Double pulsed field gradient diffusion MRI to assess skeletal muscle microstructure
被引:3
|作者:
Berry, D. B.
[1
,2
]
Galinsky, V. L.
[3
]
Hutchinson, E. B.
Galons, J. P.
[4
]
Ward, S. R.
[1
,5
,6
]
Frank, L. R.
[3
]
机构:
[1] Univ Calif San Diego, Dept Orthoped Surg, San Diego, CA USA
[2] Univ Calif San Diego, Dept Nanoengn, San Diego, CA USA
[3] Univ Calif San Diego, Ctr Sci Computat Imaging, San Diego, CA USA
[4] Univ Arizona, Dept Med Imaging, Tucson, AZ USA
[5] Univ Calif San Diego, Dept Radiol, San Diego, CA USA
[6] Univ Calif San Diego, Dept Bioengn, San Diego, CA USA
基金:
美国国家卫生研究院;
关键词:
diffusion anisotropy;
diffusion tensor imaging (DTI);
diffusion tensor subspace imaging (DiTSI);
double pulsed field gradient;
skeletal muscle;
PORE-SIZE DISTRIBUTION;
IN-VIVO;
MICROSCOPIC ANISOTROPY;
WEIGHTING EXPERIMENTS;
RELAXATION;
MODEL;
COEFFICIENT;
DEPENDENCE;
DIAMETER;
MATTER;
D O I:
10.1002/mrm.29751
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Purpose: Preliminary study to determine whether double pulsed field gradient (PFG) diffusion MRI is sensitive to key features ofmuscle microstructure related to function. Methods: The restricted diffusion profile of molecules in models of muscle microstructure derived from histology were systematically simulated using a numerical simulation approach. Diffusion tensor subspace imaging analysis of the diffusion signalwas performed, and spherical anisotropy (SA) was calculated for each model. Linear regression was used to determine the predictive capacity of SA on the fiber area, fiber diameter, and surface area to volume ratio of the models. Additionally, a rat model of muscle hypertrophy was scanned using a single PFG and a double PFG pulse sequence, and the restricted diffusion measurements were compared with histological measurements of microstructure. Results: Excellent agreement between SA and muscle fiber area (r(2) = 0.71; p < 0.0001), fiber diameter (r(2) = 0.83; p < 0.0001), and surface area to volume ratio (r(2) = 0.97; p < 0.0001) in simulated models was found. In a scanned rat leg, the distribution of these microstructural features measured from histology was broad and demonstrated that there is a wide variance in the microstructural features observed, similar to the SA distributions. However, the distribution of fractional anisotropy measurements in the same tissue was narrow. Conclusions: This study demonstrates that SA-a scalar value from diffusion tensor subspace imaging analysis-is highly sensitive to muscle microstructural features predictive of function. Furthermore, these techniques and analysis tools can be translated to real experiments in skeletal muscle. The increased dynamic range of SA compared with fractional anisotropy in the same tissue suggests increased sensitivity to detecting changes in tissue microstructure.