Comparison of central corneal thickness in healthy eyes using ultrasound pachymetry, non-contact specular microscopy and a high-resolution Scheimpflug camera

被引:2
作者
Soulantzou, Konstantina [1 ]
Plakitsi, Athina [1 ]
Chalkiadaki, Evangelia [2 ]
Anastasopoulou, Maria [3 ]
Karmiris, Efthymios [2 ]
机构
[1] Univ West Attica, Div Opt & Optometry, Dept Biomed Sci, Athens, Greece
[2] 251 Hellen Airforce Gen Hosp, Dept Ophthalmol, Athens, Greece
[3] Grad Econ Univ Piraeus, Dept Stat & Insurance Sci, Athens, Greece
关键词
Central corneal thickness; Ultrasonic pachymetry; Non-contact specular microscopy; Scheimpflug camera; Pentacam HR; OPTICAL-COHERENCE-TOMOGRAPHY; ECTASIA; REPEATABILITY; AGREEMENT; PENTACAM; SYSTEM; LASIK;
D O I
10.1007/s10792-022-02431-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare and analyze the interchangeability in measuring central corneal thickness (CCT) using ultrasonic pachymetry (USP, PACHMATE 2, DGH, Inc, Exton, PA, USA), non-contact specular microscopy (NCSM, CEM-530, Nidek CO, LTD, Gamagori, Japan) and a high-resolution Scheimpflug Camera (Pentacam HR, OCULUS, Wetzlar, Germany). Methods An observational, cross-sectional study was performed recruiting 216 volunteers, for a sample size of 216 eyes with no ocular abnormalities other than refractive errors. All subjects underwent pachymetric measurements obtained by USP, NCSM and Pentacam HR. Examinations were performed by the same examiner with USP always following the noncontact examinations. Results The mean CCT (+/- SD) was 560.30 +/- 38.80 mu m, 556.76 +/- 36.83 mu m and 547.31 +/- 35.28 mu m for USP, NCSM and Pentacam HR, respectively. The Bland-Altman analysis showed that the highest concordance was found between USP and NCSM, with differences between - 13.18 mu m and 20.26 mu m. For the differences between measurements obtained with Pentacam HR and USP, the differences at Bland-Altman plot were between - 28.25 and 13.57 mu m. The lowest concordance was found for the CCT values measured with Pentacam HR and NCSM, with differences between - 25.67 and 6.86 mu m. The intraclass correlation coefficient (ICC) between all pairs of measurements was between 0.979 and 0.987, suggesting that the mean measurements were strongly correlated. Conclusion Measurements obtained with all three devices had high correlation. USP and NCSM were found in good agreement and high concordance, too. The above results indicate that these two devices are interchangeable in clinical practice. Pentacam HR may be a useful alternative for measuring CCT; however, it significantly underestimates CCT and cannot be used interchangeably with the other devices that we studied.
引用
收藏
页码:363 / 370
页数:8
相关论文
共 41 条
  • [1] Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes
    Al-Ageel, Saleh
    Al-Muammar, Abdulrahman M.
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2009, 23 (3-4) : 181 - 187
  • [2] Ambrosio Jr R, OCULUS PENTACAM HR I
  • [3] [Anonymous], 2002, DGH 55 PACHAMATE HAN
  • [4] [Anonymous], 2012, CEM530 JAP NIDEK CO
  • [5] Corneal thickness values before and after oxybuprocaine 0.4% eye drops
    Asensio, I
    Rahhal, SM
    Alonso, L
    Palanca-Sanfrancisco, JM
    Sanchis-Gimeno, JA
    [J]. CORNEA, 2003, 22 (06) : 527 - 532
  • [6] Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry
    Barkana, Y
    Gerber, Y
    Elbaz, U
    Schwartz, S
    Ken-Dror, G
    Avni, I
    Zadok, D
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) : 1729 - 1735
  • [7] Analysis of ectasia after laser in situ keratomileusis: Risk factors
    Binder, Perry S.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (09) : 1530 - 1538
  • [8] Evaluation of a Risk Factor Scoring System for Corneal Ectasia After LASIK in Eyes With Normal Topography
    Binder, Perry S.
    Trattler, William B.
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2010, 26 (04) : 241 - 250
  • [9] The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology
    Browning, AC
    Bhan, A
    Rotchford, AP
    Shah, S
    Dua, HS
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (11) : 1395 - 1399
  • [10] Accuracy of Cirrus HD-OCT and Topcon SP-3000P for measuring central corneal thickness
    Calvo-Sanz, Jorge A.
    Ruiz-Alcocer, Javier
    Sanchez-Tena, Miguel A.
    [J]. JOURNAL OF OPTOMETRY, 2018, 11 (03) : 192 - 197