摘要
The emergence of mass spectrometry (MS)-based methods to quantify proteins for clinical applications has led to the need for accurate and consistent measurements. To meet the clinical needs of MS-based protein results, it is important that the results are traceable to higher-order standards and methods and have defined uncertainty values. Therefore, we outline a comprehensive approach for the estimation of measurement uncertainty of a MS-based procedure for the quantification of a protein biomarker. Using a bottom-up approach, which is the model outlined in the "Guide to the Expression of Uncertainty of Measurement" (GUM), we evaluated the uncertainty components of a MS-based measurement procedure for a protein biomarker in a complex matrix. The cause-and-effect diagram of the procedure is used to identify each uncertainty component, and statistical equations are derived to determine the overall combined uncertainty. Evaluation of the uncertainty components not only enables the calculation of the measurement uncertainty but can also be used to determine if the procedure needs improvement. To demonstrate the use of the bottom-up approach, the overall combined uncertainty is estimated for the National Institute of Standards and Technology (NIST) candidate reference measurement procedure for albumin in human urine. The results of the uncertainty approach are applied to the determination of uncertainty for the certified value for albumin in candidate NIST Standard Reference Material (R) (SRM) 3666. This study provides a framework for measurement uncertainty estimation of a MS-based protein procedure by identifying the uncertainty components of the procedure to derive the overall combined uncertainty.
引用
收藏
页码:3265 / 3274
页数:10
相关论文
共 50 条
  • [41] Development of human serum matrix-based unconjugated estriol-certified reference material by recommended ID-LC-MS/MS reference method
    Zhang, Qiaoxuan
    Bai, Kai
    Wu, Meiran
    Lin, Haibiao
    Yan, Jun
    Zhan, Min
    Ouyang, Fen
    Ke, Peifeng
    Han, Liqiao
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2022, 414 (07) : 2523 - 2531
  • [42] Optimization of sample preparation for transporter protein quantification in tissues by LC-MS/MS
    Jankovskaja, Skaidre
    Kamiie, Junichi
    Rezeli, Melinda
    Gustaysson, Lena
    Sugihara, Yutaka
    Miliotis, Tasso
    Ruzgas, Tautgirdas
    Marko-Varga, Gyorgy
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2019, 164 : 9 - 15
  • [43] An accurate measurement method for serum homocysteine measurement by ID-LC/MS/MS and the application of external quality assessment
    Wang, Mo
    Shi, Jie
    Song, Yichuan
    Zhang, Shunli
    Zhang, Rui
    CLINICA CHIMICA ACTA, 2025, 565
  • [44] ID-LC-MS/MS方法对被动吸烟大鼠肺组织DNA中8-oxo-dGsn的检测
    陈连国
    王蒙蒙
    干伟
    陈晓乐
    罗顺斌
    蔡剑平
    胡国新
    医学研究杂志, 2013, (06) : 30 - 33
  • [45] GLYCOSAMINOGLYCAN QUANTIFICATION BY LC/MS-MS FOR MUCOPOLYSACCHARIDOSES
    Auray-Blais, C.
    Lavoie, P.
    Gagnon, R.
    Clarke, J. T. R.
    Maranda, B.
    An, Y.
    Young, S. P.
    Zhang, H.
    Millington, D. S.
    JOURNAL OF INHERITED METABOLIC DISEASE, 2011, 34 : S217 - S217
  • [46] Development of human serum matrix-based unconjugated estriol-certified reference material by recommended ID-LC-MS/MS reference method
    Qiaoxuan Zhang
    Kai Bai
    Meiran Wu
    Haibiao Lin
    Jun Yan
    Min Zhan
    Fen Ouyang
    Peifeng Ke
    Liqiao Han
    Junhua Zhuang
    Xianzhang Huang
    Analytical and Bioanalytical Chemistry, 2022, 414 : 2523 - 2531
  • [47] Quantification of Quetiapine in Human Plasma by LC–MS–MS
    Mohan Laxman Kundlik
    Sandeep Kambli
    Viral Shah
    Yashwant Patel
    Santosh Gupta
    Ram Sharma
    Bhaskar Zaware
    Shashikant R. Kuchekar
    Chromatographia, 2009, 70 : 1587 - 1592
  • [48] Candidate Reference Measurement Procedure for the Quantification of Total Serum Cortisol with LC-MS/MS
    Hawley, James M.
    Owen, Laura J.
    MacKenzie, Finlay
    Mussell, Chris
    Cowen, Simon
    Keevil, Brian G.
    CLINICAL CHEMISTRY, 2016, 62 (01) : 262 - 269
  • [49] Optimizing charge state distribution is a prerequisite for accurate protein biomarker quantification with LC-MS/MS, as illustrated by hepcidin measurement
    Schmitz, Ellen M. H.
    Leijten, Niels M.
    van Dongen, Joost L. J.
    Broeren, Maarten A. C.
    Milroy, Lech G.
    Brunsveld, Luc
    Scharnhorst, Volkher
    van de Kerkhof, Daan
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (09) : 1490 - 1497
  • [50] Dynamics of serum testosterone during the menstrual cycle evaluated by daily measurements with an ID-LC-MS/MS method and a 2nd generation automated immunoassay
    Bui, Hong N.
    Sluss, Patrick M.
    Blincko, Stuart
    Knol, Dirk L.
    Blankenstein, Marinus A.
    Heijboer, Annemieke C.
    STEROIDS, 2013, 78 (01) : 96 - 101