High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:17
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
收藏
页码:1428 / 1435
页数:8
相关论文
共 23 条
[1]   Influence of slice thickness and reconstruction kernel on the computed tomographic attenuation of coronary atherosclerotic plaque [J].
Achenbach, Stephan ;
Boehmer, Kerstin ;
Pflederer, Tobias ;
Ropers, Dieter ;
Seltmann, Martin ;
Lell, Michael ;
Anders, Katharina ;
Kuettner, Axel ;
Uder, Michael ;
Daniel, Werner G. ;
Marwan, Mohamed .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2010, 4 (02) :110-115
[2]   A Long-Term Prognostic Value of Coronary CT Angiography in Suspected Coronary Artery Disease [J].
Andreini, Daniele ;
Pontone, Gianluca ;
Mushtaq, Saima ;
Bartorelli, Antonio L. ;
Bertella, Erika ;
Antonioli, Laura ;
Formenti, Alberto ;
Cortinovis, Sarah ;
Veglia, Fabrizio ;
Annoni, Andrea ;
Agostoni, Piergiuseppe ;
Montorsi, Piero ;
Ballerini, Giovanni ;
Fiorentini, Cesare ;
Pepi, Mauro .
JACC-CARDIOVASCULAR IMAGING, 2012, 5 (07) :690-701
[3]   Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy [J].
Benz, Dominik C. ;
Benetos, Georgios ;
Rampidis, Georgios ;
von Felten, Elia ;
Bakula, Adam ;
Sustar, Aleksandra ;
Kudura, Ken ;
Messerli, Michael ;
Fuchs, Tobias A. ;
Gebhard, Catherine ;
Pazhenkottil, Aju P. ;
Kaufmann, Philipp A. ;
Buechel, Ronny R. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (05) :444-451
[4]   Incremental Value of Pharmacological Stress Cardiac Dual-Energy CT Over Coronary CT Angiography Alone for the Assessment of Coronary Artery Disease in a High-Risk Population [J].
De Cecco, Carlo Nicola ;
Harris, Brett S. ;
Schoepf, U. Joseph ;
Silverman, Justin R. ;
McWhite, Cullen B. ;
Krazinski, Aleksander W. ;
Bayer, Richard R. ;
Meinel, Felix G. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (01) :W70-W77
[5]   Image quality and radiation dose of a prospectively electrocardiography-triggered high-pitch data acquisition strategy for coronary CT angiography: The multicenter, randomized PROTECTION IV study [J].
Deseive, Simon ;
Pugliese, Francesca ;
Meave, Aloha ;
Alexanderson, Erick ;
Martinoff, Stefan ;
Hadamitzky, Martin ;
Massberg, Steffen ;
Hausleiter, Joerg .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2015, 9 (04) :278-285
[6]   High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose [J].
Euler, Andre ;
Higashigaito, Kai ;
Mergen, Victor ;
Sartoretti, Thomas ;
Zanini, Bettina ;
Schmidt, Bernhard ;
Flohr, Thomas G. ;
Ulzheimer, Stefan ;
Eberhard, Matthias ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2022, 57 (02) :115-121
[7]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[8]   Photon-counting CT review [J].
Flohr, Thomas ;
Petersilka, Martin ;
Henning, Andre ;
Ulzheimer, Stefan ;
Ferda, Jiri ;
Schmidt, Bernhard .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 :126-136
[9]   Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: Image reconstruction and assessment of image quality [J].
Flohr, Thomas G. ;
Leng, Shuai ;
Yu, Lifeng ;
Allmendinger, Thomas ;
Bruder, Herbert ;
Petersilka, Martin ;
Eusemann, Christian D. ;
Stierstorfer, Karl ;
Schmidt, Bernhard ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2009, 36 (12) :5641-5653
[10]   Dual-energy CT and coronary imaging [J].
Kay, Fernando Uliana .
CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2020, 10 (04) :1090-1107