Developing a medical device-grade T2 phantom optimized for myocardial T2 mapping by cardiovascular magnetic resonance

被引:3
作者
Topriceanu, Constantin-Cristian [1 ,2 ,5 ]
Fornasiero, Massimiliano [3 ]
Seo, Han [4 ]
Webber, Matthew [1 ,2 ,4 ,5 ]
Keenan, Kathryn E. E. [6 ]
Stupic, Karl F. F. [6 ]
Bruehl, Ruediger [7 ]
Ittermann, Bernd [7 ]
Price, Kirsty [8 ]
McGrath, Louise [8 ]
Pang, Wenjie [9 ]
Hughes, Alun D. D. [2 ,5 ]
Nezafat, Reza [10 ]
Kellman, Peter [11 ]
Pierce, Iain [1 ,2 ]
Moon, James C. C. [1 ,2 ]
Captur, Gabriella [2 ,4 ,5 ,12 ]
机构
[1] Barts Hlth NHS Trust, Barts Heart Ctr, ECIA 7BE, London, England
[2] UCL, Inst Cardiovasc Sci, Huntley St, London WC1E 6DD, England
[3] UCL Med Sch, 74 Huntley St, London WC1E 6DE, England
[4] Royal Free London NHS Fdn Trust, Ctr Inherited Heart Muscle Condit, Dept Cardiol, Pond St, London NW3 2QG, England
[5] UCL, Med Res Council Unit Lifelong Hlth & Ageing, 1-19 Torrington Pl, London WC1E 7HB, England
[6] NIST, 325 Broadway, Boulder, CO 80305 USA
[7] Phys Tech Bundesanstalt PTB, Abbestr 2-12, D-10587 Berlin, Germany
[8] UCL Bloomsbury Ctr Clin Phenotyping, London WC1E 6HX, England
[9] Resonance Hlth RH, 141 Burswood Rd, Burswood, WA 6100, Australia
[10] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Med, Cardiovasc Div, 330 Brookline Ave, Boston, MA 02215 USA
[11] NHLBI, NIH, Bethesda, MD 20892 USA
[12] UCL, Inst Cardiovasc Sci, Inherited Heart Muscle Condit, Gower St, London WC1E 6BT, England
基金
英国医学研究理事会; 英国经济与社会研究理事会; 欧盟地平线“2020”;
关键词
T-1; mapping; T-2; Phantom; Quality control; MRI; FIELD; ECHO; SEQUENCES; VALUES; 3T;
D O I
10.1186/s12968-023-00926-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionA long T-2 relaxation time can reflect oedema, and myocardial inflammation when combined with increased plasma troponin levels. Cardiovascular magnetic resonance (CMR) T-2 mapping therefore has potential to provide a key diagnostic and prognostic biomarkers. However, T-2 varies by scanner, software, and sequence, highlighting the need for standardization and for a quality assurance system for T-2 mapping in CMR.AimTo fabricate and assess a phantom dedicated to the quality assurance of T-2 mapping in CMR.MethodA T-2 mapping phantom was manufactured to contain 9 T-1 and T-2 (T-1|T-2) tubes to mimic clinically relevant native and post-contrast T-2 in myocardium across the health to inflammation spectrum (i.e., 43-74 ms) and across both field strengths (1.5 and 3 T). We evaluated the phantom's structural integrity, B-0 and B-1 uniformity using field maps, and temperature dependence. Baseline reference T-1|T-2 were measured using inversion recovery gradient echo and single-echo spin echo (SE) sequences respectively, both with long repetition times (10 s). Long-term reproducibility of T-1|T-2 was determined by repeated T-1|T-2 mapping of the phantom at baseline and at 12 months.ResultsThe phantom embodies 9 internal agarose-containing T-1|T-2 tubes doped with nickel di-chloride (NiCl2) as the paramagnetic relaxation modifier to cover the clinically relevant spectrum of myocardial T-2. The tubes are surrounded by an agarose-gel matrix which is doped with NiCl2 and packed with high-density polyethylene (HDPE) beads. All tubes at both field strengths, showed measurement errors up to <= 7.2 ms [< 14.7%] for estimated T-2 by balanced steady-state free precession T-2 mapping compared to reference SE T-2 with the exception of the post-contrast tube of ultra-low T-1 where the deviance was up to 16 ms [40.0%]. At 12 months, the phantom remained free of air bubbles, susceptibility, and off-resonance artifacts. The inclusion of HDPE beads effectively flattened the B-0 and B-1 magnetic fields in the imaged slice. Independent temperature dependency experiments over the 13-38 degrees C range confirmed the greater stability of shorter vs longer T-1|T-2 tubes. Excellent long-term (12-month) reproducibility of measured T-1|T-2 was demonstrated across both field strengths (all coefficients of variation < 1.38%).ConclusionThe T-2 mapping phantom demonstrates excellent structural integrity, B-0 and B-1 uniformity, and reproducibility of its internal tube T-1|T-2 out to 1 year. This device may now be mass-produced to support the quality assurance of T-2 mapping in CMR.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Application of magnetic resonance T2 mapping in the temporomandibular joints
    Cao, Yang
    Xia, Chunchao
    Wang, Si
    He, Shushu
    Wamalwa, Peter
    Chen, Song
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2012, 114 (05): : 644 - 649
  • [22] Fat-suppressed T2 mapping of human knee femoral articular cartilage: comparison with conventional T2 mapping
    Kim, Bo Ram
    Yoo, Hye Jin
    Chae, Hee-Dong
    Hong, Sung Hwan
    Choi, Ja-Young
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [23] Myocardial T2* Mapping with Ultrahigh Field Magnetic Resonance: Physics and Frontier Applications
    Huelnhagen, Till
    Paul, Katharina
    Ku, Min-Chi
    Duarte, Teresa Serradas
    Niendorf, Thoralf
    [J]. FRONTIERS IN PHYSICS, 2017, 5
  • [24] T2 mapping in patellar chondromalacia
    Ruiz Santiago, Fernando
    Pozuelo Calvo, Rocio
    Almansa Lopez, Julio
    Guzman Alvarez, Luis
    Castellano Garcia, Maria del Mar
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (06) : 984 - 988
  • [25] T2 Mapping in Prostate Cancer
    Mai, Julia
    Abubrig, Mohamed
    Lehmann, Thomas
    Hilbert, Tom
    Weiland, Elisabeth
    Grimm, Marc O.
    Teichgraeber, Ulf
    Franiel, Tobias
    [J]. INVESTIGATIVE RADIOLOGY, 2019, 54 (03) : 146 - 152
  • [26] T2 mapping in myocardial disease: a comprehensive review
    O'Brien, Aaron T.
    Gil, Katarzyna E.
    Varghese, Juliet
    Simonetti, Orlando P.
    Zareba, Karolina M.
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2022, 24 (01)
  • [27] Myocardial T2 Mapping in Patients With Hypertrophic Cardiomyopathy
    Amano, Yasuo
    Yanagisawa, Fumi
    Tachi, Masaki
    Hashimoto, Hidenobu
    Imai, Shogo
    Kumita, Shinichiro
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (03) : 344 - 348
  • [28] T2 mapping cardiovascular magnetic resonance identifies the presence of myocardial inflammation in patients with dilated cardiomyopathy as compared to endomyocardial biopsy
    Spieker, M.
    Katsianos, E.
    Gastl, M.
    Behm, P.
    Horn, P.
    Jacoby, C.
    Schnackenburg, B.
    Reinecke, P.
    Kelm, M.
    Westenfeld, R.
    Boenner, F.
    [J]. EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2018, 19 (05) : 574 - 582
  • [29] Black blood myocardial T2 mapping
    Wang, Chengyan
    Jang, Jihye
    Neisius, Ulf
    Nezafat, Maryam
    Fahmy, Ahmed
    Kang, Jinkyu
    Rodriguez, Jennifer
    Goddu, Beth
    Pierce, Patrick
    Berg, Sophie
    Zhang, Jue
    Wang, Xiaoying
    Nezafat, Reza
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (01) : 153 - 166
  • [30] T1ρ, T2 and T2*mapping of lumbar intervertebral disc degeneration: a comparison study
    Yang, Li
    Sun, Cong
    Gong, Tao
    Li, Quanlin
    Chen, Xin
    Zhang, Xinjuan
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)