共 50 条
Accuracy of the Apple Watch Oxygen Saturation Measurement in Adults: A Systematic Review
被引:9
|作者:
Windisch, Paul
[1
]
Schroeder, Christina
[1
]
Foerster, Robert
[1
]
Cihoric, Nikola
[2
]
Zwahlen, Daniel R.
[1
]
机构:
[1] Kantonsspital Winterthur, Dept Radiat Oncol, Winterthur, Switzerland
[2] Univ Hosp Bern, Inselspital, Dept Radiat Oncol, Bern, Switzerland
关键词:
telehealth;
wearables;
apple watch;
pulse oximetry;
oxygen saturation;
digital health;
D O I:
10.7759/cureus.35355
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
The purpose of this review is to summarize the research on the accuracy of oxygen saturation (spO(2)) measurements using the Apple Watch (Apple Inc., Cupertino, California). The Medline and Google Scholar databases were searched for papers evaluating the spO(2) measurements of the Apple Watch vs. any kind of ground truth and records were analyzed according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines. The five publications with 973 total patients that met the inclusion criteria all used the Apple Watch Series 6 and described 95% limits of agreement of +/-2.7 to 5.9% spO(2). However, outliers of up to 15% spO(2) were reported. Only one study had patient-level data uploaded to a public repository. The Apple Watch Series 6 does not show a strong systematic bias compared to conventional, medical-grade pulse oximeters. However, outliers do occur and should not cause concern in otherwise healthy individuals. The impact of race on measurement accuracy should be investigated.
引用
收藏
页数:7
相关论文