Exploitation of the enantioselectivity space of coated amylose tris(3,5-dimethylphenylcarbamate) in mixtures of 2-propanol and acetonitrile

被引:5
作者
Horvath, Simon [1 ]
Thuy, Hong Ha Nguyen [1 ]
Eke, Zsuzsanna [2 ]
Nemeth, Gabor [1 ]
机构
[1] Egis Pharmaceut PLC, Drug Subst Dev Div, POB 100, H-1475 Budapest, Hungary
[2] Eotvos Lorand Univ, Joint Res & Training Lab Separat Sci, Pazmany Peter Stny 1-A, H-1117 Budapest, Hungary
关键词
Chiral HPLC; Eluent history; Hysteresis; Hindered structural transition; Enantiomer elution order; CHIRAL STATIONARY PHASES; VIBRATIONAL CIRCULAR-DICHROISM; SOLID-STATE NMR; LIQUID-CHROMATOGRAPHY; SEPARATION; RECOGNITION; ENANTIOSEPARATION; ENANTIOMERS; DERIVATIVES; RESOLUTION;
D O I
10.1016/j.chroma.2023.464161
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Chiral stationary phases (CSPs) with coated amylose tris(3,5-dimethylphenylcarbamate) (ADMPC) selector have long been recognized for their excellent chiral recognition ability in liquid chromatography. The conformational versatility behind this feature is the source of their known hysteretic behavior, which has been previously observed in polar organic (PO) mode eluents containing 2-propanol (IPA). Mixtures of IPA and acetonitrile (MeCN), a typical PO mode eluent system, have not been examined in this aspect yet, even though hysteresis is promising for finding unique unexplored enantioselectivities.Not only was the hysteresis detectable on ADMPC using mixtures of IPA and MeCN, but it was the typical behavior in a diverse set of test compounds. The difference in the retention time of the same analyte under conditions which only differed in the eluent history on the column can go up to 20-fold. The assumed hindered conformational changes of the selector were reflected in retention drift at certain eluent compositions. On the two sides of the transitions, distinct, useful states of the selector were de-tected. A series of IPA - MeCN compositions with defined pretreatment was selected and recommended as an extension of the preliminary, first choice method screening set that used only alcohols. The incor-poration of a solvent possessing substantially different characteristics enhances the potential in practical applications, while keeping the technical simplicity. Stability and robustness of the additional states of the CSP were characterized. The examined columns of different brands shared the observed behavior. Kinetic stability of a column state is adequate for successful application.The evaluated states of ADMPC provide multiple enantiorecognition potential by using mixtures of IPA and MeCN also considering the pretreatment of the column. Unprecedented double and triple elution order reversals along the composition range supported the versatility of the available states. Our findings further enhance the usefulness of ADMPC-containing CSPs. We provide instructions for the application of the widespread chiral selector in common eluent mixtures to avoid pitfalls regarding reproducibility and robustness.& COPY; 2023 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 31 条
  • [21] Recognition in the Domain of Molecular Chirality: From Noncovalent Interactions to Separation of Enantiomers
    Peluso, Paola
    Chankvetadze, Bezhan
    [J]. CHEMICAL REVIEWS, 2022, 122 (16) : 13235 - 13400
  • [22] Chiral recognition in separation science - an update
    Scriba, Gerhard K. E.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2016, 1467 : 56 - 78
  • [23] Enantiomeric separation of mineralocorticoid receptor (hMR) antagonists using the Chiralcel® OJ-H HPLC column with novel polar cosolvent eluent systems
    Sharp, V. Scott
    Kennedy, Joseph H.
    Belvo, Matfhew D.
    Williams, Jeffrey D.
    Risley, Donald S.
    Seest, Eric P.
    [J]. CHIRALITY, 2006, 18 (06) : 437 - 445
  • [24] Efficient Separation of Enantiomers Using Stereoregular Chiral Polymers
    Shen, Jun
    Okamoto, Yoshio
    [J]. CHEMICAL REVIEWS, 2016, 116 (03) : 1094 - 1138
  • [25] Synthesis and application of immobilized polysaccharide-based chiral stationary phases for enantioseparation by high-performance liquid chromatography
    Shen, Jun
    Ikai, Tomoyuki
    Okamoto, Yoshio
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2014, 1363 : 51 - 61
  • [26] Enantiomeric resolution of kielcorin derivatives by HPLC on polysaccharide stationary phases using multimodal elution
    Sousa, EP
    Tiritan, ME
    Oliveira, RV
    Afonso, CMM
    Cass, QB
    Pinto, MMM
    [J]. CHIRALITY, 2004, 16 (05) : 279 - 285
  • [27] Chiral chromatography method screening strategies: Past, present and future
    Tarafder, Abhijit
    Miller, Larry
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2021, 1638
  • [28] Effects of alcohol mobile-phase modifiers on the structure and chiral selectivity of amylose tris(3,5-dimethylphenylcarbamate) chiral stationary phase
    Wang, T
    Wenslow, RM
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 1015 (1-2) : 99 - 110
  • [29] Modeling Enantiomeric Separations as an Interfacial Process Using Amylose Tris(3,5-dimethylphenyl carbamate) (ADMPC) Polymers Coated on Amorphous Silica
    Wang, Xiaoyu
    Jameson, Cynthia J.
    Murad, Sohail
    [J]. LANGMUIR, 2020, 36 (05) : 1113 - 1124
  • [30] Solid-state NMR characterization of amylose tris(3,5-dimethylphenylcarbamate) chiral stationary-phase structure as a function of mobile-phase composition
    Wenslow, RM
    Wang, T