Glucose biosensors based on Michael addition crosslinked poly(ethylene glycol) hydrogels with chemo-optical sensing microdomains

被引:16
作者
Williams, Tyrell J. J. [1 ]
Jeevarathinam, Ananthakrishnan S. S. [1 ]
Jivan, Faraz [1 ]
Baldock, Victoria [1 ]
Kim, Paul [1 ]
McShane, Michael J. J. [1 ,2 ]
Alge, Daniel L. L. [1 ,2 ]
机构
[1] Texas A&M Univ, Dept Biomed Engn, College Stn, TX 77843 USA
[2] Texas A&M Univ, Dept Mat Sci & Engn, College Stn, TX 77843 USA
基金
美国国家卫生研究院;
关键词
RAMAN-SPECTROSCOPY; NANOFILM COATINGS; IN-VIVO; OXIDASE; SENSORS; DESIGN; ENZYME;
D O I
10.1039/d2tb02339c
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
Continuous glucose monitoring (CGM) devices have the potential to lead to better disease management and improved outcomes in patients with diabetes. Chemo-optical glucose sensors offer a promising, accurate, long-term alternative to the current CGMs that require frequent calibration and replacement. Recently, we have proposed glucose sensor designs using phosphorescence lifetime-based measurement of chemo-optical glucose sensing microdomains embedded within alginate hydrogels. Due to the poor long-term stability of calcium-crosslinked alginate, we propose poly(ethylene glycol) (PEG) hydrogels synthesized via thiol-Michael addition chemistry as an alternative hydrogel carrier. The objective of this study was to evaluate the suitability of Michael addition crosslinked PEG hydrogels compared to calcium crosslinked alginate hydrogels for encapsulating glucose-sensing microdomains. PEG hydrogels crosslinked via thiol-vinyl sulfone addition achieved gelation in under 5 minutes, resulting in an even distribution of sensing microdomains. The shear storage modulus of the PEG hydrogels was tunable from 2.2 +/- 0.1 kPa to 9.5 +/- 1.8 kPa, which was comparable to the alginate hydrogels (10.5 +/- 0.8 kPa), and the inclusion of microdomains did not significantly impact stiffness. The high water content of PEG hydrogels resulted in high glucose permeability that closely corresponded to the glucose permeability of alginate (D = 0.09 and 0.12 cm(2) s(-1), respectively; p = 0.47), but the PEG hydrogels exhibited superior stability. Both PEG and alginate-embedded sensors exhibited a sensing range up to similar to 200 mg dL(-1) glucose. The lower limits of detection (LOD) for PEG and alginate-based glucose sensors were 19.8 and 20.6 mg dL(-1) with a difference of just 4.2% variation. The small difference between PEG and alginate embedded sensors indicates that their sensing properties are primarily determined by the glucose sensing microdomains rather than the hydrogel matrix. Overall, the results of this study indicate that Michael addition-crosslinked PEG hydrogels are a promising platform for encapsulation of chemo-optical glucose sensing microdomains.
引用
收藏
页码:1749 / 1759
页数:11
相关论文
共 54 条
  • [1] Ahmed I., 2022, SENSORS DIAGNOSTICS, V1, P1098
  • [2] Alcantar NA, 2000, J BIOMED MATER RES, V51, P343, DOI 10.1002/1097-4636(20000905)51:3<343::AID-JBM7>3.0.CO
  • [3] 2-D
  • [4] Major Complications and Adverse Events Related to the Injection of the SpaceOAR Hydrogel System Before Radiotherapy for Prostate Cancer: Review of the Manufacturer and User Facility Device Experience Database
    Aminsharifi, Alireza
    Kotamarti, Srinath
    Silver, David
    Schulman, Ariel
    [J]. JOURNAL OF ENDOUROLOGY, 2019, 33 (10) : 868 - 871
  • [5] Characterization of Lactate Sensors Based on Lactate Oxidase and Palladium Benzoporphyrin Immobilized in Hydrogels
    Andrus, Liam P.
    Unruh, Rachel
    Wisniewski, Natalie A.
    McShane, Michael J.
    [J]. BIOSENSORS-BASEL, 2015, 5 (03): : 398 - 416
  • [6] Cutaneous Complications With Continuous or Flash Glucose Monitoring Use: Systematic Review of Trials and Observational Studies
    Asarani, Nurul A. Mohd
    Reynolds, Andrew N.
    Boucher, Sara E.
    de Bock, Martin
    Wheeler, Benjamin J.
    [J]. JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2020, 14 (02): : 328 - 337
  • [7] Composite Hydrogels Containing Bioactive Microreactors for Optical Enzymatic Lactate Sensing
    Biswas, Aniket
    Bornhoeft, Lindsey R.
    Banerjee, Swayoma
    You, Yil-Hwan
    McShane, Michael J.
    [J]. ACS SENSORS, 2017, 2 (11): : 1584 - 1588
  • [8] Cross-linked nanofilms for tunable permeability control in a composite microdomain system
    Biswas, Aniket
    Nagaraja, Ashvin T.
    You, Yil-Hwan
    Roberts, Jason R.
    McShane, Michael J.
    [J]. RSC ADVANCES, 2016, 6 (75): : 71781 - 71790
  • [9] Composite Hydrogels with Engineered Microdomains for Optical Glucose Sensing at Low Oxygen Conditions
    Bornhoeft, Lindsey R.
    Biswas, Aniket
    McShane, Michael J.
    [J]. BIOSENSORS-BASEL, 2017, 7 (01):
  • [10] Enzymatic fluorescent microsphere glucose sensors: Evaluation of response under dynamic conditions
    Brown, J. Quincy
    Srivastava, Rohit
    Zhu, Huiguang
    McShane, Michael J.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2006, 8 (03) : 288 - 295