Cuff-less blood pressure estimation using hetero-core optical fibers

被引:2
作者
Koyama, Yuya [1 ,3 ]
Kamada, Hiroki [1 ]
Watanabe, Kazuhiro [2 ]
Nishiyama, Michiko [2 ]
机构
[1] Chiba Inst Technol, Dept Elect & Elect Engn, Narashino, Chiba, Japan
[2] Soka Univ, Fac Sci & Engn, Hachioji, Tokyo, Japan
[3] Chiba Inst Technol 2, Dept Elect & Elect Engn, Tsudanuma 2-17-1, Narashino, Chiba 2750016, Japan
关键词
cuff-less blood pressure measurement; hetero-core optical fiber; noninvasive measurement; pulse transit time; SENSOR;
D O I
10.1002/ecj.12438
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this study, we proposed a cuff-less blood pressure measurement using a medical tape sensor with an embedded hetero-core optical fiber. The tape sensor pasted on the skin can directly detect pulse wave signals with a simple measuring device. Pulse transit time (PTT) was obtained from the time difference between the two pulse waveforms at the neck and foot. Systolic blood pressure (SBP) was estimated from PTT and the regression equation. The result shows that the maximum prediction error was 3.8 mmHg in five healthy young subjects. The SBP data were evaluated using a Bland-Altman method to assess the agreement of SBP data estimated by the proposed method with those measured by the cuff-based electronic sphygmomanometer. The mean +/- 1.96 standard deviation of the estimated SBP against the reference was -1.3 +/- 5.1 mmHg. The results show that the proposed noninvasive blood pressure measurement method can measure blood pressure, and its accuracy is comparable to that of the cuff-based electronic sphygmomanometer.
引用
收藏
页数:8
相关论文
共 23 条
  • [1] Chino S., 2020, T SICE, V56, P189
  • [2] Dohi T., 2016, IEEJ T SM, V136, P370
  • [3] Fujita R., 2018, IEEJ T SM, V138, P539, DOI DOI 10.1541/IEEJSMAS.138.539
  • [4] Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method
    Gesche, Heiko
    Grosskurth, Detlef
    Kuechler, Gert
    Patzak, Andreas
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (01) : 309 - 315
  • [5] Japanese Society of Hypertension, 2019, LIFE SCI PUBL
  • [6] Japanese Society of Hypertension, 2011, LIFE SCI PUBL
  • [7] Kamada H., 2021, P 38TH SENS S SENS M, P1
  • [8] Kawakami K., 2021, IEEJ T SM, V141, P186
  • [9] Kotake Y., 2009, J JPN SOC INTENSIVE, V16, P263, DOI DOI 10.3918/jsicm.16.263
  • [10] Koyama S, 2019, SEN-I GAKKAISHI, V75, P641