A vibration correction system for cold atom gravimeter

被引:4
作者
Guo, Meiying [1 ]
Bai, Jinhai [1 ]
Hu, Dong [1 ]
Tang, Zhimin [1 ]
You, Jianqi [1 ]
Chen, Runnan [1 ]
Wang, Yu [1 ]
机构
[1] Changcheng Inst Metrol & Measurement, Beijing 100095, Peoples R China
基金
中国国家自然科学基金;
关键词
cold atom gravimeter; ground vibration; vibration correction; seismometer;
D O I
10.1088/1361-6501/ad128d
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Cold atom gravimeters are used to measure the absolute value of the gravitational acceleration. To mitigate the influence of seismic noise on the mirror, a vibration correction system is used. A commercial seismometer CMG-3ESPC is employed to measure the mirror vibration. The transfer function between the mirror vibration and the output of the seismometer is regarded as a proportional element with a time delay. The best proportional element and the best time delay are obtained by minimizing the standard deviation of the fitting residuals of the interference fringe alpha-P with vibration correction. This system not only exhibits excellent environmental adaptability but also eliminates the need for rough estimation of local gravitational acceleration during the vibration correction process. The system was built on the homemade cold atom gravimeter to carry out experiments in the laboratory. Compared with the measurement results without vibration correction, the standard deviation of the gravity measurement results was reduced from 120.5 mu Gal to 4.2 mu Gal, resulting in a significant reduction by a factor of 28.7. Meanwhile, its sensitivity was significantly improved from 330.8 mu GalHz-1 to 18.0 mu GalHz-1 by performing vibration correction. In the future, this system is expected to carry out comprehensive verification experiments in more complex noisy environments.
引用
收藏
页数:10
相关论文
共 26 条
  • [1] [Anonymous], 2021, Absolute quantum gravimeter
  • [2] A cold atom pyramidal gravimeter with a single laser beam
    Bodart, Q.
    Merlet, S.
    Malossi, N.
    Dos Santos, F. Pereira
    Bouyer, P.
    Landragin, A.
    [J]. APPLIED PHYSICS LETTERS, 2010, 96 (13)
  • [3] Brown JM, 2002, IAG SYMP, V123, P223
  • [4] Canuteson E, 1997, B SEISMOL SOC AM, V87, P484
  • [5] Chen B., 2020, PhD Thesis
  • [6] Chen B., 2018, Vibration noise correction method for atomic interferometric gravimeters China, Patent No. [CN201811413684.2, 2018114136842]
  • [7] Farah T., 2011, 2011 JOINT C IEEE IN, P1
  • [8] Guralp Systems Limited, 2016, CMG-3ESP operator's guide Guralp systems limited
  • [9] ATOMIC INTERFEROMETRY USING STIMULATED RAMAN TRANSITIONS
    KASEVICH, M
    CHU, S
    [J]. PHYSICAL REVIEW LETTERS, 1991, 67 (02) : 181 - 184
  • [10] Limits to the sensitivity of a low noise compact atomic gravimeter
    Le Gouet, J.
    Mehlstaeubler, T. E.
    Kim, J.
    Merlet, S.
    Clairon, A.
    Landragin, A.
    Dos Santos, F. Pereira
    [J]. APPLIED PHYSICS B-LASERS AND OPTICS, 2008, 92 (02): : 133 - 144