Research on the temperature characteristics of SiPM-based LaBr3(Ce) detectors

被引:2
作者
Yu, Tao [1 ]
Guo, Siming [2 ]
Wu, Jinjie [2 ]
An, Zhenghua [3 ]
Qie, Xiaoyu [4 ]
Guo, Kaiyue [5 ]
机构
[1] Chengdu Univ Technol, Chengdu 610059, Peoples R China
[2] Natl Inst Metrol, Div Ionizing Radiat Metrol, Beijing 102200, Peoples R China
[3] Chinese Acad Sci, Inst High Energy Phys, Beijing 10049, Peoples R China
[4] Hebei Univ Sci & Technol, Shijiazhuang 050018, Peoples R China
[5] China Jiliang Univ, Hangzhou 310018, Zhejiang, Peoples R China
关键词
LaBr3(Ce) crystal; SiPM array; Energy resolution; Temperature response; Peak-to-total ratio;
D O I
10.1007/s40042-022-00634-4
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
When a satellite-borne detector is being ground calibrated, a standard detector is required to determine the monochromatic X-ray beam energy and intensity. For this purpose, a SiPM-based LaBr3(Ce) detector has been developed. The environmental temperature change may substantially affect the performance of the detector, thus affecting the accuracy of the ground calibration results. To study the effect of temperature change on the performance of the LaBr3(Ce) detector, the performance of the LaBr3(Ce) detector was tested with Am-241 and Cs-137 gamma-ray sources in the high- and low-temperature test chambers. The results showed that within the range of - 20 to 40 & DEG;C, the energy resolution of the LaBr3(Ce) detector does not change by more than 1%, nor does the peak-to-total ratio of Am-241 and Cs-137. The temperature change has little effect on the performance of the LaBr3(Ce) detector, which meets the requirements for use as a beam monitor. However, the peak location of the radioactive source varies significantly with temperature, and the difference with the peak location at room temperature is within 17%. To reduce the variation in the gain of the LaBr3(Ce) detector with temperature, it is necessary to optimize the compensation design of the temperature gain to ensure the stability of the detector gain at various temperatures.
引用
收藏
页码:12 / 18
页数:7
相关论文
共 16 条
[1]  
Cova S., 1996, APPL OPTICS, V35
[2]  
Dai Z., 2015, J SPACECRAFT TT C TE, V34, P5
[3]  
[高鑫 GAO Xin], 2010, [核电子学与探测技术, Nuclear Electronics and Detection Technology], V30, P5
[4]  
Gui Q., 2011, NUCL ELECT DETECT TE, V21, P1195
[5]  
Ji C.S., 1990, HDB NUCL RAD DETECTO, P6
[6]  
Lecoq P., 2006, PHYS PRINCIPLES CRYS, P90
[7]  
Leo W.R., 1987, TECHNIQUES NUCL PART, P86
[8]   A novel analog power supply for gain control of the Multi-Pixel Photon Counter (MPPC) [J].
Li, Zhengwei ;
Liu, Congzhan ;
Xu, Yupeng ;
Yan, Bo ;
Li, Yanguo ;
Lu, Xuefeng ;
Li, Xufang ;
Zhang, Shuo ;
Chang, Zhi ;
Li, Jicheng ;
Gao, He ;
Zhang, Yifei ;
Zhao, Jianling .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 850 :35-41
[9]  
Mitchell L.J., 2019, 2019 IEEE NUCL SCI S, P1, DOI [10.1109/NSS/MIC42101.2019.9059652, DOI 10.1109/NSS/MIC42101.2019.9059652]
[10]   X-ray and gamma-ray response of a 2" x 2" LaBr3:Ce scintillation detector [J].
Quarati, F. ;
Bos, A. J. J. ;
Brandenburg, S. ;
Dathy, C. ;
Dorenbos, P. ;
Kraft, S. ;
Ostendorf, R. W. ;
Ouspenski, V. ;
Owens, Alan .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 574 (01) :115-120