Impact of Fingernail Polish on Pulse Oximetry Measurements: A Systematic Review

被引:4
作者
Aggarwal, Ashutosh Nath [1 ]
Agarwal, Ritesh [1 ]
Dhooria, Sahajal [1 ]
Prasad, Kuruswamy Thurai [1 ]
Sehgal, Inderpaul Singh [1 ]
Muthu, Valliappan [1 ]
机构
[1] Postgrad Inst Med Educ & Res, Dept Pulm Med, Chandigarh 160012, India
关键词
color; nail polish; oxygen saturation; pulse oximetry; systematic review; NAIL POLISH; OXYGEN-SATURATION; INTERFERENCE; HENNA;
D O I
10.4187/respcare.10399
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The effect of application of fingernail polish on SpO2 measurement remains unclear. We conducted this systematic review to ascertain the impact of fingernail polish on SpO2 measurement. METHODS: We queried PubMed, Embase, and CINAHL databases for publica-tions indexed through December 2022. We included studies providing paired SpO2 data from fin-gertips without and after nail polish application or reporting the number of subjects whose SpO2 could not be measured due to fingernail polish. We used random effects modeling to summarize standardized mean differences (SMDs) and corresponding 95% CI for different nail polish colors from comparative studies. RESULTS: We retrieved 122 studies and included 21 publications, mostly performed on healthy volunteers. Of these, 17 (81.0%) studies had a low risk of bias. We summarized mean SMD for 10 nail polish colors (black, blue, brown, green, orange, pink, purple, red, white, and yellow) from 25 paired data sets on SpO2 across 20 studies. We found small (likely clinically insignificant) but statistically significant differences in mean SpO2 when fingers were coated with black, blue, brown, or purple nail polish (SMD -0.57, -0.47, -0.33, and -0.25, respectively; 95% CI -0.86 to -0.29, -0.84 to -0.10, -0.59 to -0.07, and -0.48 to -0.02, respectively). Only one of 4 studies reported a high proportion of unsuccessful oxim-eter readings from fingers painted with black (88.0%) or brown (36.0%) nail polish. CONCLUSIONS: Although fingernail polish of some colors can marginally reduce SpO2 reading or occasionally impede SpO2 measurement, the variability is clinically insignificant.
引用
收藏
页码:1271 / 1280
页数:10
相关论文
共 40 条
  • [1] THE USE OF A VINYL GLOVE DOES NOT AFFECT PULSE OXIMETER MONITORING
    ACKERMAN, WE
    JUNEJA, M
    BAUMANN, RC
    KACZOROWSKI, DM
    [J]. ANESTHESIOLOGY, 1989, 70 (03) : 558 - 559
  • [2] ALMAJED SA, 1994, TROP GEOGR MED, V46, P38
  • [3] [Anonymous], 1991, Respir Care, V36, P1406
  • [4] Ballesteros-Peña S, 2015, EMERGENCIAS, V27, P325
  • [5] OPERATING CHARACTERISTICS OF A BANK CORRELATION TEST FOR PUBLICATION BIAS
    BEGG, CB
    MAZUMDAR, M
    [J]. BIOMETRICS, 1994, 50 (04) : 1088 - 1101
  • [6] Enamel nail polish does not interfere with pulse oximetry among normoxic volunteers
    Brand T.M.
    Brand M.E.
    Jay G.D.
    [J]. Journal of Clinical Monitoring and Computing, 2002, 17 (02) : 93 - 96
  • [7] Perioperative management of nail polish, artificial nails and piercings in The Netherlands
    Bucx, Martin J. L.
    Krijtenburg, Piet
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2016, 33 (03) : 223 - 231
  • [8] What is the effect of fingernail polish on pulse oximetry?
    Chan, MM
    Chan, MM
    Chan, ED
    [J]. CHEST, 2003, 123 (06) : 2163 - 2164
  • [9] Effect of nail polish and henna on oxygen saturation determined by pulse oximetry in healthy young adult females
    Cicek, Hatice Sutcu
    Gumus, Seyfettin
    Deniz, Omer
    Yildiz, Safak
    Acikel, Cengiz Han
    Cakir, Erdinc
    Tozkoparan, Ergun
    Ucar, Ergun
    Bilgic, Hayati
    [J]. EMERGENCY MEDICINE JOURNAL, 2011, 28 (09) : 783 - 785
  • [10] Cohen J., 1988, STAT POWER ANAL BEHA