DESIGN OF PATIENT SPECIFIC HIP PROSTHESIS BASED ON FINITE ELEMENT ANALYSIS: A COMPARATIVE STUDY

被引:2
作者
Snekhalatha, U. [1 ,2 ,5 ]
Dhason, Raja [3 ]
Rajalakshmi, T. [4 ]
机构
[1] SRM Inst Sci & Technol, Coll Engn & Technol, Dept Biomed Engn, Kattankulathur, India
[2] Batangas State Univ, Coll Engn & Fine Arts, Batangas, Philippines
[3] SRM Inst Sci & Technol, Coll Engn & Technol, Dept Mech Engn, Kattankulathur, India
[4] SRM Inst Sci & Technol, Coll Engn & Technol, Dept Elect & Commun Engn, Chennai, India
[5] SRM Inst Sci & Technol, Dept Biomed Engn, Kattankulathur, Tamil Nadu, India
来源
BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS | 2023年 / 35卷 / 04期
关键词
Finite element analysis; Customized hip implant; Conventional hip implant; Femur bone; ARTHROPLASTY; RANGE;
D O I
10.4015/S1016237223500175
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aims to develop a patient-specific hip implant for osteoarthritis conditions and to compare with intact and conventional implant. The femoral bone with head and shaft region was segmented from the pelvic griddle and converted into 3D model. The parameters such as femoral ball diameter, shaft length, acetabular cup diameter, and neck angle were measured from the segmented 3D model. In this study, designed part of hip implant was assembled together to form a customized hip implant. The von Mises stress was measured by means of Finite element analysis (FEA) method by applying various forces applied at the distal end of hip implant. The forces applied at hip implant were based on the assumption of 500 N force for standing, 2000 N force for walking, and 3000 N force for jogging condition. The minimum stress attained at the femur bone of custom-model is 1.32 MPa for 500 N loading condition, 5.3 MPa for 2000 N and 7.96 MPa for the maximum load of 3000 N. Thus the customized model experienced better stress distribution compared to conventional model under the maximum load of 3000 N. In pelvic region, the custom model attained a lower stress of 23% compared to conventional model. Thus, the study recommends the customized hip implants for the osteoarthritis conditions to avoid revision surgery.
引用
收藏
页数:10
相关论文
共 43 条
[21]   Wear estimation of hip implants with varying chamfer geometry at the trunnion junction: a finite element analysis [J].
Gutmann, Celine ;
Shaikh, Numa ;
Shenoy, B. Satish ;
Bhat, N. Shaymasunder ;
Keni, Laxmikant G. ;
Chethan, K. N. .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (03)
[22]   Design of patient-specific hip implants based on the 3D geometry of the human femur [J].
Jun, Yongtae ;
Choi, Kuiwoon .
ADVANCES IN ENGINEERING SOFTWARE, 2010, 41 (04) :537-547
[23]   Analysis of stress distribution around total hip stems custom-designed for the standardized Asian femur configuration [J].
Jung, Jin Mu ;
Kim, Cheol Sang .
BIOTECHNOLOGY & BIOTECHNOLOGICAL EQUIPMENT, 2014, 28 (03) :525-532
[24]  
K N Chethan, 2019, J Biomed Phys Eng, V9, P507, DOI [10.31661/jbpe.v0i0.1210, 10.31661/jbpe.v0i0.1210]
[25]   Patient-specific finite element analysis of femurs with cemented hip implants [J].
Katz, Yekutiel ;
Lubovsky, Omri ;
Yosibash, Zohar .
CLINICAL BIOMECHANICS, 2018, 58 :74-89
[26]   Finite Element Analysis of Orthopedic Hip Implant with Functionally Graded Bioinspired Lattice Structures [J].
Kladovasilakis, Nikolaos ;
Tsongas, Konstantinos ;
Tzetzis, Dimitrios .
BIOMIMETICS, 2020, 5 (03)
[27]  
Kn C, 2018, International Journal of Engineering & Technology, V7, P1672, DOI [10.14419/ijet.v7i3.15231, 10.14419/ijet.v7i3.15231, DOI 10.14419/IJET.V7I3.15231]
[28]   Effect of critical shoulder angle, glenoid lateralization, and humeral inclination on range of movement in reverse shoulder arthroplasty [J].
Laedermann, A. ;
Tay, E. ;
Collin, P. ;
Piotton, S. ;
Chiu, C-H ;
Michelet, A. ;
Charbonnier, C. .
BONE & JOINT RESEARCH, 2019, 8 (08) :378-386
[29]   Femoral Stress Changes after Total Hip Arthroplasty with the Ribbed Prosthesis: A Finite Element Analysis [J].
Luo, Changqi ;
Wu, Xiang-Dong ;
Wan, Yifei ;
Liao, Junyi ;
Cheng, Qiang ;
Tian, Mian ;
Bai, Zhibiao ;
Huang, Wei .
BIOMED RESEARCH INTERNATIONAL, 2020, 2020
[30]  
Madeti Bhaskar Kumar, 2018, International Journal of Medical Engineering and Informatics, V10, P235, DOI 10.1504/ijmei.2018.093353