Motion impedance cardiography denoising method based on canonical correlation analysis and coherence analysis

被引:1
|
作者
Xie, Yao [1 ,3 ]
Yu, Honglong [2 ,3 ]
Xie, Qilian [3 ,4 ]
机构
[1] Univ Sci & Technol China, Sch Engn Sci, Hefei, Peoples R China
[2] Hefei Univ Technol, Dept Biomed Engn, Hefei, Peoples R China
[3] Anhui Tongling Bion Technol Co Ltd, Hefei, Peoples R China
[4] Anhui Med Univ, Hefei, Peoples R China
关键词
Impedance cardiography; Artifacts; Canonical correlation analysis; Coherence analysis; CARDIAC-OUTPUT; STROKE VOLUME; PARAMETERS; FILTER; TOOL;
D O I
10.1016/j.bspc.2023.105300
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Impedance cardiography (ICG) is an attractive noninvasive method for measuring stroke volume and cardiac output. However, it is easily disturbed by the artifacts such as respiration and body shaking. Considering the correlation between physiological and motion signals, and the synchronization relationship among the physiological signals, this paper introduces an ICG denoising method based on canonical correlation analysis (CCA) and coherence analysis to remove the artifacts during the measurement. First, the CCA extracts the shared components between the electrocardiogram (ECG) and the motion signal, as well as the shared components between ICG and the motion signal. Then, set those shared components to zero to suppress the primary motion artifacts. Next, the coherence analysis was used to calculate the synchronization relationship between obtained ECG and ICG components. Finally, the components with strong synchronization relationships were used to reconstruct the ICG signal. The denoising method was evaluated for 54 subjects during lying and walking. Experimental results show that after removing the artifacts, the signal quality index beat contribution factor (BCF) was increased from the original 78.1% to 97.8%, and the physiological parameters measured based on the proposed method were in good agreement with those measured by the standard instrument. The proposed denoising method effectively improves the reliability of analysis and diagnosis on cardiovascular diseases relying on ICG signals.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Research on motion impedance cardiography de-noising method based on two-step spectral ensemble empirical mode decomposition and canonical correlation analysis
    Xie, Yao
    Yang, Dong
    Yu, Honglong
    Xie, Qilian
    Shengwu Yixue Gongchengxue Zazhi/Journal of Biomedical Engineering, 2024, 41 (05): : 986 - 994
  • [2] A CANONICAL CORRELATION ANALYSIS BASED MOTION MODEL FOR PROBABILISTIC VISUAL TRACKING
    Heyman, Tom
    Spruyt, Vincent
    Grunwedel, Sebastian
    Ledda, Alessandro
    Philips, Wilfried
    2012 IEEE VISUAL COMMUNICATIONS AND IMAGE PROCESSING (VCIP), 2012,
  • [3] Stroke volume assessment by impedance cardiography: Comparative analysis with transthoracic echocardiography
    Silluta, S.
    Pilt, K.
    Bischler, E.
    Koots, K.
    Meigas, K.
    Viigimaa, M.
    EMBEC & NBC 2017, 2018, 65 : 398 - 401
  • [4] COMPUTER-SUPPORTED ANALYSIS OF CARDIOVASCULAR PARAMETERS BY IMPEDANCE CARDIOGRAPHY AND PLETHYSMOGRAPHY
    HUNOLD, S
    HEXAMER, M
    WERNER, J
    GEHRT, M
    MEDICAL PROGRESS THROUGH TECHNOLOGY, 1993, 19 (02) : 105 - 113
  • [5] Time Delay Estimation Method Based on Canonical Correlation Analysis
    Lim, Jun-seok
    Pang, Hee-Suk
    CIRCUITS SYSTEMS AND SIGNAL PROCESSING, 2013, 32 (05) : 2527 - 2538
  • [6] Time Delay Estimation Method Based on Canonical Correlation Analysis
    Jun-seok Lim
    Hee-Suk Pang
    Circuits, Systems, and Signal Processing, 2013, 32 : 2527 - 2538
  • [7] Denoising of HD-sEMG signals using canonical correlation analysis
    Al Harrach, M.
    Boudaoud, S.
    Hassan, M.
    Ayachi, F. S.
    Gamet, D.
    Grosset, J. F.
    Marin, F.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2017, 55 (03) : 375 - 388
  • [8] Denoising of HD-sEMG signals using canonical correlation analysis
    M. Al Harrach
    S. Boudaoud
    M. Hassan
    F. S. Ayachi
    D. Gamet
    J. F. Grosset
    F. Marin
    Medical & Biological Engineering & Computing, 2017, 55 : 375 - 388
  • [9] A fault detection method based on sparse dynamic canonical correlation analysis
    Hu, Xuguang
    Wu, Ping
    Pan, Haipeng
    He, Yuchen
    CANADIAN JOURNAL OF CHEMICAL ENGINEERING, 2024, 102 (03) : 1188 - 1202
  • [10] A meta-analysis of three decades of validating thoracic impedance cardiography
    Raaijmakers, E
    Faes, TJC
    Scholten, RJPM
    Goovaerts, HG
    Heethaar, RM
    CRITICAL CARE MEDICINE, 1999, 27 (06) : 1203 - 1213