Study and experimental investigation of insertion force modeling and tissue deformation phenomenon during surgical needle-soft tissue interaction

被引:1
|
作者
Barua, Ranjit [1 ,6 ]
Das, Surajit [2 ]
Datta, Sudipto [3 ]
Datta, Pallab [4 ]
Chowdhury, Amit Roy [5 ]
机构
[1] IIEST Shibpur, Ctr Healthcare Sci & Technol, Howrah, West Bengal, India
[2] RG Kar Med Coll & Hosp, Kolkata, West Bengal, India
[3] Indian Inst Technol, Roorkee, Uttaranchal, India
[4] Natl Inst Pharmaceut Educ & Res, Kolkata, West Bengal, India
[5] IIEST Shibpur, Aerosp Engn & Appl Mech, Howrah, West Bengal, India
[6] IIEST Shibpur, Ctr Healthcare Sci & Technol, Near be Garden, Howrah 711103, West Bengal, India
关键词
Needle insertion; bevel tip; insertion force; tissue deformation; MIS; frictional stress; DEFLECTION; SIMULATION;
D O I
10.1177/09544062221126628
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Surgical needle insertion is generally used in the current advanced surgery process, particularly in the area of biopsy, MIS or minimally invasive surgery, brachytherapy, etc. During the needle insertion performance into the soft tissue, one of the most important issues is the tissue deformation that affects the needle engagement inside the tissue material. We have presented an energy-based insertion model, and the conical shape of deformed tissue is assumed. The tests are performed on PVA gel samples, and the model is analyzed to determine the tissue deformation volume during needle insertion into soft tissue. The procedures for evaluating the puncture force which creates the deformation of the contact point. With the help of the needle insertion experimental investigation, a geometric model of tissue deformation phenomenon and insertion force was investigated. The active needle-soft tissue contacts are studied, and with the help of the energy-based insertion model, the different factors like strain energy, potential, and dissipated energies are investigated. The tests are completed on the tissue mimic PVA gel samples, and the results show the volume and area of tissue deformation; at the time of initial needle insertion, the maximum tissue deformation arises and during the needle movement in the post-perforation stage, the deformation gradually decreases as the more peripheral work is altered into the vicious and degenerate energies. The maximum insertion force was noticed in 14 degrees bevel angle needle tip and at 3 mm/s needle insertion speed, the volume of tissue deformation was maximum. The maximum deformation of the issue arises at the initial perforate position, and the distortion reduces along with the raises of the needle movement during the post-insertion period as the needle insertion work is converted into vicious and dissolute energies.
引用
收藏
页码:1007 / 1014
页数:8
相关论文
共 26 条
  • [1] Study of the surgical needle and biological soft tissue interaction phenomenon during insertion process for medical application: A Survey
    Barua, Ranjit
    Datta, Sudipto
    RoyChowdhury, Amit
    Datta, Pallab
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (10) : 1465 - 1477
  • [2] Experimental study on needle insertion force to minimize tissue deformation in tongue tissue
    Aaboubout, Y.
    Soares, M. R. Nunes
    Barroso, E. M.
    van der Sar, L. C.
    Bocharnikov, A.
    Usenov, I
    Artyushenko, V
    Caspers, P. J.
    Koljenovic, S.
    Schut, T. C. Bakker
    van den Dobbelsteen, J. J.
    Puppels, G. J.
    MEDICAL ENGINEERING & PHYSICS, 2021, 97 (97) : 40 - 46
  • [3] Experimental study of mosquito-inspired needle to minimize insertion force and tissue deformation
    Gidde, Sai Teja Reddy
    Islam, Sayemul
    Kim, Albert
    Hutapea, Parsaoran
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2023, 237 (01) : 113 - 123
  • [4] Control of soft tissue deformation during robotic needle insertion
    Abolhassani, Niki
    Patel, Rajni
    Moallem, Mehrdad
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (03) : 165 - 176
  • [5] Soft Tissue Deformation Modeling in the Procedure of Needle Insertion: A Kriging-Based Method
    Lei, Yong
    Li, Murong
    Gao, Dedong
    CHINESE JOURNAL OF MECHANICAL ENGINEERING, 2022, 35 (01)
  • [6] Experimental study of robotic needle insertion in soft tissue
    Abolhassani, N
    Patel, R
    Moallem, M
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 797 - 802
  • [7] Simulation for Needle Deflection and Soft Tissue Deformation in Needle Insertion
    Gao, Dedong
    Zheng, Haojun
    MANUFACTURING ENGINEERING AND AUTOMATION I, PTS 1-3, 2011, 139-141 : 889 - +
  • [8] Experimental study of needle-tissue interaction forces: Effect of needle geometries, insertion methods and tissue characteristics
    Jiang, Shan
    Li, Pan
    Yu, Yan
    Liu, Jun
    Yang, Zhiyong
    JOURNAL OF BIOMECHANICS, 2014, 47 (13) : 3344 - 3353
  • [9] INTERACTIVE SIMULATION OF SURGICAL NEEDLE INSERTION INTO SOFT TISSUE
    Jin, Xin
    Gao, Dedong
    Lei, Yong
    PROCEEDINGS OF THE ASME 5TH ANNUAL DYNAMIC SYSTEMS AND CONTROL DIVISION CONFERENCE AND JSME 11TH MOTION AND VIBRATION CONFERENCE, DSCC 2012, VOL 3, 2013, : 457 - 464
  • [10] Deflection modeling for a needle actuated by lateral force and axial rotation during insertion in soft phantom tissue
    Lehmann, Thomas
    Rossa, Carlos
    Usmani, Nawaid
    Sloboda, Ronald
    Tavakoli, Mandi
    MECHATRONICS, 2017, 48 : 42 - 53