A Highly Versatile X-ray and Electron Beam Diamond Dosimeter for Radiation Therapy and Protection

被引:18
|
作者
Pettinato, Sara [1 ,2 ]
Girolami, Marco [2 ]
Stravato, Antonella [3 ]
Serpente, Valerio [2 ]
Musio, Daniela [3 ]
Rossi, Maria C. [4 ]
Trucchi, Daniele M. [2 ]
Olivieri, Riccardo [3 ]
Salvatori, Stefano [1 ,2 ]
机构
[1] Niccolo Cusano Univ, Engn Fac, Via Don Carlo Gnocchi 3, I-00166 Rome, Italy
[2] Consiglio Nazl Ric ISM CNR, Ist Struttura Mat, DiaTHEMA Lab, Str Prov 35D 9, I-00010 Rome, Italy
[3] Azienda Osped San Giovanni Addolorata, Via Amba Aradam 8, I-00184 Rome, Italy
[4] Univ Roma Tre, Dept Ind Elect & Mech Engn, Via Vito Volterra 62, I-00146 Rome, Italy
关键词
radiation therapy; radiation protection; dosimetry; CVD diamond detectors; electron beams; X-rays; SINGLE-CRYSTAL; DETECTORS; PHOTON;
D O I
10.3390/ma16020824
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Radiotherapy is now recognized as a pillar in the fight against cancer. Two different types are currently used in clinical practice: (1) external beam radiotherapy, using high-energy X-rays or electron beams, both in the MeV-range, and (2) intraoperative radiotherapy, using low-energy X-rays (up to 50 keV) and MeV-range electron beams. Versatile detectors able to measure the radiation dose independently from the radiation nature and energy are therefore extremely appealing to medical physicists. In this work, a dosimeter based on a high-quality single-crystal synthetic diamond sample was designed, fabricated and characterized under low-energy X-rays, as well as under high-energy pulsed X-rays and electron beams, demonstrating excellent linearity with radiation dose and dose-rate. Detector sensitivity was measured to be 0.299 +/- 0.002 mu C/Gy under 6 MeV X-ray photons, and 0.298 +/- 0.004 mu C/Gy under 6 MeV electrons, highlighting that the response of the diamond dosimeter is independent of the radiation nature. Moreover, in the case of low-energy X-rays, an extremely low limit of detection (23 nGy/s) was evaluated, pointing out the suitability of the device to radiation protection dosimetry.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] X-ray acoustic imaging for external beam radiation therapy dosimetry using a commercial ultrasound scanner
    Sampaio, Diego R. T.
    Uliana, Joao H.
    Carneiro, Antonio A. O.
    Pavoni, Juliana F.
    Pavan, Theo Z.
    Borges, Leandro F.
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [22] Optimization of X-ray microplanar beam radiation therapy for deep-seated tumors by a simulation study
    Shinohara, Kunio
    Kondoh, Takeshi
    Nariyama, Nobuteru
    Fujita, Hajime
    Washio, Masakazu
    Aoki, Yukimasa
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (03) : 395 - 406
  • [23] Delivery confirmation of bolus electron conformal therapy combined with intensity modulated x-ray therapy
    Kavanaugh, James A.
    Hogstrom, Kenneth R.
    Chu, Connel
    Carver, Robert A.
    Fontenot, Jonas P.
    Henkelmann, Gregory
    MEDICAL PHYSICS, 2013, 40 (02)
  • [24] X-ray radiation protection aspects during ultrashort laser processing
    Legall, Herbert
    Schwanke, Christoph
    Bonse, Joern
    Krueger, Joerg
    JOURNAL OF LASER APPLICATIONS, 2020, 32 (02)
  • [25] A versatile diamond anvil cell for X-ray inelastic, diffraction and imaging studies at synchrotron facilities
    Petitgirard, Sylvain
    Jacobs, Jeroen
    Cerantola, Valerio
    Collings, Ines E.
    Tucoulou, Remi
    Dubrovinsky, Leonid
    Sahle, Christoph J.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2019, 90 (09)
  • [26] X-ray radiation generated by a beam of relativistic electrons in composite structure
    Blazhevich, S. V.
    Noskov, A. V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2018, 421 : 18 - 26
  • [27] Experimental optimisation of the X-ray energy in microbeam radiation therapy
    Livingstone, Jayde
    Stevenson, Andrew W.
    Hausermann, Daniel
    Adam, Jean-Francois
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 45 : 156 - 161
  • [28] Effect of gamma-ray, electron-beam, and X-ray irradiation on antioxidant activity of mugwort extracts
    Hwang, Ko-Eun
    Ham, Youn-Kyung
    Song, Dong-Heon
    Kim, Hyun-Wook
    Lee, Mi-Ai
    Jeong, Jong-Yeon
    Choi, Yun-Sang
    RADIATION PHYSICS AND CHEMISTRY, 2021, 186
  • [29] A TECHNIQUE FOR CALIBRATING AN ELECTRON-BEAM EVAPORATOR X-RAY SOURCE
    KINZIG, RJ
    JOURNAL OF ELECTRONIC MATERIALS, 1991, 20 (09) : 681 - 686
  • [30] Image intensifier X-ray beam collimation and its effect on radiation dose during ureteroscopy
    Horsburgh, Ben A.
    Babajews, Wesley
    Altham, Samantha
    Cowan, Helen
    JOURNAL OF CLINICAL UROLOGY, 2016, 9 (05) : 335 - 339