Metal artifacts and artifact reduction of neurovascular coils in photon-counting detector CT versus energy-integrating detector CT - in vitro comparison of a standard brain imaging protocol

被引:13
|
作者
Schmitt, Niclas [1 ]
Wucherpfennig, Lena [2 ]
Rotkopf, Lukas T. [3 ]
Sawall, Stefan [4 ]
Kauczor, Hans-Ulrich [2 ]
Bendszus, Martin [1 ]
Moehlenbruch, Markus A. [1 ]
Schlemmer, Heinz-Peter [3 ]
Vollherbst, Dominik F. [1 ]
机构
[1] Heidelberg Univ Hosp, Dept Neuroradiol, Neuenheimer Feld 400, D-69120 Heidelberg, Germany
[2] Heidelberg Univ Hosp, Dept Diagnost & Intervent Radiol, Neuenheimer Feld 420, D-69120 Heidelberg, Germany
[3] German Canc Res Ctr, Div Radiol, Neuenheimer Feld 280, D-69120 Heidelberg, Germany
[4] German Canc Res Ctr, Div Xray Imaging & Computed Tomog, Neuenheimer Feld 280, D-69120 Heidelberg, Germany
关键词
Metal artifact reduction; Coils; Photon-counting CT; Aneurysm; Brain imaging; COMPUTED-TOMOGRAPHY;
D O I
10.1007/s00330-022-09073-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives Photon-counting detector computed tomography (PCD-CT) is a promising new technique for CT imaging. The aim of the present study was the in vitro comparison of coil-related artifacts in PCD-CT and conventional energy-integrating detector CT (EID-CT) using a comparable standard brain imaging protocol before and after metal artifact reduction (MAR). Methods A nidus-shaped rubber latex, resembling an aneurysm of the cerebral arteries, was filled with neurovascular platinum coils and inserted into a brain imaging phantom. Image acquisition and reconstruction were repeatedly performed for PCD-CT and EID-CT (n = 10, respectively) using a standard brain imaging protocol. Moreover, linear interpolation MAR was performed for PCD-CT and EID-CT images. The degree of artifacts was analyzed quantitatively (standard deviation in a donut-shaped region of interest) and qualitatively (5-point scale analysis). Results Quantitative and qualitative analysis demonstrated a lower degree of metal artifacts in the EID-CT images compared to the total-energy PCD-CT images (e.g., 82.99 +/- 7.89 Hounsfield units (HU) versus 90.35 +/- 6.28 HU; p < 0.001) with no qualitative difference between the high-energy bin PCD-CT images and the EID-CT images (4.18 +/- 0.37 and 3.70 +/- 0.64; p = 0.575). After MAR, artifacts were more profoundly reduced in the PCD-CT images compared to the EID-CT images in both analyses (e.g., 2.35 +/- 0.43 and 3.18 +/- 0.34; p < 0.001). Conclusion PCD-CT in combination with MAR have the potential to provide an improved option for reduction of coil-related artifacts in cerebral imaging in this in vitro study.
引用
收藏
页码:803 / 811
页数:9
相关论文
共 50 条
  • [31] Energy-resolved CT imaging with a photon-counting silicon-strip detector
    Persson, Mats
    Huber, Ben
    Karlsson, Staffan
    Liu, Xuejin
    Chen, Han
    Xu, Cheng
    Yveborg, Moa
    Bornefalk, Hans
    Danielsson, Mats
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [32] Comparison of low-energy virtual monoenergetic images between photon-counting CT and energy-integrating detectors CT: A phantom study
    Greffier, Joel
    Faby, Sebastian
    Pastor, Maxime
    Frandon, Julien
    Erath, Julien
    Beregi, Jean Paul
    Dabli, Djamel
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2024, 105 (09) : 311 - 318
  • [33] Visualization of bone details in a novel photon-counting dual-source CT scanner-comparison with energy-integrating CT
    Bette, Stefanie J.
    Braun, Franziska M.
    Haerting, Mark
    Decker, Josua A.
    Luitjens, Jan H.
    Scheurig-Muenkler, Christian
    Kroencke, Thomas J.
    Schwarz, Florian
    EUROPEAN RADIOLOGY, 2022, 32 (05) : 2930 - 2936
  • [34] Abdominal image quality and dose reduction with energy-integrating or photon-counting detectors dual-source CT: A phantom study
    Greffiera, Joel
    Dabli, Djamel
    Faby, Sebastian
    Pastor, Maxime
    Croisille, Cedric
    de Oliveira, Fabien
    Erath, Julien
    Beregi, Jean Paul
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2024, 105 (10) : 379 - 385
  • [35] Multi-energy CT imaging for large patients using dual-source photon-counting detector CT
    Tao, Shengzhen
    Marsh, Jeffrey F.
    Tao, Ashley
    Michalak, Greg J.
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17)
  • [36] Photon-Counting Detector CT Virtual Monoenergetic Images in Cervical Trauma Imaging-Optimization of Dental Metal Artifacts and Image Quality
    Dillinger, Daniel
    Overhoff, Daniel
    Froelich, Matthias F.
    Kaatsch, Hanns L.
    Booz, Christian
    Hagen, Achim
    Vogl, Thomas J.
    Schoenberg, Stefan O.
    Waldeck, Stephan
    DIAGNOSTICS, 2024, 14 (06)
  • [37] Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study
    Siegel, Marilyn J.
    Bugenhagen, Scott M.
    Sanchez, Adrian
    Kim, Stacy
    Abadia, Andres
    Ramirez-Giraldo, Juan C.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 221 (03) : 363 - 371
  • [38] High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose
    Euler, Andre
    Higashigaito, Kai
    Mergen, Victor
    Sartoretti, Thomas
    Zanini, Bettina
    Schmidt, Bernhard
    Flohr, Thomas G.
    Ulzheimer, Stefan
    Eberhard, Matthias
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2022, 57 (02) : 115 - 121
  • [39] Noise performance of low-dose CT: comparison between an energy integrating detector and a photon counting detector using a whole-body research photon counting CT scanner
    Yu, Zhicong
    Leng, Shuai
    Kappler, Steffen
    Hahn, Katharina
    Li, Zhoubo
    Halaweish, Ahmed F.
    Henning, Andre
    McCollough, Cynthia H.
    JOURNAL OF MEDICAL IMAGING, 2016, 3 (04)
  • [40] Ultra-high Versus Standard Resolution Photon-Counting Detector CT Angiography for Imaging of Femoral Stents in a Cadaveric Perfusion Model
    Hartung, Viktor
    Hendel, Robin
    Huflage, Henner
    Augustin, Anne Marie
    Grunz, Jan-Peter
    Kleefeldt, Florian
    Peter, Dominik
    Lichthardt, Sven
    Erguen, Sueleyman
    Bley, Thorsten Alexander
    Gruschwitz, Philipp
    ACADEMIC RADIOLOGY, 2025, 32 (01) : 556 - 564