategy for Quantitating Tacrolimus Concentrations by Liquid Chromatography-Tandem Mass Spectrometry [J].
Brister-Smith, Allister ;
Young, Jeffrey A. ;
Saitman, Alec .
JOURNAL OF APPLIED LABORATORY MEDICINE, 2021, 6 (05) :1293-1298
[6]  
CLSI, 2014, C62A CLSI
[7]  
CLSI, 2018, CLSI
[8]  
CLSI, 2020, EP06ED2 CLSI, V2nd
[9]   Compensate for or Minimize Matrix Effects? Strategies for Overcoming Matrix Effects in Liquid Chromatography-Mass Spectrometry Technique: A Tutorial Review [J].
Cortese, Manuela ;
Gigliobianco, Maria Rosa ;
Magnoni, Federico ;
Censi, Roberta ;
Di Martino, Piera .
MOLECULES, 2020, 25 (13)
[10]   A novel approach to quantitative LC-MS/MS: therapeutic drug monitoring of clozapine and norclozapine using isotopic internal calibration [J].
Couchman, Lewis ;
Belsey, Sarah L. ;
Handley, Simon A. ;
Flanagan, Robert J. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (29) :9455-9466