Estimation of measurement uncertainty for the quantification of protein by ID-LC-MS/MS

被引:5
|
作者
Beasley-Green, Ashley [1 ]
Heckert, N. Alan [2 ]
机构
[1] Natl Inst Stand & Technol, Biomol Measurement Div, Mat Measurement Lab, 100 Bur Dr, Gaithersburg, MD 20899 USA
[2] Natl Inst Stand & Technol, Stat Engn Div, Informat Technol Lab, 100 Bur Dr, Gaithersburg, MD 20899 USA
关键词
Measurement uncertainty; Isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS; MS); Albumin; Design of experiments (DOE); Reference measurement procedure (RMP); Standard Reference Material (R) (SRM);
D O I
10.1007/s00216-023-04705-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The emergence of mass spectrometry (MS)-based methods to quantify proteins for clinical applications has led to the need for accurate and consistent measurements. To meet the clinical needs of MS-based protein results, it is important that the results are traceable to higher-order standards and methods and have defined uncertainty values. Therefore, we outline a comprehensive approach for the estimation of measurement uncertainty of a MS-based procedure for the quantification of a protein biomarker. Using a bottom-up approach, which is the model outlined in the "Guide to the Expression of Uncertainty of Measurement" (GUM), we evaluated the uncertainty components of a MS-based measurement procedure for a protein biomarker in a complex matrix. The cause-and-effect diagram of the procedure is used to identify each uncertainty component, and statistical equations are derived to determine the overall combined uncertainty. Evaluation of the uncertainty components not only enables the calculation of the measurement uncertainty but can also be used to determine if the procedure needs improvement. To demonstrate the use of the bottom-up approach, the overall combined uncertainty is estimated for the National Institute of Standards and Technology (NIST) candidate reference measurement procedure for albumin in human urine. The results of the uncertainty approach are applied to the determination of uncertainty for the certified value for albumin in candidate NIST Standard Reference Material (R) (SRM) 3666. This study provides a framework for measurement uncertainty estimation of a MS-based protein procedure by identifying the uncertainty components of the procedure to derive the overall combined uncertainty.
引用
收藏
页码:3265 / 3274
页数:10
相关论文
共 50 条
  • [31] Measurement of dehydroepiandrosterone sulphate (DHEAS): A comparison of Isotope-Dilution Liquid Chromatography Tandem Mass Spectrometry (ID-LC-MS/MS) and seven currently available immunoassays
    Buttler, Rahel M.
    Kruit, Adrian
    Blankenstein, Marinus A.
    Heijboer, Annemieke C.
    CLINICA CHIMICA ACTA, 2013, 424 : 22 - 26
  • [32] Development of matrix-based reference materials for 17 beta-estradiol by the recommended reference method of ID-LC-MS/MS
    Haibiao Lin
    Dan Liu
    Li Deng
    Jun Yan
    Liqiao Han
    Min Zhan
    Pengwei Zhang
    Jianbing Wang
    Min He
    Qiaoxuan Zhang
    Xianzhang Huang
    Analytical and Bioanalytical Chemistry, 2023, 415 : 5637 - 5644
  • [33] LC-MS quantification of protein drugs: validating protein LC-MS methods with predigestion immunocapture
    Duggan, Jeffrey
    Ren, Bailuo
    Mao, Yan
    Chen, Lin-Zhi
    Philip, Elsy
    BIOANALYSIS, 2016, 8 (18) : 1951 - 1964
  • [34] Targeted protein quantification by LC-MS/MS: Applications in the agricultural biotechnology
    Hu, Tiger
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2016, 252
  • [35] An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of carbamazepine-10,11-epoxide in human serum and plasma
    Schierscher, Tobias
    Singh, Neeraj
    Kobel, Anja
    Wild, Janik
    Bauland, Friederike
    Geistanger, Andrea
    Risch, Lorenz
    Geletneky, Christian
    Seger, Christoph
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (07) : 1339 - 1351
  • [36] Quantification of Urinary Albumin by Using Protein Cleavage and LC-MS/MS
    Seegmiller, Jesse C.
    Barnidge, David R.
    Burns, Bradley E.
    Larson, Timothy S.
    Lieske, John C.
    Kumar, Rajiv
    CLINICAL CHEMISTRY, 2009, 55 (06) : 1100 - 1107
  • [37] Antibody-free workflows for protein quantification by LC-MS/MS
    Wilffert, Daniel
    Bischoff, Rainer
    van de Merbel, Nico C.
    BIOANALYSIS, 2015, 7 (06) : 763 - 779
  • [38] ID-LC-MS/MS候选参考方法测定血清载脂蛋白A-Ⅰ和B的优化
    李卿
    居漪
    孙贺伟
    金中淦
    临床检验杂志, 2018, 36 (12) : 899 - 902
  • [39] Rapid LC-MS/MS Bosutinib Quantification with Applications in Metabolic Stability Estimation
    Attwa, Mohamed W.
    Alanazi, Mohammed M.
    MOLECULES, 2023, 28 (04):
  • [40] LC/MS AND LC/MS/MS DETERMINATION OF PROTEIN TRYPTIC DIGESTS
    HUANG, EC