Two-dimensional time-resolved scintillating sheet monitoring of proton pencil beam scanning FLASH mouse irradiations

被引:3
|
作者
Kanouta, Eleni [1 ,2 ]
Bruza, Petr [3 ]
Johansen, Jacob Graversen [1 ,2 ,4 ]
Kristensen, Line [1 ,2 ,5 ]
Sorensen, Brita Singers [1 ,2 ,5 ]
Poulsen, Per Rugaard [1 ,2 ,4 ]
机构
[1] Aarhus Univ Hosp, Danish Ctr Particle Therapy, Aarhus, Denmark
[2] Aarhus Univ, Fac Hlth, Dept Clin Med, Aarhus, Denmark
[3] Dartmouth Coll, Thayer Sch Engn, Hanover, NH USA
[4] Aarhus Univ Hosp, Dept Oncol, Aarhus, Denmark
[5] Aarhus Univ Hosp, Dept Expt Clin Oncol, Aarhus, Denmark
关键词
FLASH radiotherapy; in vivo measurement; proton PBS; scintillation imaging; DOSIMETRY;
D O I
10.1002/mp.17049
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundDosimetry in pre-clinical FLASH studies is essential for understanding the beam delivery conditions that trigger the FLASH effect. Resolving the spatial and temporal characteristics of proton pencil beam scanning (PBS) irradiations with ultra-high dose rates (UHDR) requires a detector with high spatial and temporal resolution.PurposeTo implement a novel camera-based system for time-resolved two-dimensional (2D) monitoring and apply it in vivo during pre-clinical proton PBS mouse irradiations.MethodsTime-resolved 2D beam monitoring was performed with a scintillation imaging system consisting of a 1 mm thick transparent scintillating sheet, imaged by a CMOS camera. The sheet was placed in a water bath perpendicular to a horizontal PBS proton beam axis. The scintillation light was reflected through a system of mirrors and captured by the camera with 500 frames per second (fps) for UHDR and 4 fps for conventional dose rates. The raw images were background subtracted, geometrically transformed, flat field corrected, and spatially filtered. The system was used for 2D spot and field profile measurements and compared to radiochromic films. Furthermore, spot positions were measured for UHDR irradiations. The measured spot positions were compared to the planned positions and the relative instantaneous dose rate to equivalent fiber-coupled point scintillator measurements. For in vivo application, the scintillating sheet was placed 1 cm upstream the right hind leg of non-anaesthetized mice submerged in the water bath. The mouse leg and sheet were both placed in a 5 cm wide spread-out Bragg peak formed from the mono-energetic proton beam by a 2D range modulator. The mouse leg position within the field was identified for both conventional and FLASH irradiations. For the conventional irradiations, the mouse foot position was tracked throughout the beam delivery, which took place through repainting. For FLASH irradiations, the delivered spot positions and relative instantaneous dose rate were measured.ResultsThe pixel size was 0.1 mm for all measurements. The spot and field profiles measured with the scintillating sheet agreed with radiochromic films within 0.4 mm. The standard deviation between measured and planned spot positions was 0.26 mm and 0.35 mm in the horizontal and vertical direction, respectively. The measured relative instantaneous dose rate showed a linear relation with the fiber-coupled scintillator measurements. For in vivo use, the leg position within the field varied between mice, and leg movement up to 3 mm was detected during the prolonged conventional irradiations.ConclusionsThe scintillation imaging system allowed for monitoring of UHDR proton PBS delivery in vivo with 0.1 mm pixel size and 2 ms temporal resolution. The feasibility of instantaneous dose rate measurements was demonstrated, and the system was used for validation of the mouse leg position within the field.
引用
收藏
页码:5119 / 5129
页数:11
相关论文
共 5 条
  • [1] Experimental characterization of two-dimensional pencil beam scanning proton spot profiles
    Lin, Liyong
    Ainsley, Christopher G.
    McDonough, James E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (17) : 6193 - 6204
  • [2] Experimental characterization of two-dimensional spot profiles for two proton pencil beam scanning nozzles
    Lin, Liyong
    Ainsley, Christopher G.
    Solberg, Timothy D.
    McDonough, James E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (02) : 493 - 504
  • [3] Feasibility study of a plastic scintillating plate-based treatment beam fluence monitoring system for use in pencil beam scanning proton therapy
    Jeong, Seonghoon
    Chung, Kwangzoo
    Ahn, Sung Hwan
    Lee, Boram
    Seo, Jaehyeon
    Yoon, Myonggeun
    MEDICAL PHYSICS, 2020, 47 (02) : 703 - 712
  • [4] Development of a new 3-dimensional scintillating detector for patient treatment quality control in pencil beam scanning proton therapy
    Frelin, Anne-Marie
    Daviau, G.
    Bui, My Hoang Hoa
    Fontbonne, Cathy
    Fontbonne, Jean-Marc
    Lebhertz, Dorothee
    Mainguy, Erwan
    Moignier, Cyril
    Vela, Anthony
    Thariat, Juliette
    13TH INTERNATIONAL CONFERENCE ON 3D DOSIMETRY, IC3DDOSE 2024, 2024, 2799
  • [5] A dose rate independent 2D Ce-doped YAG scintillating dosimetry system for time resolved beam monitoring in ultra-high dose rate electron "FLASH"radiation therapy
    Vanreusel, Verdi
    Heinrich, Sophie
    De Kerf, Thomas
    Leblans, Paul
    Vandenbroucke, Dirk
    Vanlanduit, Steve
    Verellen, Dirk
    Gasparini, Alessia
    Nascimento, Luana de Freitas
    SENSORS AND ACTUATORS A-PHYSICAL, 2024, 371