A Robust HPLC Approach for Quantitation of Camptothecin in Mesoporous Silica Nanoparticles Matrix and in the Presence of Its Degradation Products<A3B2/><A3B2/><A3B2/><A3B2/>

被引:6
作者
Dave, Richa [1 ]
Savaliya, Neel [1 ]
Dobariya, Ashish [1 ,2 ]
Patel, Rashmin [1 ]
Patel, Mrunali [1 ,3 ]
机构
[1] Charotar Univ Sci & Technol, Ramanbhai Patel Coll Pharm, Anand 388421, Gujarat, India
[2] Kashiv Biosci LLC, Analyt Res & Dev, Ahmadabad 382210, Gujarat, India
[3] Charotar Univ Sci & Technol, Ramanbhai Patel Coll Pharm, Dept Pharmaceut, Anand 388421, Gujarat, India
关键词
VALIDATION; STABILITY;
D O I
10.1093/jaoacint/qsae013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Background Camptothecin is a potent anti-cancer drug used for the treatment of various cancers. Objective The goal of this research investigation was to develop and validate a new stability indicating HPLC technique for the quantitative assessment of camptothecin in in-house developed mesoporous silica nanoparticles, a novel nanoformulation matrix for the treatment of cancer. Method The Waters Inertsil (R) HPLC column (C18) was used for the chromatographic separation, with a flow rate of 1 mL/min, a column oven temperature of 40 degrees C, an injection volume of 10 mu L, a detection wavelength of 216 nm, and a 10 min runtime overall. An isocratic blend of phosphate buffer (10 mM, pH7.0) and acetonitrile (60:40 v/v) served as the mobile phase. Various stress conditions including acid, alkali, oxidative, photolytic, thermal, and humidity environments were tested for the quantitative estimation of the camptothecin through the proposed method. Results The results demonstrated that the proposed method is specific (peak purity >= 0.999), accurate (99.69-100.64% w/w), precise (%RSD < 2.0), and sensitive (LOD-0.17 mu g and LOQ-0.56 mu g) in accordance with ICH guideline Q2(R1). Any unidentified degradation products did not interfere with the drug's estimation. Furthermore, the current method of analysis has eliminated any excipient interference from the matrix effect caused by the numerous excipients of the formulation matrix. Conclusions To quantify camptothecin for routine assay purposes, this research work offers a novel and straightforward HPLC methodology with optimized chromatographic parameters, contributing to the research and development community while ensuring an appropriate and efficient use of the drug through a variety of nanoformulation for cancer treatment. Highlights Stability-indicating HPLC method was found to be specific and suitable for routine analysis of camptothecin. The absence of any interference from excipients was confirmed by forced degradation studies.
引用
收藏
页码:387 / 395
页数:9
相关论文
共 31 条
[1]   An analytical "quality by design" approach in RP-HPLC method development and validation for reliable and rapid estimation of irinotecan in an injectable formulation [J].
Bhaskaran, Navya Ajitkumar ;
Kumar, Lalit ;
Reddy, M. Sreenivasa ;
Pai, Girish K. .
ACTA PHARMACEUTICA, 2021, 71 (01) :57-79
[2]  
Blessy M, 2014, J Pharm Anal, V4, P159, DOI 10.1016/j.jpha.2013.09.003
[3]  
Chimagave S. S., 2023, Journal of Applied Pharmaceutical Science, V13, P039, DOI [10.7324/JAPS.2023.130205, DOI 10.7324/JAPS.2023.130205]
[4]  
Dagron S., 2014, HDB ETHIK R FORSCH A, P541, DOI DOI 10.1007/978-3-642-35099-3_86
[5]   Development and validation of an HPLC method to quantify 3,4-dideoxyglucosone-3-ene in peritoneal dialysis fluids [J].
Frischmann, Matthias ;
Spitzer, Johanna ;
Fuenfrocken, Michael ;
Mittelmaier, Stefan ;
Deckert, Melanie ;
Fichert, Thomas ;
Pischetsrieder, Monika .
BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (08) :843-851
[6]   Characterization of camptothecin by analytical methods and determination of anticancer potential against prostate cancer [J].
Galatage, Sunil T. ;
Trivedi, Rahul ;
Bhagwat, Durgacharan A. .
FUTURE JOURNAL OF PHARMACEUTICAL SCIENCES, 2021, 7 (01)
[7]  
Harron D W. G., 2013, Technical Requirements for Registration of Pharmaceuticals for Human Use: The ICH Process, The Textbook of Pharmaceutical Medicine, pag, P447, DOI DOI 10.1002/9781118532331.CH23
[8]  
Hua H., 2022, Grain Oil Sci. Technol., V5, P107, DOI DOI 10.1016/J.GAOST.2022.06.001
[9]  
ICH Q1A(R2), 2003, Q1AR2 ICH IFPMA, V4, P24
[10]   Development and Validation of HPLC Method to Determine Colchicine in Pharmaceutical Formulations and its Application for Analysis of Solid Lipid Nanoparticles [J].
Joshi, Sumit Ashok ;
Jalalpure, Sunil Satyappa ;
Kempwade, Amolkumar Ashok ;
Peram, Malleswara Rao .
CURRENT PHARMACEUTICAL ANALYSIS, 2018, 14 (01) :76-83