Evaluation of the Malalignment Varus- Valgus in Total Knee Arthroplasty Designed for Deep Knee Flexion Used Knee Kinematic Motion Simulator

被引:1
作者
Triwardono, Joko [1 ]
Rokhmanto, Fendy [1 ]
Roberto, Rahadian [1 ]
Kartika, Ika [1 ]
Saragih, Agung Shamsuddin [2 ]
Whulanza, Yudan [2 ]
Shuib, Solehuddin [3 ]
Supriadi, Sugeng [2 ]
机构
[1] Natl Res & Innovat Agcy, Res Ctr Met, KST BJ Habibie, South Tangerang, Indonesia
[2] Univ Indonesia, Dept Mech Engn, Depok, Indonesia
[3] Univ Teknol MARA, Coll Engn, Sch Mech Engn, Shah Alam, Selangor, Malaysia
关键词
malalignment; hyperflexion; subluxations; dislocations; wear; polyethylene; PATIENT SATISFACTION; FAILURE; THICKNESS; COMPONENT; ALIGNMENT; FRACTURE; OUTCOMES; JOINT; POST;
D O I
10.4028/p-l6p55i
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Revision total knee arthroplasties cause performed aseptic loosening, instability, and polyethylene wear. Separation or removal of the femoral component has been observed and this has the potential to severely damage the polyethylene component. In most cases 90% of the patients examined experienced significant medial or lateral condylar lift at some stage during the gait cycle. Using the MRI, a normal knee has maximum lateral lift is approx. 6.7 mm and maximum medial lift is approx. 2.1 mm, when a varus strees applied at a 90 degrees knee flexion. Elevation of the lateral condyle due to valgus malalignment will distribute more contact force on the medial condyle. In this study, a polyethylene component of a posterior-stabilized right knee joint implant was developed to facilitate a high range of motion (ROM). Malalignment valgus was observed with the axes of knee motion joint implants were varied from 0 degrees, 2 degrees, 3 degrees to 5 and knee bend measurements at 30 degrees, 60 degrees, 90 degrees, 120 degrees, and 150 degrees of knee flexion. Using the knee kinematic motion simulator, the modified polyethylene component resulted in 0 degrees malalignment there is no gap of the femoral component with the polyethylene component, from 30 degrees to 150 degrees of knee flexion. At 2 degrees malalignment, the femoral component was raised by 0.5 mm at a 90 degrees to 150 degrees knee flexion and increased with increasing knee flexion. Maximum gap occurs at 5 degrees malalignment in the amount of 5 mm at 150 degrees of knee flexion. The aim of this study was therefore to evaluation malalignment valgus of the flexed knee using knee kinematic motion simulator, with reference to the tibiofemoral flexion gap. The result that the modified design is expected in an narrow down gap between femoral and polyethylene component used knee kinematic motion simulator, this accommodate deep knee flexion movement in daily activities and reduce the possibility of subluxation and dislocation at the polyethylene component during deep knee flexion. A wide gap between the femoral component and the polyethylene component and a significant amount of contact force in the medial condyle region might be the explanation for polyethylene component damage. It is expected that potential medial or lateral condylar lift at some stage during the gait cycle can be reduced.
引用
收藏
页码:119 / 130
页数:12
相关论文
共 30 条
[21]   Patient satisfaction after knee arthroplasty - A report on 27,372 knees operated on between 1981 and 1995 in Sweden [J].
Robertsson, O ;
Dunbar, M ;
Pehrsson, T ;
Knutson, K ;
Lidgren, L .
ACTA ORTHOPAEDICA SCANDINAVICA, 2000, 71 (03) :262-267
[22]  
Stiehl J. B., 2013, BIOMED RES INT, DOI [10.1155/2013/717546, DOI 10.1155/2013/717546]
[23]   A prospective comparative analysis of mobility in osteoarthritic knees -: Does lifestyle have an influence? [J].
Szabó, G ;
Lovász, G ;
Kustos, T ;
Bener, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (08) :1167-1169
[24]   Influence of loading and activity on the primary stability of cementless tibial trays [J].
Taylor, Mark ;
Barrett, David S. ;
Deffenbaugh, Daren .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (09) :1362-1368
[25]   Current Failure Mechanisms After Knee Arthroplasty Have Changed: Polyethylene Wear Is Less Common in Revision Surgery [J].
Thiele, Kathi ;
Perka, Carsten ;
Matziolis, Georg ;
Mayr, Hermann Otto ;
Sostheim, Michael ;
Hube, Robert .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (09) :715-720
[26]   The flexion gap in normal knees - An MRI study [J].
Tokuhara, Y ;
Kadoya, Y ;
Nakagawa, S ;
Kobayashi, A ;
Takaoka, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (08) :1133-1136
[27]   Finite Element Analysis Contact Stresses on Tibiofemoral Joint and Post Polyethylene Components Used to Evaluated Predesign Knee Implant [J].
Triwardono, Joko ;
Saragih, Agung Shamsuddin ;
Abrar, Fuad ;
Utomo, M. Satrio ;
Kartika, Ika ;
Whulanza, Yudan ;
Supriadi, Sugeng .
JOURNAL OF BIOMIMETICS BIOMATERIALS AND BIOMEDICAL ENGINEERING, 2022, 55 :46-55
[28]   Evaluation of the Contact Area in Total Knee Arthroplasty Designed for Deep Knee Flexion [J].
Triwardono, Joko ;
Supriadi, Sugeng ;
Whulanza, Yudan ;
Saragih, Agung Shamsuddin ;
Novalianita, Deva Ariana ;
Utomo, Muhammad Satrio ;
Kartika, Ika .
INTERNATIONAL JOURNAL OF TECHNOLOGY, 2021, 12 (06) :1312-1322
[29]   Biomechanical Consequences of Tibial Insert Thickness after Total Knee Arthroplasty: A Musculoskeletal Simulation Study [J].
Tzanetis, Periklis ;
Marra, Marco A. ;
Fluit, Rene ;
Koopman, Bart ;
Verdonschot, Nico .
APPLIED SCIENCES-BASEL, 2021, 11 (05)
[30]   Kinematics of lower limbs of healthy Chinese people sitting cross-legged [J].
Zhou, Hai ;
Liu, Anmin ;
Wang, Dongmei ;
Zeng, Xiangsen ;
Wei, San ;
Wang, Chengtao .
PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2013, 37 (05) :369-374