High-Speed On-Site Deep Learning-Based FFR-CT Algorithm: Evaluation Using Invasive Angiography as the Reference Standard

被引:6
作者
Giannopoulos, Andreas A. [1 ]
Keller, Lukas [1 ]
Sepulcri, Daniel [1 ]
Boehm, Reto [1 ]
Garefa, Chrysoula [1 ]
Venugopal, Prem [2 ]
Mitra, Jhimli [2 ]
Ghose, Soumya [2 ]
Deak, Paul [2 ]
Pack, Jed D. [2 ]
Davis, Cynthia L. [2 ]
Stahli, Barbara E. [3 ]
Stehli, Julia [3 ]
Pazhenkottil, Aju P. [1 ]
Kaufmann, Philipp A. [1 ]
Buechel, Ronny R. [1 ]
机构
[1] Univ Zurich Hosp, Dept Nucl Med, Cardiac Imaging, Ramistr 100, CH-8091 Zurich, Switzerland
[2] GE Res, Schenectady, NY USA
[3] Univ Zurich Hosp, Univ Heart Ctr, Dept Cardiol, Zurich, Switzerland
关键词
coronary CTA; deep learning; FFR-CT; fractional flow reserve; invasive angiography; validation study; FRACTIONAL FLOW RESERVE; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; DIAGNOSTIC PERFORMANCE; CARDIAC CT; DISEASE; RADIATION;
D O I
10.2214/AJR.23.29156
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BACKGROUND. Estimation of fractional flow reserve from coronary CTA (FFR-CT) is an established method of assessing the hemodynamic significance of coronary lesions. However, clinical implementation has progressed slowly, partly because of off-site data transfer with long turnaround times for results. OBJECTIVE. The purpose of this study was to evaluate the diagnostic performance of FFR-CT computed on-site with a high-speed deep learning-based algorithm with invasive hemodynamic indexes as the reference standard. METHODS. This retrospective study included 59 patients (46 men, 13 women; mean age, 66.5 +/- 10.2 years) who underwent coronary CTA (including calcium scoring) followed within 90 days by invasive angiography with invasive fractional flow reserve (FFR) and/ or instantaneous wave-free ratio measurements from December 2014 to October 2021. Coronary artery lesions were considered to have hemodynamically significant stenosis in the presence of invasive FFR of 0.80 or less and/or instantaneous wave-free ratio of 0.89 or less. A single cardiologist evaluated the CTA images using an on-site deep learning-based semiautomated algorithm entailing a 3D computational flow dynamics model to determine FFR-CT for coronary artery lesions detected with invasive angiography. Time for FFR-CT analysis was recorded. FFR-CT analysis was repeated by the same cardiologist in 26 randomly selected examinations and by a different cardiologist in 45 randomly selected examinations. Diagnostic performance and agreement were assessed. RESULTS. A total of 74 lesions were identified with invasive angiography. FFR-CT and invasive FFR had strong correlation (r = 0.81) and, in Bland-Altman analysis, bias of 0.01 and 95% limits of agreement of - 0.13 to 0.15. FFR-CT had AUC for hemodynamically significant stenosis of 0.975. At a cutoff of 0.80 or less, FFR-CT had 95.9% accuracy, 93.5% sensitivity, and 97.7% specificity. In 39 lesions with severe calcifications (>= 400 Agatston units), FFR-CT had AUC of 0.991 and at a cutoff of 0.80, 94.7% sensitivity, 95.0% specificity, and 94.9% accuracy. Mean analysis time per patient was 7 minutes 54 seconds. Intraobserver agreement (intraclass correlation coefficient, 0.85; bias, -0.01; 95% limits of agreement, -0.12 and 0.10) and interobserver agreement (intraclass correlation coefficient, 0.94; bias, - 0.01; 95% limits of agreement, -0.08 and 0.07) were good to excellent. CONCLUSION. A high-speed on-site deep learning-based FFR-CT algorithm had excellent diagnostic performance for hemodynamically significant stenosis with high reproducibility. CLINICAL IMPACT. The algorithm should facilitate implementation of FFR-CT technology into routine clinical practice.
引用
收藏
页码:460 / 470
页数:11
相关论文
共 45 条
  • [31] Deep learning-based noise reduction for coronary CT angiography: using four-dimensional noise-reduction images as the ground truth
    Kobayashi, Takuma
    Nishii, Tatsuya
    Umehara, Kensuke
    Ota, Junko
    Ohta, Yasutoshi
    Fukuda, Tetsuya
    Ishida, Takayuki
    ACTA RADIOLOGICA, 2023, 64 (05) : 1831 - 1840
  • [32] Deep learning-based and hybrid-type iterative reconstructions for CT: comparison of capability for quantitative and qualitative image quality improvements and small vessel evaluation at dynamic CE-abdominal CT with ultra-high and standard resolutions
    Matsukiyo, Ryo
    Ohno, Yoshiharu
    Matsuyama, Takahiro
    Nagata, Hiroyuki
    Kimata, Hirona
    Ito, Yuya
    Ogawa, Yukihiro
    Murayama, Kazuhiro
    Kato, Ryoichi
    Toyama, Hiroshi
    JAPANESE JOURNAL OF RADIOLOGY, 2021, 39 (02) : 186 - 197
  • [33] Fast high-quality MRI protocol of the lumbar spine with deep learning-based algorithm: an image quality and scanning time comparison with standard protocol
    Marta Zerunian
    Francesco Pucciarelli
    Damiano Caruso
    Domenico De Santis
    Michela Polici
    Benedetta Masci
    Ilaria Nacci
    Antonella Del Gaudio
    Giuseppe Argento
    Andrea Redler
    Andrea Laghi
    Skeletal Radiology, 2024, 53 : 151 - 159
  • [34] Reducing windmill artifacts in clinical spiral CT using a deep learning-based projection raw data upsampling: Method and robustness evaluation
    Magonov, Jan
    Maier, Joscha
    Erath, Julien
    Sunnegardh, Johan
    Fournie, Eric
    Stierstorfer, Karl
    Kachelriess, Marc
    MEDICAL PHYSICS, 2024, 51 (03) : 1597 - 1616
  • [35] Fast high-quality MRI protocol of the lumbar spine with deep learning-based algorithm: an image quality and scanning time comparison with standard protocol
    Zerunian, Marta
    Pucciarelli, Francesco
    Caruso, Damiano
    De Santis, Domenico
    Polici, Michela
    Masci, Benedetta
    Nacci, Ilaria
    Del Gaudio, Antonella
    Argento, Giuseppe
    Redler, Andrea
    Laghi, Andrea
    SKELETAL RADIOLOGY, 2024, 53 (01) : 151 - 159
  • [36] Application of Deep Learning-Based Denoising Technique for Radiation Dose Reduction in Dynamic Abdominal CT: Comparison with Standard-Dose CT Using Hybrid Iterative Reconstruction Method
    Nagata, Motonori
    Ichikawa, Yasutaka
    Domae, Kensuke
    Yoshikawa, Kazuya
    Kanii, Yoshinori
    Yamazaki, Akio
    Nagasawa, Naoki
    Ishida, Masaki
    Sakuma, Hajime
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (04) : 1578 - 1587
  • [37] Application of Deep Learning-Based Denoising Technique for Radiation Dose Reduction in Dynamic Abdominal CT: Comparison with Standard-Dose CT Using Hybrid Iterative Reconstruction Method
    Motonori Nagata
    Yasutaka Ichikawa
    Kensuke Domae
    Kazuya Yoshikawa
    Yoshinori Kanii
    Akio Yamazaki
    Naoki Nagasawa
    Masaki Ishida
    Hajime Sakuma
    Journal of Digital Imaging, 2023, 36 : 1578 - 1587
  • [38] Deep learning-based reconstruction of virtual monoenergetic images of kVp-switching dual energy CT for evaluation of hypervascular liver lesions: Comparison with standard reconstruction technique
    Seo, June Young
    Joo, Ijin
    Yoon, Jeong Hee
    Kang, Hyo Jin
    Kim, Sewoo
    Kim, Jong Hyo
    Ahn, Chulkyun
    Lee, Jeong Min
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 154
  • [39] Deep learning-based and hybrid-type iterative reconstructions for CT: comparison of capability for quantitative and qualitative image quality improvements and small vessel evaluation at dynamic CE-abdominal CT with ultra-high and standard resolutions
    Ryo Matsukiyo
    Yoshiharu Ohno
    Takahiro Matsuyama
    Hiroyuki Nagata
    Hirona Kimata
    Yuya Ito
    Yukihiro Ogawa
    Kazuhiro Murayama
    Ryoichi Kato
    Hiroshi Toyama
    Japanese Journal of Radiology, 2021, 39 : 186 - 197
  • [40] Ultra-low-dose hepatic multiphase CT using deep learning-based image reconstruction algorithm focused on arterial phase in chronic liver disease: A non-inferiority study
    Lee, Hyun Joo
    Kim, Jin Sil
    Lee, Jeong Kyong
    Lee, Hye Ah
    Pak, Seongyong
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 159