High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:14
|
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
收藏
页码:1428 / 1435
页数:8
相关论文
共 50 条
  • [31] Multi-Energy CT with Triple X-ray Beams and Photon-Counting-Detector CT for Simultaneous Imaging of Two Contrast Agents: An Experimental Comparison
    Ren, Liqiang
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [32] Improved Pulmonary Artery Evaluation Using High-Pitch Photon-Counting CT Compared to High-Pitch Conventional or Routine-Pitch Conventional Dual-Energy CT
    Yalon, Mariana
    Hoodeshenas, Safa
    Chan, Alex
    Horst, Kelly K.
    Crum, Isaac
    Thorne, Jamison E.
    Lee, Yong S.
    Yu, Lifeng
    Mccollough, Cynthia H.
    Fletcher, Joel G.
    Rajiah, Prabhakar Shantha
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2024, 48 (06) : 897 - 905
  • [33] Full Field Image Reconstruction Is Suitable for High-Pitch Dual-Source Computed Tomography
    Mahnken, Andreas H.
    Allmendinger, Thomas
    Sedlmair, Martin
    Tamm, Miriam
    Reinartz, Sebastian D.
    Flohr, Thomas
    INVESTIGATIVE RADIOLOGY, 2012, 47 (11) : 642 - 648
  • [34] Radiation dose and image quality of high-pitch emergency abdominal CT in obese patients using third-generation dual-source CT (DSCT)
    Forbrig, Robert
    Ingrisch, Michael
    Stahl, Robert
    Winter, Katharina Stella
    Reiser, Maximilian
    Trumm, Christoph G.
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [35] Ultra-high-resolution imaging of the shoulder and pelvis using photon-counting-detector CT: a feasibility study in patients
    Baffour, Francis, I
    Rajendran, Kishore
    Glazebrook, Katrina N.
    Thorne, Jamison E.
    Larson, Nicholas B.
    Leng, Shuai
    McCollough, Cynthia H.
    Fletcher, Joel G.
    EUROPEAN RADIOLOGY, 2022, 32 (10) : 7079 - 7086
  • [36] Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience
    Morsbach, Fabian
    Gordic, Sonja
    Desbiolles, Lotus
    Husarik, Daniela
    Frauenfelder, Thomas
    Schmidt, Bernhard
    Allmendinger, Thomas
    Wildermuth, Simon
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2014, 24 (08) : 1889 - 1895
  • [37] High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population
    Sabel, Bastian O.
    Buric, Kristijan
    Karara, Nora
    Thierfelder, Kolja M.
    Dinkel, Julien
    Sommer, Wieland H.
    Meinel, Felix G.
    PLOS ONE, 2016, 11 (02):
  • [38] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Florian Wolf
    Sebastian Leschka
    Christian Loewe
    Peter Homolka
    Christina Plank
    Ruediger Schernthaner
    Dominik Bercaczy
    Robert Goetti
    Johannes Lammer
    Guy Friedrich
    Borut Marincek
    Hatem Alkadhi
    Gudrun Feuchtner
    European Radiology, 2010, 20 : 2084 - 2091
  • [39] Consistency of spectral results in cardiac dual-source photon-counting CT
    Liu, Leening P.
    Shapira, Nadav
    Sahbaee, Pooyan
    Gang, Grace J.
    Knollman, Friedrich D.
    Chen, Marcus Y.
    Litt, Harold I.
    Noel, Peter B.
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [40] Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience
    Leschka, Sebastian
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Schertler, Thomas
    Scheffel, Hans
    Plass, Andre
    Falk, Volkmar
    Feuchtner, Gudrun
    Marincek, Borut
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2009, 19 (12) : 2896 - 2903