Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study

被引:8
|
作者
Romero, Marta [1 ]
Herrero-Climent, Mariano [1 ,2 ]
Rios-Carrasco, Blanca [1 ]
Brizuela, Aritza [3 ]
Romero, Manuel Maria [1 ]
Gil, Javier [4 ]
机构
[1] Univ Seville, Sch Dent, Dept Periodontol, C-Avicena S-N, Seville 41009, Spain
[2] Porto Dent Inst, Periodontol Dept, Symmetrya Prothesis, Ave Montevideu 810, P-4150518 Porto, Portugal
[3] Univ Europea Miguel Cervantes, Fac Ciencias Salud, Densia Reserach Grp, C Padre Julio Chevalier 2, Valladolid 47012, Spain
[4] Univ Int Catalunya, Bioengn Inst Technol, Fac Med & Hlth Sci, C Josep Trueta S-N, Sant Cugat Del Valles 08195, Barcelona, Spain
关键词
dental implant; titanium; primary stability; insertion torque; implant stability quotient; RESONANCE FREQUENCY-ANALYSIS; CRESTAL BONE CHANGES; SURFACE-ROUGHNESS; SURGICAL TECHNIQUE; TITANIUM IMPLANTS; FATIGUE; OSSEOINTEGRATION; WETTABILITY; TOPOGRAPHY; GEOMETRY;
D O I
10.3390/jcm12134190
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In the placement of dental implants, the primary fixation between the dental implant and the bone is of great importance and corresponds to compressive mechanical fixation that aims to prevent micromovement of the implant. The aim of this research was to determine the role of roughness and the type of dental implant (tissue-level or bone-level) in implant stability, measured using resonance frequency analysis (RFA) and insertion torque (IT). We analyzed 234 titanium dental implants, placed in fresh calf ribs, at the half-tissue level and half-bone level. The implant surface was subjected to grit-blasting treatments with alumina particles of 120, 300, and 600 & mu;m at a projection pressure of 2.5 bar, resulting in three types of roughness. Roughness was determined via optical interferometry. The wettability of the surfaces was also determined. Implant stability was measured using a high-precision torquemeter to obtain IT, and RFA was used to determine the implant stability quotient (ISQ). The results show that rough surfaces with Sa values of 0.5 to 4 & mu;m do not affect the primary stability. However, the type of implant is important; bone-level implants obtained the highest primary stability values. A good correlation between the primary stability values obtained via IT and ISQ was demonstrated. New in vivo studies are necessary to know whether these results can be maintained in the long term.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Effect of Implant Macro-Design and Magnetodynamic Surgical Preparation on Primary Implant Stability: An In Vitro Investigation
    Antonelli, Alessandro
    Barone, Selene
    Attanasio, Ferdinando
    Salviati, Marianna
    Cerra, Maria Giulia
    Calabria, Elena
    Bennardo, Francesco
    Giudice, Amerigo
    DENTISTRY JOURNAL, 2023, 11 (10)
  • [32] Effect of photobiomodulation therapy of overprepared dental implant bed on torque removal and implant stability quotient: an experimental study in sheep
    Xailani, Kocher K.
    Hamad, Shehab A.
    MINERVA DENTAL AND ORAL SCIENCE, 2024, 73 (02) : 96 - 101
  • [33] Influence of implant length and insertion depth on primary stability of short dental implants: An in vitro study of a novel mandibular artificial bone model
    Wu, Hsiang-Chun
    Huang, Heng-Li
    Fuh, Lih-Jyh
    Tsai, Ming-Tzu
    Hsu, Jui-Ting
    JOURNAL OF DENTAL SCIENCES, 2024, 19 (01) : 139 - 147
  • [34] Relation between initial implant stability quotient and bone-implant contact percentage: an in vitro model study
    Huang, Heng-Li
    Tsai, Ming-Tzu
    Su, Kuo-Chih
    Li, Yu-Fen
    Hsu, Jui-Ting
    Chang, Chih-Han
    Fuh, Lih-Jyh
    Wu, Aaron Yu-Jen
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 116 (05): : E356 - E361
  • [35] Insertion Torque Function Analysis of Novel Dental Implant Geometry
    Pammer, D.
    Bognar, E.
    FIRST EUROPEAN BIOMEDICAL ENGINEERING CONFERENCE FOR YOUNG INVESTIGATORS, 2015, 50 : 83 - 86
  • [36] Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study
    Tabassum, Afsheen
    Meijer, Gert J.
    Wolke, Johannes G. C.
    Jansen, John A.
    CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (04) : 327 - 332
  • [37] Effect of bone quality, implant design, and surgical technique on primary implant stability
    Karl, Matthias
    Grobecker-Karl, Tanja
    QUINTESSENCE INTERNATIONAL, 2018, 49 (03): : 189 - 198
  • [38] The Effect of Insertion Angles and Depths of Dental Implant on the Initial Stability
    Wu, Hsiang-Chun
    Tsai, Ming-Tzu
    Hsu, Jui-Ting
    APPLIED SCIENCES-BASEL, 2020, 10 (09):
  • [39] Evaluation of Insertion Energy as Novel Parameter for Dental Implant Stability
    Grobecker-Karl, Tanja
    Dickinson, Anthony
    Heckmann, Siegfried
    Karl, Matthias
    Steiner, Constanze
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (09) : 1 - 7
  • [40] Implant Primary Stability Determined by Resonance Frequency Analysis: Correlation With Insertion Torque, Histologic Bone Volume, and Torsional Stability at 6 Weeks
    Makary, Christian
    Rebaudi, Alberto
    Sammartino, Gilberto
    Naaman, Nada
    IMPLANT DENTISTRY, 2012, 21 (06) : 474 - 480