Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer

被引:5
|
作者
Larcher, Romaric [1 ,2 ]
Lottelier, Maxence [3 ]
Badiou, Stephanie [1 ]
Dupuy, Anne-Marie [3 ]
Bargnoux, Anne-Sophie [1 ]
Cristol, Jean-Paul [1 ]
机构
[1] Univ Montpellier, Montpellier Univ Hosp, INSERM French Inst Hlth & Med Res, CNRS French Natl Ctr Sci Res,PhyMedExp Physiol & E, F-34000 Montpellier, France
[2] Nimes Univ Hosp, Infect & Trop Dis Dept, F-30000 Nimes, France
[3] Montpellier Univ Hosp, Biochem & Hormonol Dept, F-34000 Montpellier, France
关键词
analytical performance; Point-of-Care Analyzer; i-STAT Alinity; blood gas; lactate; chemistry; urea; creatinine; hematology; coagulation; BLOOD; COAGULATION; AGREEMENT; THERAPY;
D O I
10.3390/diagnostics13020297
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Many Point-of-Care devices have been released over the past decade. However, data regarding their analytical performances in real-world situations remains scarce. Herein, we aimed to assess the analytical performances of the i-STAT Alinity system. We conducted an analytical performances study with the i-STAT Alinity device using cartridges CG4+ (pH, Pco(2), Po-2, lactate, bicarbonate and base excess); CHEM8+ (Na, K, Cl, ionized Ca, urea, creatinine, glucose, hematocrit and hemoglobin) and PT/INR (prothrombin time and international normalized ratio). We assessed the imprecision and compared the results to those obtained on existing instruments in the central laboratory. We found that the within-lab coefficients of variation (CV) were very low (<2%) or low (2-5%), except for creatinine and PT (CV = 5.2% and CV = 6.3%, respectively). For almost all the parameters, the results were strongly (R-2 = 90-95%) or very strongly (R-2 > 95%) correlated with those of the existing laboratory instruments, and the biases were very low (<2%) or low (2-5%). However, correlations of the PT and INR measurements with existing instruments were lower (R-2 = 86.0% and 89.7%), and biases in the Po-2 (7.9%), creatinine (5.4%) and PT (-6.6%) measurements were higher. The i-STAT Alinity appeared as a convenient device for measurements of numerous parameters. However, clinicians should interpret Po-2, creatinine and PT results with caution.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Evaluation of the i-STAT Alinity Point-of-Care Analyzer
    Lee, Mee-Yin
    Lim, Sian-Foong
    Lam, Leslie
    POINT OF CARE, 2019, 18 (02): : 48 - 55
  • [2] Analytical performances of the i-STAT portable clinical analyzer for the measurement of PO2 and PCO2
    Sachs, C
    Finetti, P
    Rabouine, P
    Abdoulaye, F
    ANNALES DE BIOLOGIE CLINIQUE, 2002, 60 (04) : 411 - 420
  • [3] The rise and fall of i-STAT point-of-care blood gas testing in an acute care hospital
    Ng, VL
    Kraemer, R
    Hogan, C
    Eckman, D
    Siobal, M
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 114 (01) : 128 - 138
  • [4] Evaluation of the i-STAT portable point-of-care analyzer for determination of blood gases and acid-base status in newborn calves
    Bleul, Ulrich
    Goetz, Eva
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2014, 24 (05) : 519 - 528
  • [5] Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT® point-of-care analyser
    Veldhoen, Esther S.
    de Vooght, Karen M. K.
    Slieker, Martijn G.
    Versluys, Anne B.
    Turner, Nigel Mcb.
    RESUSCITATION, 2014, 85 (03) : 359 - 363
  • [6] Evaluation of the i-STAT Alinity v in a veterinary clinical setting
    Burke, Jasper E.
    Nguyen, Thuy Hien T.
    Davis, Tanya
    Koenig, Amie
    Lane, Selena L.
    Good, Jennifer
    Brainard, Benjamin M.
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2021, 33 (04) : 703 - 710
  • [7] Evaluation of the i-STAT Portable Clinical Analyzer in dogs
    Matburger, C
    Henke, J
    Hirschberger, J
    Matburger, S
    Erhardt, W
    TIERARZTLICHE PRAXIS AUSGABE KLEINTIERE HEIMTIERE, 2000, 28 (02): : 132 - 137
  • [8] Comparison of Hematocrit and Biochemical Analytes among Two Point-of-Care Analyzers (EPOC and i-STAT Alinity v) and a Veterinary Diagnostic Laboratory in the African Savanna Elephant (Loxodonta africana) and the Southern White Rhinoceros (Ceratotherium simum simum)
    Stratton, Hayley S.
    Ange-van Heugten, Kimberly D.
    Minter, Larry J.
    JOURNAL OF ZOOLOGICAL AND BOTANICAL GARDENS, 2022, 3 (04): : 653 - 664
  • [9] Evaluation of i-STAT portable clinical analyzer in a hemodialysis unit
    Gault, MH
    Harding, CE
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 117 - 124
  • [10] Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: agreement, concordance, and clinical reliability
    Dirkmann, Daniel
    Nagy, Elisabeth
    Britten, Martin W.
    Peters, Juergen
    BMC ANESTHESIOLOGY, 2019, 19 (01)