Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer

被引:5
|
作者
Larcher, Romaric [1 ,2 ]
Lottelier, Maxence [3 ]
Badiou, Stephanie [1 ]
Dupuy, Anne-Marie [3 ]
Bargnoux, Anne-Sophie [1 ]
Cristol, Jean-Paul [1 ]
机构
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
关键词
analytical performance; Point-of-Care Analyzer; i-STAT Alinity; blood gas; lactate; chemistry; urea; creatinine; hematology; coagulation; BLOOD; COAGULATION; AGREEMENT; THERAPY;
D O I
10.3390/diagnostics13020297
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Analytical and clinical validation of a new point-of-care testing system for determination of D-Dimer in human blood
    Sidelmann, Johannes J.
    Gram, Jorgen
    Larsen, Anette
    Overgaard, Kathrine
    Jespersen, Jorgen
    THROMBOSIS RESEARCH, 2010, 126 (06) : 524 - 530
  • [42] Multicentre Analytical Evaluation of a New Point-Of-Care System for the Determination of Cardiac and Thromboembolic Markers
    Bertsch, Thomas
    Chapelle, Jean-Paul
    Dempfle, Carl-Erik
    Giannitsis, Evangelos
    Schwab, Michael
    Zerback, Rainer
    CLINICAL LABORATORY, 2010, 56 (1-2) : 37 - 49
  • [43] Analytical evaluation of the epoc® point-of-care blood analysis system in cardiopulmonary bypass patients
    Chen, Jianing
    Gorman, Monique
    O'Reilly, Bill
    Chen, Yu
    CLINICAL BIOCHEMISTRY, 2016, 49 (09) : 708 - 712
  • [44] Measuring the Level of Agreement for Lactate Measurements in Hispaniolan Amazon Parrots (Amazona ventralis) Among 2 Point-of-Care Analyzers and a Benchtop Analyzer
    Mason, Alexandra K.
    Padlo, Jordan
    Mitchell, Mark A.
    Heatley, J. Jill
    Tully, Thomas N., Jr.
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2023, 37 (04) : 314 - 320
  • [45] Performance evaluation of SD A1cCare as a HbA1c analyzer for point-of-care testing
    Lee, Kyunghoon
    Jun, Sun-Hee
    Han, Minje
    Song, Sang Hoon
    Park, Kyoung Un
    Song, Woon Heung
    Song, Junghan
    CLINICAL BIOCHEMISTRY, 2015, 48 (09) : 625 - 627
  • [46] Method Comparison of β-Hydroxybutyrate Using a Point-of-Care Device and Dry Chemistry Analyzer in Three Sea Turtle Species
    Soehnlein, Cassidy
    Hirsch, Sarah E.
    Myre, Brianna L.
    Stacy, Nicole, I
    Clark, Samantha
    Cray, Carolyn
    Aoki, Derek
    Kaleel, Kristina
    Klingshirn, Skyler
    Toonder, Madison
    Perrault, Justin R.
    JOURNAL OF WILDLIFE DISEASES, 2022, 58 (03) : 670 - 674
  • [47] Analytical Performance Evaluation and Comparability of Patient Results Within a Point-of-Care Blood Gas Network
    Oliver, Paloma
    Fernandez-Calle, Pilar
    Rico, Natividad
    Jose Alcaide, Maria
    Gomez-Rioja, Ruben
    Buno, Antonio
    Manuel Iturzaeta, Jose
    POINT OF CARE, 2013, 12 (03): : 144 - 149
  • [48] Inosine and hypoxanthine as novel biomarkers for cardiac ischemia: From bench to point-of-care
    Farthing, Don E.
    Farthing, Christine A.
    Xi, Lei
    EXPERIMENTAL BIOLOGY AND MEDICINE, 2015, 240 (06) : 821 - 831
  • [49] Novel Point-of-Care Diagnostic Method for Neonatal Encephalopathy Using Purine Nucleosides
    Beamer, Edward
    O'Dea, Mary Isabel
    Garvey, Aisling A.
    Smith, Jonathon
    Menendez-Mendez, Aida
    Kelly, Lynne
    Pavel, Andreea
    Quinlan, Sean
    Alves, Mariana
    Jimenez-Mateos, Eva M.
    Tian, Faming
    Dempsey, Eugene
    Dale, Nicholas
    Murray, Deirdre M.
    Boylan, Geraldine B.
    Molloy, Eleanor J.
    Engel, Tobias
    FRONTIERS IN MOLECULAR NEUROSCIENCE, 2021, 14
  • [50] Clinically relevant analytical techniques, organizational concepts for application and future perspectives of point-of-care testing
    Luppa, Peter B.
    Bietenbeck, Andreas
    Beaudoin, Christopher
    Giannetti, Ambra