Novel Radiofrequency Balloon Catheter-Impact of Ablation Parameters on Single-Shot Isolation

被引:4
|
作者
My, Ilaria [1 ,5 ]
Bordignon, Stefano [2 ]
Butt, Mahi [1 ]
Rottner, Laura [1 ]
Marc, Lemoine [1 ]
Moser, Fabian [1 ]
Wenzel, Jan -Per [1 ]
Obergassel, Julius [1 ]
Schleberger, Ruben [1 ]
Moser, Julia [1 ]
Dinshaw, Leon [1 ]
Kirchhof, Paulus [1 ,3 ]
Reissmann, Bruno [1 ]
Ouyang, Feifan [1 ]
Chun, Kyoung-Ryul Julian [2 ]
Schmidt, Boris [2 ,4 ]
Rillig, Andreas [1 ]
Metzner, Andreas [1 ]
机构
[1] Univ Heart & Vasc Ctr Hamburg, Dept Cardiol, Hamburg, Germany
[2] Cardioangiol Ctr Bethanien, Frankfurt, Germany
[3] Univ Birmingham, Inst Cardiovasc Sci, Birmingham, England
[4] Univ Klinikum Frankfurt, Med Klin 3, Kardiol, Frankfurt, Germany
[5] Univ Heart & Vasc Ctr Hamburg, Dept Cardiol, Martinistr 52, D-20246 Hamburg, Germany
关键词
Atrial fibrillation; Pulmonary vein isolation; Radiofrequency balloon; Single-shot isolation; PAROXYSMAL ATRIAL-FIBRILLATION; PULMONARY VEIN ISOLATION; CRYOBALLOON;
D O I
10.1253/circj.CJ-23-0220
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A novel multielectrode radiofrequency balloon (RFB) catheter has been released for pulmonary vein isolation (PVI). Methods and Results: In this observational study consecutive patients with drug-refractory paroxysmal or persistent atrial fibrillation (AF) undergoing first-time PVI were enrolled in 2 high-volume ablation centers. All procedures were conducted in conjunction with a 3D-mapping system. Clinical, procedural and ablation parameters were systematically analyzed. 105 patients (58% male; 52% paroxysmal AF, 68 & PLUSMN;11.3 years mean age, left atrial volume index 38.6 & PLUSMN;14.8 mL/m2) were included. 241/412 (58.5%) PVs were suc-cessfully isolated with a single shot (SS), with a time-to-isolation of 11.6 & PLUSMN;8 s. Total number of radiofrequency applications was 892 (mean 2.2/PV), resulting in successful isolation of 408/412 (99%) PVs at the end of the procedure. Mean electrodes' impedance drop was significantly higher in the SS-PVI compared with non-SS applications (21.5 & PLUSMN; 6.6 vs. 18.6 & PLUSMN;6.5 Ohm). Concordantly, higher tem-perature rise was observed in the SS vs. non-SS applications (10.9 & PLUSMN; 4.9 & DEG;C vs. 9.6 & PLUSMN;4.7 & DEG;C).Conclusions: In this multicenter real-world study, mean impedance drop and temperature rise were associated with successful SS-PVI applying the novel RFB catheter. These parameters may help to guide efficient usage of the new RF balloon.
引用
收藏
页码:950 / 956
页数:7
相关论文
共 50 条
  • [21] Preclinical evaluation of a novel single-shot pulsed field ablation system for pulmonary vein and atrial ablation
    Aryana, Arash
    Ji, Sang Yong
    Hata, Cary
    de la Rama, Alan
    Nguyen, Ken
    Panescu, Dorin
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (11) : 2203 - 2212
  • [22] Characterization of durability and reconnection patterns at time of repeat ablation after single-shot pulsed field pulmonary vein isolation
    Ruwald, Martin H.
    Haugdal, Martin
    Worck, Rene
    Johannessen, Arne
    Hansen, Morten Lock
    Sorensen, Samuel K.
    Hansen, Jim
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2024, 67 (02) : 379 - 387
  • [23] Cerebral microembolization during atrial fibrillation ablation: Comparison of different single-shot ablation techniques
    Kiss, Alexandra
    Nagy-Balo, Edina
    Sandorfi, Gabor
    Edes, Istvan
    Csanadi, Zoltan
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2014, 174 (02) : 276 - 281
  • [24] The Anatomic Impact of Sequential, Additional, Ostial Radiofrequency Ablation Following Pulmonary Vein Cryo-Isolation
    Traulle, Sarah
    Miyazaki, Shinsuke
    Mouquet, Vincent
    Kubala, Maciej
    Hermida, Jean-Sylvain
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (12): : 1420 - 1427
  • [25] Pulmonary vein isolation using second-generation single-shot devices: not all the same?
    Seidl, Philipp
    Steinborn, Frank
    Costello-Boerrigter, Lisa
    Surber, Ralf
    Schulze, Paul C.
    Bottcher, Christine
    Sommermeier, Andreas
    Mattea, Violeta
    Simeoni, Roland
    Malur, Frank Michael
    Lapp, Harald
    Schade, Anja
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 60 (03) : 521 - 528
  • [26] Novel Cryoballoon Ablation System for Single Shot Pulmonary Vein Isolation - The Prospective ICE-AGE-X Study -
    Tilz, Roland Richard
    Meyer-Saraei, Roza
    Eitel, Charlotte
    Fink, Thomas
    Sciacca, Vanessa
    Lopez, Lisbeth Delgado
    Kirstein, Bettina
    Schluter, Michael
    Vogler, Julia
    Kuck, Karl-Heinz
    Heeger, Christian-H
    CIRCULATION JOURNAL, 2021, 85 (08) : 1296 - 1304
  • [27] Comparison between a Novel Radiofrequency-Balloon and a Standard Cryo-Balloon in Pulmonary Vein Isolation: A Propensity-Score-Matched Analysis
    Teumer, Yannick
    Miesbichler, Clemens
    Katov, Lyuboslav
    Mayer, Benjamin
    Rottbauer, Wolfgang
    Bothner, Carlo
    Weinmann, Karolina
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (04)
  • [28] Atrial Fibrillation Ablation with a Novel Fully 3D-Mapping-Integrated Multi-Electrode Radiofrequency Balloon Catheter
    Teumer, Yannick
    Miesbichler, Clemens
    Hauke, Andreas
    Katov, Lyuboslav
    Bothner, Carlo
    Pott, Alexander
    Mueller, Martin
    Walter, Benjamin
    Rottbauer, Wolfgang
    Dahme, Tillman
    Weinmann, Karolina
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (01)
  • [29] Lesion Characteristics Using a Novel Multi-Electrode Radiofrequency Balloon Ablation Catheter
    Lee, Byron K.
    Al-Ahmad, Amin
    Rippy, Marian
    Reddy, Vivek
    Narayan, Girish
    Kuck, Karl-Heinz
    Crozier, Ian
    Daly, Matt
    Melton, Iain
    Walters, Tomas E.
    Lee, Randall J.
    Badhwar, Nitish
    CIRCULATION, 2016, 134
  • [30] Impact of Left Atrial Bipolar Electrogram Voltage on First Pass Pulmonary Vein Isolation During Radiofrequency Catheter Ablation
    Garg, Lohit
    Pothineni, Naga Venkata K.
    Daw, J. Michael
    Hyman, Matthew C.
    Arkles, Jeffrey
    Tschabrunn, Cory M.
    Santangeli, Pasquale
    Marchlinski, Francis E.
    FRONTIERS IN PHYSIOLOGY, 2020, 11