Adhesive Carbon Paste Electrode for Convenient and Long-term Electrocardiogram Monitoring: A Pilot Study

被引:0
作者
Wang, Xin [1 ,2 ,3 ]
Tian, Qiong [1 ,3 ]
Zhang, Biying [4 ]
Xu, Yangjie [5 ]
Ji, Junyu [1 ,2 ,3 ]
He, Yuchao [1 ,3 ]
Huang, Jinzhu [6 ]
Zhu, Mingxing [1 ,3 ,7 ]
Wang, Lin [1 ,3 ]
Chen, Shixiong [1 ,3 ]
Liu, Zhiyuan [1 ,3 ]
Li, Guanglin [1 ,3 ]
机构
[1] Chinese Acad Sci, Shenzhen Inst Adv Technol, CAS Key Lab Human Machine Intelligence Synergy Sy, Shenzhen 518055, Guangdong, Peoples R China
[2] Univ Chinese Acad Sci, Shenzhen Coll Adv Technol, Shenzhen 518055, Guangdong, Peoples R China
[3] Guangdong Hong Kong Macao Joint Lab Human Machine, Shenzhen 518055, Guangdong, Peoples R China
[4] Ningbo 2 Hosp, Dept Gastroenterol, Ningbo 315099, Peoples R China
[5] Univ Luxembourg, Interdisciplinary Ctr Secur Reliabil & Trust, L-1855 Luxembourg, Luxembourg
[6] Guangdong Med Univ, Dept Endocrine, Dongguan Affiliated Hosp 1, Dongguan 523710, Peoples R China
[7] Harbin Inst Technol, Sch Elect & Informat Engn, Shenzhen 518055, Peoples R China
来源
2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC | 2023年
基金
中国国家自然科学基金; 中国博士后科学基金;
关键词
D O I
10.1109/EMBC40787.2023.10341094
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Long-term electrocardiogram (ECG) monitoring is an important and widely-used technique in the clinic that helps with the diagnosis of possible diseases that cannot be detected in a short time monitoring. However, the clinically used electrode needs conductive gel to reduce the impedance between the skin and the electrodes, which easily causes the possibility of allergy. Moreover, as the conductive gel becomes dry, the signal's quality will decrease accordingly. In this paper, we proposed a novel adhesive Carbon Paste Electrode (CPE) to achieve convenient and long-term ECG monitoring. By comparing the time-domain waveforms, the R-R peak intervals difference, and the Signal-to-Noise Ratio (SNR) of ECG with the traditional conductive gel-based electrode (Gel) in fixed and unfixed conditions, the performance of the proposed CPE was investigated. The results showed that the CPE could achieve similar ECG monitoring both in fixed and unfixed conditions. When on Day 2, the quality acquired by Gel began to decrease while CPE was still stable, which was obvious especially in unfixed condition. The R-R peak intervals showed that on Day 2, the Gel was unreliable with some abnormal points occurring. Besides, the results of SNR and average heart rate (AHR) also confirmed that the CPE could achieve similar results as Gel on Day 1 and outperformed Gel on Day 2. It is believed that the proposed CPE opens a window of high-quality long-term ECG monitoring with more convenience.
引用
收藏
页数:4
相关论文
共 19 条
[1]  
[Anonymous], About us
[2]  
Brahme A., 2014, Comprehensive Biomedical Physics
[3]   Vital Signs Monitoring with Wearable Sensors in High-risk Surgical Patients A Clinical Validation Study [J].
Breteler, Martine J. M. ;
KleinJan, Eline J. ;
Dohmen, Daan A. J. ;
Leenen, Luke P. H. ;
van Hillegersberg, Richard ;
Ruurda, Jelle P. ;
van Loon, Kim ;
Blokhuis, Taco J. ;
Kalkman, Cor J. .
ANESTHESIOLOGY, 2020, 132 (03) :424-439
[4]   Non-contact Low Power EEG/ECG Electrode for High Density Wearable Biopotential Sensor Networks [J].
Chi, Yu M. ;
Deiss, Stephen R. ;
Cauwenberghs, Gert .
SIXTH INTERNATIONAL WORKSHOP ON WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS, PROCEEDINGS, 2009, :246-250
[5]   A Highly Sensitive Force Sensor with Fast Response Based on Interlocked Arrays of Indium Tin Oxide Nanosprings toward Human Tactile Perception [J].
Chun, Sungwoo ;
Choi, Il Yong ;
Son, Wonkyeong ;
Bae, Gi Yoon ;
Lee, Eun Jae ;
Kwon, Hyunah ;
Jung, Jaimyun ;
Kim, Hyoung Seop ;
Kim, Jong Kyu ;
Park, Wanjun .
ADVANCED FUNCTIONAL MATERIALS, 2018, 28 (42)
[6]  
Einthoven W., 1902, ATIVIDADE ELETRICA C
[7]   A HISTORY OF THE ORIGIN, EVOLUTION, AND IMPACT OF ELECTROCARDIOGRAPHY [J].
FYE, WB .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (13) :937-949
[8]   A highly conductive and stretchable wearable liquid metal electronic skin for long-term conformable health monitoring [J].
Guo Rui ;
Wang XueLin ;
Yu WenZhuo ;
Tang JianBo ;
Liu Jing .
SCIENCE CHINA-TECHNOLOGICAL SCIENCES, 2018, 61 (07) :1031-1037
[9]   CNT/PDMS Composite Flexible Dry Electrodes for Long-Term ECG Monitoring [J].
Jung, Ha-Chul ;
Moon, Jin-Hee ;
Baek, Dong-Hyun ;
Lee, Jae-Hee ;
Choi, Yoon-Young ;
Hong, Joung-Sook ;
Lee, Sang-Hoon .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (05) :1472-1479
[10]   Bioinspired, Highly Stretchable, and Conductive Dry Adhesives Based on 1D-2D Hybrid Carbon Nanocomposites for All-in-One ECG Electrodes [J].
Kim, Taehoon ;
Park, Junyong ;
Sohn, Jongmoo ;
Cho, Donghwi ;
Jeon, Seokwoo .
ACS NANO, 2016, 10 (04) :4770-4778