Investigation of the Influence of Roughness and Dental Implant Design on Primary Stability via Analysis of Insertion Torque and Implant Stability Quotient: An In Vitro Study

被引:12
作者
Romero, Marta [1 ]
Herrero-Climent, Mariano [1 ,2 ]
Rios-Carrasco, Blanca [1 ]
Brizuela, Aritza [3 ]
Romero, Manuel Maria [1 ]
Gil, Javier [4 ]
机构
[1] Univ Seville, Sch Dent, Dept Periodontol, C-Avicena S-N, Seville 41009, Spain
[2] Porto Dent Inst, Periodontol Dept, Symmetrya Prothesis, Ave Montevideu 810, P-4150518 Porto, Portugal
[3] Univ Europea Miguel Cervantes, Fac Ciencias Salud, Densia Reserach Grp, C Padre Julio Chevalier 2, Valladolid 47012, Spain
[4] Univ Int Catalunya, Bioengn Inst Technol, Fac Med & Hlth Sci, C Josep Trueta S-N, Sant Cugat Del Valles 08195, Barcelona, Spain
关键词
dental implant; titanium; primary stability; insertion torque; implant stability quotient; RESONANCE FREQUENCY-ANALYSIS; CRESTAL BONE CHANGES; SURFACE-ROUGHNESS; SURGICAL TECHNIQUE; TITANIUM IMPLANTS; FATIGUE; OSSEOINTEGRATION; WETTABILITY; TOPOGRAPHY; GEOMETRY;
D O I
10.3390/jcm12134190
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In the placement of dental implants, the primary fixation between the dental implant and the bone is of great importance and corresponds to compressive mechanical fixation that aims to prevent micromovement of the implant. The aim of this research was to determine the role of roughness and the type of dental implant (tissue-level or bone-level) in implant stability, measured using resonance frequency analysis (RFA) and insertion torque (IT). We analyzed 234 titanium dental implants, placed in fresh calf ribs, at the half-tissue level and half-bone level. The implant surface was subjected to grit-blasting treatments with alumina particles of 120, 300, and 600 & mu;m at a projection pressure of 2.5 bar, resulting in three types of roughness. Roughness was determined via optical interferometry. The wettability of the surfaces was also determined. Implant stability was measured using a high-precision torquemeter to obtain IT, and RFA was used to determine the implant stability quotient (ISQ). The results show that rough surfaces with Sa values of 0.5 to 4 & mu;m do not affect the primary stability. However, the type of implant is important; bone-level implants obtained the highest primary stability values. A good correlation between the primary stability values obtained via IT and ISQ was demonstrated. New in vivo studies are necessary to know whether these results can be maintained in the long term.
引用
收藏
页数:14
相关论文
共 60 条
[1]   The effect of thread pattern upon implant osseointegration [J].
Abuhussein, Heba ;
Pagni, Giorgio ;
Rebaudi, Alberto ;
Wang, Hom-Lay .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (02) :129-136
[2]   Comparative histomorphometry and resonance frequency analysis of implants with moderately rough surfaces in a loaded animal model [J].
Al-Nawas, B. ;
Groetz, K. A. ;
Goetz, H. ;
Duschner, H. ;
Wagner, W. .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (01) :1-8
[3]   Turned, machined versus double-etched dental implants in vivo [J].
Al-Nawas, B. ;
Hangen, U. ;
Duschner, H. ;
Krummenauer, F. ;
Wagner, W. .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2007, 9 (02) :71-78
[4]  
Al-Sabbagh Mohanad, 2019, Dent Clin North Am, V63, P461, DOI 10.1016/j.cden.2019.02.001
[5]   Microthreaded Implants and Crestal Bone Loss: A Systematic Review [J].
Al-Thobity, Ahmad M. ;
Kutkut, Ahmad ;
Almas, Khalid .
JOURNAL OF ORAL IMPLANTOLOGY, 2017, 43 (02) :157-166
[6]   Advances in surfaces and osseointegration in implantology. Biomimetic surfaces [J].
Albertini, Matteo ;
Fernandez-Yague, Marc ;
Lazaro, Pedro ;
Herrero-Climent, Mariano ;
Rios-Santos, Jose-Vicente ;
Bullon, Pedro ;
Gil, Francisco-Javier .
MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2015, 20 (03) :E316-E325
[7]   Relevant Aspects of Titanium and Zirconia Dental Implants for Their Fatigue and Osseointegration Behaviors [J].
Aragoneses, Javier ;
Lopez Valverde, Nansi ;
Fernandez-Dominguez, Manuel ;
Mena-Alvarez, Jesus ;
Rodriguez, Cinthia ;
Gil, Javier ;
Manuel Aragoneses, Juan .
MATERIALS, 2022, 15 (11)
[8]   Peri-implant diseases and conditions: Consensus report of workgroup 4 of the 2017 World Workshop on the Classification of Periodontal and Peri-Implant Diseases and Conditions [J].
Berglundh, Tord ;
Armitage, Gary ;
Araujo, Mauricio G. ;
Avila-Ortiz, Gustavo ;
Blanco, Juan ;
Camargo, Paulo M. ;
Chen, Stephen ;
Cochran, David ;
Derks, Jan ;
Figuero, Elena ;
Hammerle, Christoph H. F. ;
Heitz-Mayfield, Lisa J. A. ;
Huynh-Ba, Guy ;
Iacono, Vincent ;
Koo, Ki-Tae ;
Lambert, France ;
McCauley, Laurie ;
Quirynen, Marc ;
Renvert, Stefan ;
Salvi, Giovanni E. ;
Schwarz, Frank ;
Tarnow, Dennis ;
Tomasi, Cristiano ;
Wang, Hom-Lay ;
Zitzmann, Nicola .
JOURNAL OF PERIODONTOLOGY, 2018, 89 :S313-S318
[9]   Influence of hex geometry and prosthetic table width on static and fatigue strength of dental implants [J].
Boggan, RS ;
Strong, JT ;
Misch, CE ;
Bidez, MW .
JOURNAL OF PROSTHETIC DENTISTRY, 1999, 82 (04) :436-440
[10]   Relationship Between Insertion Torque and Resonance Frequency Measurements, Performed by Resonance Frequency Analysis, in Micromobility of Dental Implants: An In Vitro Study [J].
Brizuela-Velasco, Aritza ;
Alvarez-Arenal, Angel ;
Javier Gil-Mur, Francisco ;
Herrero-Climent, Mariano ;
Chavarri-Prado, David ;
Chento-Valiente, Yelko ;
Dieguez-Pereira, Markel .
IMPLANT DENTISTRY, 2015, 24 (05) :607-611