Characterisation of HEXITECMHz- A 1 MHz continuous frame rate spectroscopic X-ray imaging detector system

被引:3
|
作者
Cline, B. D. [1 ]
Banks, D. [1 ]
Bell, S. [1 ]
Church, I. [1 ]
Cross, S. [1 ]
Davis, A. [1 ]
Day, C. [1 ]
French, M. [1 ]
Gardiner, T. [1 ]
Ghorbanian, N. [1 ]
Harris, J. [1 ]
Hart, M. [1 ]
Holden, J. [1 ]
Jones, L. [1 ]
Lipp, J. [1 ]
Nicholls, T. [1 ]
Nobes, J. [1 ]
Pradeep, S. [1 ]
Prydderch, M. [1 ]
Roberts, M. [1 ]
Schneider, A. [1 ]
Seller, P. [1 ]
Sole, D. [1 ]
Veale, M. C. [1 ]
Wilson, M. D. [1 ]
Helsby, W. [2 ]
Dhamgaye, V. [3 ]
Fox, O. [3 ]
Sawhney, K. [3 ]
机构
[1] UKRI STFC Rutherford Appleton Lab, Didcot OX11 0QX, Oxon, England
[2] UKRI STFC Daresbury Lab, Daresbury WA4 4AD, Warrington, England
[3] Diamond Light Source, Didcot OX11 0DE, Oxon, England
关键词
Pixel detector; MHz imaging; P-type Si; Resolution; X-ray spectroscopy; X-ray imaging; PERFORMANCE; ENERGY;
D O I
10.1016/j.nima.2023.168718
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Spectroscopic X-ray imaging techniques including Compton X-ray Imaging, X-ray Fluorescence Imaging and Hyperspectral X-ray Tomography require energy-resolving detectors capable of operating at high incident X-ray fluxes to make time resolved measurements. HEXITECMHz, operates at a continuous 1 MHz frame rate and can make fully spectroscopic measurements at >10(6) ph s(-1) mm(-2). This is enabled by an integrating Front End, in-pixel digitisation and high-speed serialisers. A 300 mu m thick p-type Si HEXITECMHz detector was characterised on the B16 Test Beamline at the Diamond Light Source and are the first measurements taken at a 1 MHz frame rate. At 10 keV and 15 keV) the device displayed average FWHM of 656 eV and 682 eV respectively, with minimal changes in spectroscopic performance over similar to 8 h. Analysis of charge-sharing events show low charge loss and a linear energy-signal response. Higher-flux measurements illustrated the capability of the ASIC to operate as a photon-counting device.
引用
收藏
页数:12
相关论文
共 21 条
  • [21] Stroke Evaluation in the Interventional Suite Using Dual-Layer Detector Cone-Beam CT: a First-in-human Prospective Cohort Study (the Next Generation X-ray Imaging System Trial)
    Stahl, Fredrik
    Almqvist, Hakan
    Aspelin, Asa
    Kolloch, Jens
    Ghalamkari, Odett
    Gontu, Vamsi
    Schaefer, Dirk
    van de Haar, Peter
    Engel, Klaus-Juergen
    van Nijnatten, Fred
    Holmberg, Ake
    Mazya, Michael V.
    Soderman, Michael
    Delgado, Anna Falk
    CLINICAL NEURORADIOLOGY, 2024, 34 (04) : 929 - 937