Simple-Z: A Low-Cost Portable Impedance Analyzer

被引:4
作者
Buscaglia, Lorenzo A. [1 ]
Carmo, Joao Paulo [2 ]
Oliveira, Osvaldo N., Jr. [1 ]
机构
[1] Univ Sao Paulo, Sao Carlos Inst Phys, Dept Phys & Mat Sci, Bernhard Gross Polymers Grp, BR-13560970 Sao Carlos, Brazil
[2] Univ Sao Paulo, Sao Carlos Sch Engn, Dept Elect Engn, BR-13560970 Sao Carlos, Brazil
基金
巴西圣保罗研究基金会;
关键词
AD5933; biosensing; biosensor; calibration; electronic; impedance analyzer; impedance spectroscopy; sensor; signal processing; Simple-Z; WIRELESS POTENTIOSTAT; SPECTROSCOPY; SYSTEM;
D O I
10.1109/JSEN.2023.3312039
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The use of electrical impedance spectroscopy in point-of-care systems or for online monitoring has been hampered by the high cost of typical benchtop impedance analyzers. In this article, we introduce a portable, low-cost impedance analyzer, referred to as "Simple-Z," which contains a signal processing unit that allows for a precise generation and reading of the excitation and response signals. Simple-Z is based on the integrated circuit AD5933 in conjunction with peripheral circuits for a fine output amplitude regulation, flexible response amplification, sampling rate control, automatic calibration, and external communication. Software solutions were implemented, including the strategical register overflow to allow decreasing frequency sweeps, the control of the number of acquisition cycles to reduce the spectral leakage of the discrete Fourier transform (DFT), and the iterative approximation for defining proper response amplification and calibration resistances. The software also included a graphical user interface (GUI) with entries, buttons, and graphs that allow for an intuitive usage of the device. Results generated with Simple-Z in equivalent circuits mimicking sensing systems and in aqueous solution of different Na2SO4 concentrations were essentially identical to those obtained with a commercial benchtop impedance analyzer. Since the cost of the circuitry and components to fabricate an instrument is near U.S. $100 and because Simple-Z was designed to allow for easy upscale to mass production, one may expect it to be deployed in pointof-care diagnosis systems. Furthermore, it can also be used in teaching laboratories for training students in electrical impedance spectroscopy.
引用
收藏
页码:26067 / 26074
页数:8
相关论文
共 31 条
  • [1] Open-Source Potentiostat for Wireless Electrochemical Detection with Smartphones
    Ainla, Alar
    Mousavi, Maral P. S.
    Tsaloglou, Maria-Nefeli
    Redston, Julia
    Bell, Jeffrey G.
    Teresa Fernandez-Abedul, M.
    Whitesides, George M.
    [J]. ANALYTICAL CHEMISTRY, 2018, 90 (10) : 6240 - 6246
  • [2] Bogonez-Franco P., 2010, Journal of Physics: Conference Series, V224, DOI 10.1088/1742-6596/224/1/012163
  • [3] Low-cost bacterial nanocellulose-based interdigitated biosensor to detect the p53 cancer biomarker
    Bondancia, Thalita J.
    Soares, Andrey Coatrini
    Popolin-Neto, Mario
    Gomes, Nathalia O.
    Raymundo-Pereira, Paulo A.
    Barud, Hernane S.
    Machado, Sergio A. S.
    Ribeiro, Sidney J. L.
    Melendez, Matias E.
    Carvalho, Andre L.
    Reis, Rui M.
    Paulovich, Fernando, V
    Oliveira Jr, Osvaldo N.
    [J]. BIOMATERIALS ADVANCES, 2022, 134
  • [4] Photocatalysis of TiO2 Sensitized with Graphitic Carbon Nitride and Electrodeposited Aryl Diazonium on Screen-Printed Electrodes to Detect Prostate Specific Antigen under Visible Light
    Bott-Neto, Jose L.
    Martins, Thiago S.
    Buscaglia, Lorenzo A.
    Machado, Sergio A. S.
    Oliveira, Osvaldo N. Jr Jr
    [J]. ACS APPLIED MATERIALS & INTERFACES, 2022, 14 (19) : 22114 - 22121
  • [5] Buscaglia L. A., 2022, M.S. thesis, DOI [10.11606/D.76.2022.tde-19072022-120227, DOI 10.11606/D.76.2022.TDE-19072022-120227]
  • [6] Roadmap for Electrical Impedance Spectroscopy for Sensing: A Tutorial
    Buscaglia, Lorenzo A.
    Oliveira, Osvaldo N.
    Carmo, Joao Paulo
    [J]. IEEE SENSORS JOURNAL, 2021, 21 (20) : 22246 - 22257
  • [7] Caytak H, 2019, ENCYCLOPEDIA OF BIOMEDICAL ENGINEERING, VOL 3, P265, DOI 10.1016/B978-0-12-801238-3.10884-0
  • [8] SIMPLE WIDE FREQUENCY RANGE IMPEDANCE METER BASED ON AD5933 INTEGRATED CIRCUIT
    Chabowski, Konrad
    Piasecki, Tomasz
    Dzierka, Andrzej
    Nitsch, Karol
    [J]. METROLOGY AND MEASUREMENT SYSTEMS, 2015, 22 (01) : 13 - 24
  • [9] Microfluidic E-tongue to diagnose bovine mastitis with milk samples using Machine learning with Decision Tree models
    Coatrini-Soares, Andrey
    Coatrini-Soares, Juliana
    Neto, Mario Popolin
    de Mello, Suelen Scarpa
    Pinto, Danielle Dos Santos Cinelli
    Carvalho, Wanessa Araujo
    Gilmore, Michael S.
    Piazzetta, Maria Helena Oliveira
    Gobbi, Angelo Luiz
    Branda, Humberto de Mello
    Paulovich, Fernando Vieira
    Oliveira Jr, Osvaldo N.
    Mattoso, Luiz Henrique Capparelli
    [J]. CHEMICAL ENGINEERING JOURNAL, 2023, 451 (451)
  • [10] In Vivo Blood Characterization From Bioimpedance Spectroscopy of Blood Pooling
    Dai, Tao
    Adler, Andy
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (11) : 3831 - 3838