High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:14
|
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
收藏
页码:1428 / 1435
页数:8
相关论文
共 50 条
  • [41] Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography
    Yang, Junjie
    Yang, Xiaobo
    De Cecco, Carlo N.
    Duguay, Taylor M.
    Chen, Zhiye
    Tesche, Christian
    Schoepf, U. Joseph
    Chen, Yundai
    SCIENTIFIC REPORTS, 2017, 7
  • [42] Using Dual-source Photon-counting Detector CT to Simultaneously Quantify Fat and Iron Content: A Phantom Study
    Hao, Wanting
    Xu, Zhihan
    Lin, Huimin
    Yan, Fuhua
    ACADEMIC RADIOLOGY, 2024, 31 (10) : 4119 - 4128
  • [43] The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography
    Liu, Ying
    Xu, Jian
    Li, Jian
    Ren, Jing
    Liu, Hongtao
    Xu, Junqing
    Wei, Mengqi
    Hao, Yuewen
    Zheng, Minwen
    JOURNAL OF CARDIOTHORACIC SURGERY, 2013, 8
  • [44] Solid-state photon-counting hybrid detector array for high-resolution multi-energy X-ray imaging
    Sia, R.
    Kleinfelder, S.
    Nagarkar, V. V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) : 470 - 473
  • [45] Quantitative Analysis of Motion Artifacts in High-pitch Dual-source Computed Tomography of the Thorax
    Schulz, Boris
    Jacobi, Volkmar
    Beeres, Martin
    Bodelle, Boris
    Gruber, Tatjana
    Lee, Clara
    Bauer, Ralf
    Kerl, Matthias
    Vogl, Thomas
    Zangos, Stefan
    JOURNAL OF THORACIC IMAGING, 2012, 27 (06) : 382 - 386
  • [46] Extracellular Volume Quantification With Cardiac Late Enhancement Scanning Using Dual-Source Photon-Counting Detector CT
    Mergen, Victor
    Sartoretti, Thomas
    Klotz, Ernst
    Schmidt, Bernhard
    Jungblut, Lisa
    Higashigaito, Kai
    Manka, Robert
    Euler, Andre
    Kasel, Markus
    Eberhard, Matthias
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2022, 57 (06) : 406 - 411
  • [47] Spectral performance for iodine quantification of a dual-source, dual-kV photon counting detector CT
    Winfree, Tim
    Treb, Kevin
    Mccollough, Cynthia
    Leng, Shuai
    MEDICAL PHYSICS, 2025,
  • [48] Characterization of single- and multi-energy CT performance of an oral dark borosilicate contrast media using a clinical photon-counting-detector CT platform
    Ren, Liqiang
    Sun, Yuxin
    Yeh, Benjamin
    Marsh, Jeffrey F.
    Winfree, Timothy N.
    Burke, Kristin A.
    Rajendran, Kishore
    Mccollough, Cynthia H.
    Mileto, Achille
    Fletcher, Joel G.
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (11) : 6779 - 6788
  • [49] Radiation doses from phantom measurements at high-pitch dual-source computed tomography coronary angiography
    Goetti, Robert
    Leschka, Sebastian
    Boschung, Markus
    Mayer, Sabine
    Wyss, Christophe
    Stolzmann, Paul
    Frauenfelder, Thomas
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (04) : 773 - 779
  • [50] Technical note: Exploring the detectability of coronary calcification using ultra-high-resolution photon-counting-detector CT
    Chang, Shaojie
    Ren, Liqiang
    Tang, Shanshan
    Marsh Jr, Jeffrey F.
    Hsieh, Scott
    Mccollough, Cynthia H.
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (11) : 6836 - 6843