Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer

被引:5
|
作者
Larcher, Romaric [1 ,2 ]
Lottelier, Maxence [3 ]
Badiou, Stephanie [1 ]
Dupuy, Anne-Marie [3 ]
Bargnoux, Anne-Sophie [1 ]
Cristol, Jean-Paul [1 ]
机构
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
关键词
analytical performance; Point-of-Care Analyzer; i-STAT Alinity; blood gas; lactate; chemistry; urea; creatinine; hematology; coagulation; BLOOD; COAGULATION; AGREEMENT; THERAPY;
D O I
10.3390/diagnostics13020297
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Evaluation of PIMA Point-of-care CD4 Analyzer in Yunnan, China
    Liang, Jun
    Duan, Song
    Ma, Yan-Ling
    Wang, Ji-Bao
    Su, Ying-Zhen
    Zhang, Hui
    Ou, Chin-Yih
    Hao, Ling
    Qi, Ming-Shan
    Bulterys, Marc
    Westerman, Larry
    Jiang, Yan
    Xiao, Yao
    CHINESE MEDICAL JOURNAL, 2015, 128 (07) : 890 - 895
  • [32] Comparison of a Point-of-care Cholesterol Meter With a Reference Laboratory Analyzer in Companion Psittaciformes
    Barboza, Trinita
    Beaufrere, Hugues
    JOURNAL OF AVIAN MEDICINE AND SURGERY, 2019, 33 (01) : 7 - 14
  • [33] Analytical performance and user-friendliness of five novel point-of-care D-dimer assays
    Heerink, Jorn
    Gemen, Eugenie
    Oudega, Ruud
    Hopstaken, Rogier
    Geersing, Geert-Jan
    Kusters, Ron
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2020, 80 (05) : 433 - 440
  • [34] Plasma lactate concentrations and comparison of two point-of-care lactate analyzers to a laboratory analyzer in a population of healthy cats
    Tynan, Beth
    Kerl, Marie E.
    Jackson, Mary L.
    Mann, F. A.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2015, 25 (04) : 521 - 527
  • [35] PORTABLE SIMULTANEOUS MULTIPLE ANALYTE WHOLE-BLOOD ANALYZER FOR POINT-OF-CARE TESTING
    SCHEMBRI, CT
    OSTOICH, V
    LINGANE, PJ
    BURD, TL
    BUHL, SN
    CLINICAL CHEMISTRY, 1992, 38 (09) : 1665 - 1670
  • [36] A Point-of-Care Prothrombin Time Test on a Microfluidic Disk Analyzer Using Alternate Spinning
    Lin, Chia-Hui
    Lin, Kun-Wei
    Yen, Daniel
    Shih, Chih-Hsin
    Lu, Chien-Hsing
    Wang, Jiunn-Min
    Lin, Chi-Yu
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2015, 15 (02) : 1401 - 1407
  • [37] Agreement between hematocrits measured with electrical conductivity on the i-STAT analyzer and manual packed cell volume measurements in 148 chelonian patients
    Wong, Shanna
    Di Girolamo, Nicola
    VETERINARY CLINICAL PATHOLOGY, 2021, 50 (03) : 342 - 347
  • [38] Impact of regression modeling on the assessment and harmonization of a point-of-care analyzer and commercial laboratory analyzer for feline plasma biochemistry testing
    Baral, Randolph M.
    Flatland, Bente
    Jaensch, Susan M.
    Hayward, Douglas A.
    Freeman, Kathleen P.
    VETERINARY CLINICAL PATHOLOGY, 2024, 53 (03) : 325 - 342
  • [39] Comprehensive Coagulation Profiling at the Point-of-Care Using a Novel Laser-Based Approach
    Nadkarni, Seemantini K.
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2019, 45 (03) : 264 - 274
  • [40] Measuring circulating triglycerides in free-living birds: evaluation of a field-usable point-of-care analyzer for American woodcock
    Roelandt, Christopher M.
    Roth, Amber M.
    McWilliams, Scott R.
    Witt, Jill C.
    JOURNAL OF ORNITHOLOGY, 2022, 163 (01) : 235 - 241