Understanding the ablation rate of Holmium:YAG and thulium fiber lasers. Perspectives from an in vitro study

被引:9
|
作者
Basulto-Martinez, Mario [1 ,2 ,3 ]
Proietti, Silvia [1 ,2 ]
Pavia, Maria Pia [1 ,2 ,4 ]
Yeow, Yuyi [2 ]
Eisner, Brian H. H. [5 ]
Giusti, Guido [1 ,2 ,6 ]
机构
[1] IRCCS San Raffaele Hosp, Dept Urol, Milan, Italy
[2] European Training Ctr Endourol, Milan, Italy
[3] Hosp Reg Alta Especial Peninsula Yucatan, Dept Urol, Merida, Mexico
[4] Marche Polytech Univ, Osped Riuniti Univ Hosp, Dept Urol, Ancona, Italy
[5] Massachusetts Gen Hosp, Dept Urol, Boston, MA USA
[6] Osped San Raffaele Srl, IRCCS, Via Olgettina 60, I-20132 Milan, Italy
关键词
Kidney stones; Urolithiasis; Lithotripsy; Holmium laser; Thulium fiber laser; LITHOTRIPSY;
D O I
10.1007/s00240-022-01402-6
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
This study sought to analyze the ablation rates of Holmium:YAG (Ho:YAG) and thulium fiber laser (TFL) under different settings combinations. Lasering was carried out by a computed spiral pattern through a stepper motor bearing a laser fiber. BegoStones were placed inside a water container and lasered therein and the ablation rate was calculated. Different combinations were tested for similar to 12 Watts (W) for Ho:YAG and TFL lasers. Further 25 W combination for TFL were conducted testing higher frequencies. One-hundred-one experiments were conducted. Under similar to 12 W, the highest ablation rate for Ho:YAG was obtained with 40 Hz/0.3 J, and Virtual Basket (TM) (VB) emission mode [114.35 (88.30 - 126.40) mg/min] whereas for TFL, using 40 Hz/0.3 J, and long pulse [143.40 (137.40 - 146) mg/min]. A matched-comparison for 12 Hz/1 J settings showed that TFL outperformed Ho:YAG using medium pulse [63.30 (55.30 - 81.30) vs 132.00 (115.70 - 228.60) mg/min, p=0.016] and long pulse [62.90 (51.60 - 78.90) vs 134.70 (110.60 - 149.30) mg/min, p=0.016]. In the 40 Hz/0.3 J settings, TFL outperformed Ho:YAG (p=0.034). Nonetheless, Ho:YAG with VB's ablation rate increased [114.35 (88.30 - 126.40)], being comparable to that of TFL with medium pulse [127.5 (88.90 - 141.70)] (p=0.400). In TFL 25 W experiments, the ablation rate of 500 Hz / 0.05 J, was higher than those of 1000 Hz / 0.025 J (p=0.049). TFL ablation rate is higher than that of Ho:YAG. Moreover, the Virtual Basket (TM) emission mode, increased Ho:YAG ablation rates, resulting reaching similar to that of TFL in certain modalities.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Understanding the ablation rate of Holmium:YAG and thulium fiber lasers. Perspectives from an in vitro study
    Mario Basulto-Martínez
    Silvia Proietti
    Maria Pia Pavia
    Yuyi Yeow
    Brian H. Eisner
    Guido Giusti
    Urolithiasis, 51
  • [2] Assessment of Holmium:YAG, Pulsed-Thulium:YAG and Thulium Fiber Lasers for Urinary Stone Ablation. In Vitro Study
    Sierra, Alba
    Panthier, Frederic
    Castillo, Esther
    Mercade, Asier
    Peri, Lluis
    Alcaraz, Antonio
    Traxer, Olivier
    Lopez, Juan Manuel
    Luque, Maria Pilar
    JOURNAL OF ENDOUROLOGY, 2024, 38 (11) : 1209 - 1216
  • [3] An in vitro analysis of holmium: YAG and thulium fiber laser ablation rates
    Basulto-Martinez, M.
    Proietti, S.
    Pavia, M. P.
    Rico, L.
    Yeow, Y.
    Flores-Tapia, J. P.
    Esqueda-Mendoza, A.
    Gaboardi, F.
    Giusti, G.
    EUROPEAN UROLOGY, 2021, 79 : S305 - S305
  • [4] Comparison of Holmium: YAG and Thulium Fiber Lasers for Lithotripsy
    Blackmon, Richard L.
    Irby, Pierce B.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS VI, 2010, 7548
  • [5] In Vitro Comparison of Pulsed-Thulium:YAG, Holmium:YAG, and Thulium Fiber Laser
    Chicaud, Marie
    Kutchukian, Stessy
    Berthe, Laurent
    Corrales, Mariela
    Solano, Catalina
    Candela, Luigi
    Doizi, Steeve
    Smith, Daron
    Traxer, Olivier
    Panthier, Frederic
    JOURNAL OF ENDOUROLOGY, 2024, 38 (12) : 1427 - 1435
  • [6] Comparison of Holmium:YAG and Thulium Fiber Lasers on the Risk of Laser Fiber Fracture
    Uzan, Audrey
    Chiron, Paul
    Panthier, Frederic
    Haddad, Mattieu
    Berthe, Laurent
    Traxer, Olivier
    Doizi, Steeve
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (13)
  • [7] Ablation rates with Holmium:YAG and Thulium Fiber Laser: Influence of the stone phantom homogeneity. An in vitro study
    Kutchukian, S.
    Doizi, S.
    Lapouge, P.
    Germain, T.
    Dragos, L.
    Berthe, L.
    Solano, C.
    Candela, L.
    Corrales, M.
    Chicaud, M.
    Traxer, O.
    Panthier, F.
    PROGRES EN UROLOGIE, 2023, 33 (8-9): : 456 - 462
  • [8] In-vitro comparison of stone ablation volume using Pulsed Thulium:YAG and modulated pulsed holmium: YAG lasers
    Balado, Bravo A. C.
    Sanchez-Puy, A.
    Izquierdo, P.
    Diana, P.
    Piana, A.
    Kanashiro, A. S.
    Balana, J.
    Millan, F. L.
    Gallioli, A.
    Gracia, S.
    Angerri, O.
    Emiliani, E.
    EUROPEAN UROLOGY, 2024, 85 : S1301 - S1302
  • [9] Comparison of Holmium:YAG and Thulium Fiber Lasers on Soft Tissue: An Ex Vivo Study
    Doizi, Steeve
    Germain, Thibault
    Panthier, Frederic
    Comperat, Eva
    Traxer, Olivier
    Berthe, Laurent
    JOURNAL OF ENDOUROLOGY, 2022, 36 (02) : 251 - 258
  • [10] PULSED-THULIUM:YAG LASER FOR ENDOSCOPIC TREATMENT OF KIDNEY STONE. COMPARATIVE ANALYSIS WITH HOLMIUM:YAG AND THULIUM FIBER LASERS IN AN IN VITRO MODEL
    Astroza, Gaston M.
    Manterola, Carla
    Navarro, Renato
    Salvado, Jose A.
    Pauchard, Felipe
    JOURNAL OF UROLOGY, 2025, 213 (5S):