Comparison of four commercial dose calculation algorithms in different evaluation tests

被引:3
|
作者
Rostami, Aram [1 ]
De Castro Neto, Aluisio Jose [1 ]
Paloor, Satheesh Prasad [1 ]
Khalid, Abdul Sattar [1 ]
Hammoud, Rabih [1 ]
机构
[1] Natl Ctr Canc Care & Res, Dept Radiat Oncol, Doha, Qatar
关键词
Treatment Planning System (TPS); Anisotropic Analytical Algorithm (AAA); Acuros (AXB); collapsed cone; convolution (CCC); Monte Carlo (MC); TREATMENT PLANNING SYSTEM; MONTE-CARLO; CLINICAL IMPLEMENTATION; DOSIMETRIC ACCURACY; ACUROS XB; PHOTON; VALIDATION; ENERGY; VERIFICATION; RADIOTHERAPY;
D O I
10.3233/XST-230079
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Accurate and fast dose calculation is crucial in modern radiation therapy. Four dose calculation algorithms (AAA, AXB, CCC, and MC) are available in Varian Eclipse and RaySearch Laboratories RayStation Treatment Planning Systems (TPSs). OBJECTIVES: This study aims to evaluate and compare dosimetric accuracy of the four dose calculation algorithms applying to homogeneous and heterogeneous media, VMAT plans (based on AAPM TG-119 test cases), and the surface and buildup regions. METHODS: The four algorithms are assessed in homogeneous (IAEA-TECDOCE 1540) and heterogeneous (IAEATECDOC 1583) media. Dosimetric evaluation accuracy for VMAT plans is then analyzed, along with the evaluation of the accuracy of algorithms applying to the surface and buildup regions. RESULTS: Tests conducted in homogeneous media revealed that all algorithms exhibit dose deviations within 5% for various conditions, with pass rates exceeding 95% based on recommended tolerances. Additionally, the tests conducted in heterogeneous media demonstrate high pass rates for all algorithms, with a 100% pass rate observed for 6MV and mostly 100% pass rate for 15MV, except for CCC, which achieves a pass rate of 94%. The results of gamma index pass rate (GIPR) for dose calculation algorithms in IMRT fields show that GIPR (3%/3 mm) for all four algorithms in all evaluated tests based on TG119, are greater than 97%. The results of the algorithm testing for the accuracy of superficial dose reveal variations in dose differences, ranging from -11.9% to 7.03% for 15MV and -9.5% to 3.3% for 6MV, respectively. It is noteworthy that the AXB and MC algorithms demonstrate relatively lower discrepancies compared to the other algorithms. CONCLUSIONS: This study shows that generally, two dose calculation algorithms (AXB and MC) that calculate dose in medium have better accuracy than other two dose calculation algorithms (CCC and AAA) that calculate dose to water.
引用
收藏
页码:1013 / 1033
页数:21
相关论文
共 50 条
  • [1] Superficial dose evaluation of four dose calculation algorithms
    Cao, Ying
    Yang, Xiaoyu
    Yang, Zhen
    Qiu, Xiaoping
    Lv, Zhiping
    Lei, Mingjun
    Liu, Gui
    Zhang, Zijian
    Hu, Yongmei
    RADIATION PHYSICS AND CHEMISTRY, 2017, 137 : 23 - 28
  • [2] Evaluation of the accuracy of various dose calculation algorithms of a commercial treatment planning system in the presence of hip prosthesis and comparison with Monte Carlo
    Mohammadi, Kheirollah
    Hassani, Mohsen
    Ghorbani, Mahdi
    Farhood, Bagher
    Knaup, Courtney
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (03) : 501 - 509
  • [3] Dose distribution evaluation of various dose calculation algorithms in inhomogeneous media
    Kim, Y. L.
    Suh, T. S.
    Choe, B. Y.
    Choi, B. O.
    Chung, J. B.
    Lee, J. W.
    Bae, Y. K.
    Park, B. M.
    Jung, J. Y.
    Shin, Y. J.
    Kim, J. Y.
    Moon, S. K.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (04): : 269 - 278
  • [4] Comparison of five dose calculation algorithms in a heterogeneous media using design of experiment
    Vangvichith, M.
    Autret, D.
    Tiplica, T.
    Barreau, M.
    Dufreneix, S.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 61 : 103 - 111
  • [5] Evaluation of a commercial Monte Carlo dose calculation algorithm for electron treatment planning
    Huang, Jessie Y.
    Dunkerley, David
    Smilowitz, Jennifer B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06): : 184 - 193
  • [6] Evaluation of heterogeneity dose distributions for Stereotactic Radiotherapy (SRT): comparison of commercially available Monte Carlo dose calculation with other algorithms
    Takahashi, Wataru
    Yamashita, Hideomi
    Saotome, Naoya
    Iwai, Yoshio
    Sakumi, Akira
    Haga, Akihiro
    Nakagawa, Keiichi
    RADIATION ONCOLOGY, 2012, 7
  • [7] Stereotactic radiotherapy of intrapulmonary lesions: comparison of different dose calculation algorithms for Oncentra MasterPlan®
    Troeller, Almut
    Garny, Sylvia
    Pachmann, Sophia
    Kantz, Steffi
    Gerum, Sabine
    Manapov, Farkhad
    Ganswindt, Ute
    Belka, Claus
    Soehn, Matthias
    RADIATION ONCOLOGY, 2015, 10
  • [8] Evaluation of a new commercial Monte Carlo dose calculation algorithm for electron beams
    Vandervoort, Eric J.
    Tchistiakova, Ekaterina
    La Russa, Daniel J.
    Cygler, Joanna E.
    MEDICAL PHYSICS, 2014, 41 (02) : 021711
  • [9] Dosimetric evaluation of the dose calculation accuracy of different algorithms for two different treatment techniques during whole breast irradiation
    Acar, Hilal
    Altinok, Ayse Yildirim
    Cebe, Mehmet Siddik
    JOURNAL OF RADIATION ONCOLOGY, 2019, 8 (03) : 337 - 347
  • [10] Comparison between Acuros XB and Brainlab Monte Carlo algorithms for photon dose calculation
    Misslbeck, M.
    Kneschaurek, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 (07) : 599 - 605