Performance study of a 360° CZT camera for monitoring 177Lu-PSMA treatment

被引:12
作者
Vergnaud, Laure [1 ,2 ]
Badel, Jean-Noel [1 ,2 ]
Giraudet, Anne-Laure [1 ]
Kryza, David [1 ,3 ]
Mognetti, Thomas [1 ]
Baudier, Thomas [1 ,2 ]
Rida, Hanan [4 ]
Dieudonne, Arnaud [4 ]
Sarrut, David [1 ,2 ]
机构
[1] Ctr Lutte Canc Leon Berard, Lyon, France
[2] Univ Lyon 1, INSA Lyon, INSERM U1044, CREATIS,CNRS UMR 5220,Univ Lyon, Lyon, France
[3] Univ Claude Bernard Lyon 1, Hosp Civils Lyon, Univ Lyon, LAGEPP UMR 5007 CNRS, Lyon, France
[4] Ctr Henri Becquerel, Dept Med Nucl, Rouen, France
关键词
CZT camera; SPECT; Lu-177; Theranostic; Internal radiotherapy; RECEPTOR RADIONUCLIDE THERAPY; RESISTANT PROSTATE-CANCER; RADIOLIGAND THERAPY; QUANTITATIVE SPECT; IN-VIVO; DOSIMETRY; VALIDATION;
D O I
10.1186/s40658-023-00576-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The aim of this study was to investigate the quantification performance of a 360 degrees CZT camera for Lu-177-based treatment monitoring.Methods: Three phantoms with known Lu-177 activity concentrations were acquired: (1) a uniform cylindrical phantom for calibration, (2) a NEMA IEC body phantom for analysis of different-sized spheres to optimise quantification parameters and (3) a phantom containing two large vials simulating organs at risk for tests. Four sets of reconstruction parameters were tested: (1) Scatter, (2) Scatter and Point Spread Function Recovery (PSFR), (3) PSFR only and (4) Penalised likelihood option and Scatter, varying the number of updates (iterations x subsets) with CT-based attenuation correction only. For each, activity concentration (ARC) and contrast recovery coefficients (CRC) were estimated as well as root mean square. Visualisation and quantification parameters were applied to reconstructed patient image data.Results: Optimised quantification parameters were determined to be: CT-based attenuation correction, scatter correction, 12 iterations, 8 subsets and no filter. ARC, CRC and RMS results were dependant on the methodology used for calculations. Two different reconstruction parameters were recommended for visualisation and for quantification. 3D whole-body SPECT images were acquired and reconstructed for Lu-177-PSMA patients in 2-3 times faster than the time taken for a conventional gamma camera.Conclusion: Quantification of whole-body 3D images of patients treated with Lu-177-PSMA is feasible and an optimised set of parameters has been determined. This camera greatly reduces procedure time for whole-body SPECT.
引用
收藏
页数:18
相关论文
共 49 条
[1]   Therapeutic response and side effects of repeated radioligand therapy with 177Lu-PSMA-DKFZ-617 of castrate-resistant metastatic prostate cancer [J].
Ahmadzadehfar, Hojjat ;
Eppard, Elisabeth ;
Kuerpig, Stefan ;
Fimmers, Rolf ;
Yordanova, Anna ;
Schlenkhoff, Carl Diedrich ;
Gaertner, Florian ;
Rogenhofer, Sebastian ;
Essler, Markus .
ONCOTARGET, 2016, 7 (11) :12477-12488
[2]   Quantitative 177Lu SPECT (QSPECT) imaging using a commercially available SPECT/CT system [J].
Beauregard, Jean-Mathieu ;
Hofman, Michael S. ;
Pereira, Jucilene M. ;
Eu, Peter ;
Hicks, Rodney J. .
CANCER IMAGING, 2011, 11 (01) :56-66
[3]   Highlights in nuclear cardiology: Applications of new CZT cameras [J].
Ben Bouallegue, F. ;
D'estanque, E. ;
Mariano-Goulart, D. .
MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2015, 39 (05) :435-443
[4]   High-quality brain perfusion SPECT images may be achieved with a high-speed recording using 360° CZT camera [J].
Bordonne, Manon ;
Chawki, Mohammad B. ;
Marie, Pierre-Yves ;
Zaragori, Timothee ;
Roch, Veronique ;
Grignon, Rachel ;
Imbert, Laetitia ;
Verger, Antoine .
EJNMMI PHYSICS, 2020, 7 (01)
[5]   Feasibility of 177Lu Therapy Monitoring Using Fast Whole-Body SPECT Recordings Provided by a High-Speed 360° CZT Camera [J].
Chevalier, Elodie ;
Boursier, Caroline ;
Claudin, Marine ;
Marie, Pierre-Yves ;
Imbert, Laetitia .
CLINICAL NUCLEAR MEDICINE, 2020, 45 (11) :E493-E494
[6]   Gamma camera calibration and validation for quantitative SPECT imaging with 177Lu [J].
D'Arienzo, M. ;
Cazzato, M. ;
Cozzella, M. L. ;
Cox, M. ;
D'Andrea, M. ;
Fazio, A. ;
Fenwick, A. ;
Iaccarino, G. ;
Johansson, L. ;
Strigari, L. ;
Ungania, S. ;
De Felice, P. .
APPLIED RADIATION AND ISOTOPES, 2016, 112 :156-164
[7]   Improving quantitative dosimetry in 177 Lu-DOTATATE SPECT by energy window-based scatter corrections [J].
de Nijs, Robin ;
Lagerburg, Vera ;
Klausen, Thomas L. ;
Holm, Soren .
NUCLEAR MEDICINE COMMUNICATIONS, 2014, 35 (05) :522-533
[8]   Dosimetry for 177Lu-DKFZ-PSMA-617: a new radiopharmaceutical for the treatment of metastatic prostate cancer [J].
Delker, Andreas ;
Fendler, Wolfgang Peter ;
Kratochwil, Clemens ;
Brunegraf, Anika ;
Gosewisch, Astrid ;
Gildehaus, Franz Josef ;
Tritschler, Stefan ;
Stief, Christian Georg ;
Kopka, Klaus ;
Haberkorn, Uwe ;
Bartenstein, Peter ;
Boening, Guido .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (01) :42-51
[9]   Biodistribution and dosimetry for combined [177Lu]Lu-PSMA-I&T/[225Ac]Ac-PSMA-I&T therapy using multi-isotope quantitative SPECT imaging [J].
Delker, Astrid ;
Schleske, Mirjam ;
Liubchenko, Grigory ;
Berg, Isabella ;
Zacherl, Mathias Johannes ;
Brendel, Matthias ;
Gildehaus, Franz Josef ;
Rumiantcev, Mikhail ;
Resch, Sandra ;
Huerkamp, Kerstin ;
Wenter, Vera ;
Unterrainer, Lena M. ;
Bartenstein, Peter ;
Ziegler, Sibylle, I ;
Beyer, Leonie ;
Boening, Guido .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (05) :1280-1290
[10]   A MODIFIED EXPECTATION MAXIMIZATION ALGORITHM FOR PENALIZED LIKELIHOOD ESTIMATION IN EMISSION TOMOGRAPHY [J].
DEPIERRO, AR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (01) :132-137