Forehead monitoring of heart rate in neonatal intensive care

被引:1
作者
Stockwell, S. J. [1 ]
Kwok, T. C. [2 ]
Morgan, S. P. [1 ]
Sharkey, D. [2 ]
Hayes-Gill, B. R. [1 ]
机构
[1] Univ Nottingham, Fac Engn, Opt & Photon Res Grp, Nottingham, England
[2] Univ Nottingham, Ctr Perinatal Res Lifespan & Populat Hlth, Sch Med, Nottingham, England
基金
英国工程与自然科学研究理事会; 英国医学研究理事会;
关键词
photoplethysmogram; PPG; neonatal; heart rate; forehead; reflectance-mode; pulse oximeter Min; 5-Max; 8; NICU (neonatal intensive care unit);
D O I
10.3389/fphys.2023.1127419
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Heart rate is an extremely important physiological parameter to measure in critically unwell infants, as it is the main physiological marker that changes in response to a change in infant condition. Heart rate is routinely measured peripherally on a limb with a pulse oximeter. However, when infants are critically unwell, the blood supply to these peripheries is reduced in preference for central perfusion of vital organs such as the brain and heart. Measurement of heart rate with a reflection mode photoplethysmogram (PPG) sensor on the forehead could help minimise this problem and make it easier for other important medical equipment, such as cannulas, to be placed on the limbs. This study compares heart rates measured with a forehead-based PPG sensor against a wrist-based PPG sensor in 19 critically unwell infants in neonatal intensive care collecting 198 h of data. The two heart rates were compared using positive percentage agreement, Spearman's correlation coefficient and Bland-Altman analysis. The forehead PPG sensor showed good agreement with the wrist-based PPG sensor with limits of agreement of 8.44 bpm, bias of -0.22 bpm; positive percentage agreement of 98.87%; and Spearman's correlation coefficient of 0.9816. The analysis demonstrates that the forehead is a reliable alternative location for measuring vital signs using the PPG.
引用
收藏
页数:7
相关论文
共 22 条
  • [1] Microvascular imaging: techniques and opportunities for clinical physiological measurements
    Allen, John
    Howell, Kevin
    [J]. PHYSIOLOGICAL MEASUREMENT, 2014, 35 (07) : R91 - R141
  • [2] Berkenbosch JW, 2006, RESPIR CARE, V51, P726
  • [3] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [4] Bonner Oliver, 2017, BMJ Innov, V3, P12, DOI 10.1136/bmjinnov-2016-000145
  • [5] BS EN ISO 80601-2-61:2019, 2019, 806012612019 BS EN I
  • [6] Neonatal hemodynamics and management of hypotension in newborns
    Dilli, Dilek
    Soylu, Hanifi
    Tekin, Neslihan
    [J]. TURK PEDIATRI ARSIVI-TURKISH ARCHIVES OF PEDIATRICS, 2018, 53 : S65 - S75
  • [7] The use of photoplethysmography for assessing hypertension
    Elgendi, Mohamed
    Fletcher, Richard
    Liang, Yongbo
    Howard, Newton
    Lovell, Nigel H.
    Abbott, Derek
    Lim, Kenneth
    Ward, Rabab
    [J]. NPJ DIGITAL MEDICINE, 2019, 2 (1)
  • [8] Forehead reflectance photoplethysmography to monitor heart rate: preliminary results from neonatal patients
    Grubb, M. R.
    Carpenter, J.
    Crowe, J. A.
    Teoh, J.
    Marlow, N.
    Ward, C.
    Mann, C.
    Sharkey, D.
    Hayes-Gill, B. R.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2014, 35 (05) : 881 - 893
  • [9] Accurate neonatal heart rate monitoring using a new wireless, cap mounted device
    Henry, Caroline
    Shipley, Lara
    Ward, Carole
    Mirahmadi, Siavash
    Liu, Chong
    Morgan, Steve
    Crowe, John
    Carpenter, James
    Hayes-Gill, Barrie
    Sharkey, Don
    [J]. ACTA PAEDIATRICA, 2021, 110 (01) : 72 - 78
  • [10] Monitoring of heart and respiratory rates in newborn infants using a new photoplethysmographic technique
    Johansson, A
    Öberg, PÅ
    Sedin, G
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (7-8) : 461 - 467