Evaluation of the Malalignment Varus- Valgus in Total Knee Arthroplasty Designed for Deep Knee Flexion Used Knee Kinematic Motion Simulator

被引:1
作者
Triwardono, Joko [1 ]
Rokhmanto, Fendy [1 ]
Roberto, Rahadian [1 ]
Kartika, Ika [1 ]
Saragih, Agung Shamsuddin [2 ]
Whulanza, Yudan [2 ]
Shuib, Solehuddin [3 ]
Supriadi, Sugeng [2 ]
机构
[1] Natl Res & Innovat Agcy, Res Ctr Met, KST BJ Habibie, South Tangerang, Indonesia
[2] Univ Indonesia, Dept Mech Engn, Depok, Indonesia
[3] Univ Teknol MARA, Coll Engn, Sch Mech Engn, Shah Alam, Selangor, Malaysia
关键词
malalignment; hyperflexion; subluxations; dislocations; wear; polyethylene; PATIENT SATISFACTION; FAILURE; THICKNESS; COMPONENT; ALIGNMENT; FRACTURE; OUTCOMES; JOINT; POST;
D O I
10.4028/p-l6p55i
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Revision total knee arthroplasties cause performed aseptic loosening, instability, and polyethylene wear. Separation or removal of the femoral component has been observed and this has the potential to severely damage the polyethylene component. In most cases 90% of the patients examined experienced significant medial or lateral condylar lift at some stage during the gait cycle. Using the MRI, a normal knee has maximum lateral lift is approx. 6.7 mm and maximum medial lift is approx. 2.1 mm, when a varus strees applied at a 90 degrees knee flexion. Elevation of the lateral condyle due to valgus malalignment will distribute more contact force on the medial condyle. In this study, a polyethylene component of a posterior-stabilized right knee joint implant was developed to facilitate a high range of motion (ROM). Malalignment valgus was observed with the axes of knee motion joint implants were varied from 0 degrees, 2 degrees, 3 degrees to 5 and knee bend measurements at 30 degrees, 60 degrees, 90 degrees, 120 degrees, and 150 degrees of knee flexion. Using the knee kinematic motion simulator, the modified polyethylene component resulted in 0 degrees malalignment there is no gap of the femoral component with the polyethylene component, from 30 degrees to 150 degrees of knee flexion. At 2 degrees malalignment, the femoral component was raised by 0.5 mm at a 90 degrees to 150 degrees knee flexion and increased with increasing knee flexion. Maximum gap occurs at 5 degrees malalignment in the amount of 5 mm at 150 degrees of knee flexion. The aim of this study was therefore to evaluation malalignment valgus of the flexed knee using knee kinematic motion simulator, with reference to the tibiofemoral flexion gap. The result that the modified design is expected in an narrow down gap between femoral and polyethylene component used knee kinematic motion simulator, this accommodate deep knee flexion movement in daily activities and reduce the possibility of subluxation and dislocation at the polyethylene component during deep knee flexion. A wide gap between the femoral component and the polyethylene component and a significant amount of contact force in the medial condyle region might be the explanation for polyethylene component damage. It is expected that potential medial or lateral condylar lift at some stage during the gait cycle can be reduced.
引用
收藏
页码:119 / 130
页数:12
相关论文
共 30 条
  • [1] Traumatic fracture of the polyethylene tibial post and cone in a posterior-stabilized total knee arthroplasty: A case report
    Alkheraiji, Abdulaziz
    Borai, Sultana
    Alfadhil, Rheema
    AlJassir, Fawzi
    [J]. INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2022, 97
  • [2] Patient-reported Outcome Correlates With Knee Function After a Single-design Mobile-bearing TKA
    Argenson, Jean-Noel
    Parratte, Sebastien
    Ashour, Abdullah
    Komistek, Richard D.
    Scuderi, Giles R.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (11) : 2669 - 2676
  • [3] Choi Young-Joon, 2016, Knee Surg Relat Res, V28, P1, DOI 10.5792/ksrr.2016.28.1.1
  • [4] Impact of Polyethylene Thickness on Clinical Outcomes and Survivorship in Medial Mobile-Bearing Unicondylar Knee Arthroplasty
    Crawford, David A.
    Lapsley, Lauren
    Hurst, Jason M.
    Morris, Michael J.
    Lombardi, Adolph V., Jr.
    Berend, Keith R.
    [J]. JOURNAL OF ARTHROPLASTY, 2021, 36 (07) : 2440 - 2444
  • [5] The Relationship Between Expectations and Satisfaction in Patients Undergoing Primary Total Knee Arthroplasty
    Culliton, Sharon E.
    Bryant, Dianne M.
    Overend, Tom J.
    MacDonald, Steven J.
    Chesworth, Bert M.
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (03) : 490 - 492
  • [6] Femoral condylar lift-off in vivo in total knee arthroplasty
    Dennis, DA
    Komistek, RD
    Walker, SA
    Cheal, EJ
    Stiehl, JB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (01): : 33 - 39
  • [7] The Effect of Polyethylene Liner Thickness on Patient Outcomes and Failure After Primary Total Knee Arthroplasty
    Garceau, Simon P.
    Warschawski, Yaniv S.
    Tang, Alex
    Sanders, Ethan B.
    Schwarzkopf, Ran M.
    Backstein, David J.
    [J]. JOURNAL OF ARTHROPLASTY, 2020, 35 (08) : 2072 - 2075
  • [8] Kinematics of the knee joint in deep flexion: a radiographic assessment
    Hefzy, MS
    Kelly, BP
    Cooke, TDV
    [J]. MEDICAL ENGINEERING & PHYSICS, 1998, 20 (04) : 302 - 307
  • [9] Manipulation of Tibial Component to Ensure Avoidance of Bearing Separation from the Vertical Wall of Tibial Component in Oxford Unicompartmental Arthroplasty
    Hiranaka, Takafumi
    Tanaka, Toshikazu
    Okimura, Kenjiro
    Fujishiro, Takaaki
    Shigemoto, Rika
    Araki, Shotaro
    Okada, Ryo
    Nako, Ryohei
    Kamenaga, Tomoyuki
    Okamoto, Koji
    [J]. CLINICS IN ORTHOPEDIC SURGERY, 2021, 13 (01) : 123 - 126
  • [10] Femoral component sizing in total knee arthroplasty - Size matched resection versus flexion space balancing
    Incavo, SJ
    Coughlin, KM
    Beynnon, BD
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (04) : 493 - 497