High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:14
|
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
收藏
页码:1428 / 1435
页数:8
相关论文
共 50 条
  • [21] Approaches, advantages, and challenges to photon counting detector and multi-energy CT
    Toia, Giuseppe V.
    Mileto, Achille
    Borhani, Amir A.
    Chen, Guang-Hong
    Ren, Liqiang
    Uyeda, Jennifer W.
    Marin, Daniele
    ABDOMINAL RADIOLOGY, 2024, 49 (09) : 3251 - 3260
  • [22] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Robert Goetti
    Gudrun Feuchtner
    Paul Stolzmann
    Lotus Desbiolles
    Michael Alexander Fischer
    Christoph Karlo
    Stephan Baumueller
    Hans Scheffel
    Hatem Alkadhi
    Sebastian Leschka
    European Radiology, 2010, 20 : 2565 - 2571
  • [23] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Goetti, Robert
    Feuchtner, Gudrun
    Stolzmann, Paul
    Desbiolles, Lotus
    Fischer, Michael Alexander
    Karlo, Christoph
    Baumueller, Stephan
    Scheffel, Hans
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2010, 20 (11) : 2565 - 2571
  • [24] High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model
    Donati, Olivio F.
    Burg, Matthias C.
    Desbiolles, Lotus
    Karlo, Christoph
    Stolzmann, Paul
    Bunck, Alexander
    Baumueller, Stephan
    Marincek, Borut
    Seifarth, Harald
    Alkadhi, Hatem
    Leschka, Sebastian
    Maintz, David
    ACADEMIC RADIOLOGY, 2010, 17 (11) : 1366 - 1374
  • [25] Temporal resolution and motion artifacts in single-source and dual-source cardiac CT
    Schoendube, Harald
    Allmendinger, Thomas
    Stierstorfer, Karl
    Bruder, Herbert
    Flohr, Thomas
    MEDICAL PHYSICS, 2013, 40 (03)
  • [26] High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate
    Gordic, Sonja
    Husarik, Daniela B.
    Desbiolles, Lotus
    Leschka, Sebastian
    Frauenfelder, Thomas
    Alkadhi, Hatem
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 (06) : 1173 - 1179
  • [27] High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate
    Sonja Gordic
    Daniela B. Husarik
    Lotus Desbiolles
    Sebastian Leschka
    Thomas Frauenfelder
    Hatem Alkadhi
    The International Journal of Cardiovascular Imaging, 2014, 30 : 1173 - 1179
  • [28] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Wolf, Florian
    Leschka, Sebastian
    Loewe, Christian
    Homolka, Peter
    Plank, Christina
    Schernthaner, Ruediger
    Bercaczy, Dominik
    Goetti, Robert
    Lammer, Johannes
    Friedrich, Guy
    Marincek, Borut
    Alkadhi, Hatem
    Feuchtner, Gudrun
    EUROPEAN RADIOLOGY, 2010, 20 (09) : 2084 - 2091
  • [29] Quantitative assessment of motion effects in dual-source dual-energy CT and dual-source photon-counting detector CT
    Ahmed, Zaki
    Rajendran, Kishore
    Gong, Hao
    McCollough, Cynthia
    Leng, Shuai
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [30] Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance
    Tao, Ashley
    Huang, Richard
    Tao, Shengzhen
    Michalak, Gregory J.
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11)