Surgical Navigation System for Pedicle Screw Placement Based on Mixed Reality

被引:1
|
作者
Hwang, Seokbin [1 ]
Lee, Suk-joong [2 ,3 ]
Kim, Sungmin [4 ]
机构
[1] Univ Ulsan, Dept Elect Elect & Comp Engn, 93 Daehak Ro, Ulsan 44610, South Korea
[2] Gyeongsang Natl Univ, Coll Med, Dept Orthopaed Surg, 11 Samjeongja Ro, Chang Won, Gyeongsangnam D, South Korea
[3] Gyeongsang Natl Univ, Changwon Hosp, 11 Samjeongja Ro, Chang Won, Gyeongsangnam D, South Korea
[4] Univ Ulsan, Dept Biomed Engn, 93 Daehak Ro, Ulsan 44610, South Korea
基金
新加坡国家研究基金会;
关键词
HoloLens; mixed reality; pedicle screw placement; surgical navigation; ACCURACY; METAANALYSIS;
D O I
10.1007/s12555-023-0083-6
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Inaccurate screw insertion has been a major concern in pedicle screw placement in spinal surgery because of the potential for postoperative complications. Efforts to improve screw insertion have led to the development of novel surgical navigation systems. However, current surgical navigation systems have several problems in that the attention of surgeons is frequently interrupted during surgery, and complications occur in their depth perception because of the limitations of 2D monitors. This study proposes a surgical navigation system to address these issues using mixed reality (MR)-based smart glasses. We developed a navigation system that provides 3D visualization of the surgical tool with respect to the patient's anatomy. It utilizes preoperative surgical planning data to provide visual feedback and guidance. An orthopedic surgeon performed pedicle screw placement with the navigation system in two different environments, using Microsoft HoloLens (HoloLens) and a 2D monitor. The results were evaluated according to procedural time, translational error, angular error, and clinical accuracy. The mean procedural time was 111.3 +/- 52.7 s with the HoloLens and 192.1 +/- 104.0 s with the 2D monitor. The mean translational error was 2.14 +/- 1.13 mm at the entry and 3.14 +/- 0.90 mm at the target with HoloLens. With the 2D monitor, the mean translational error was 2.10 +/- 0.97 mm and 3.41 +/- 2.16 mm at the entry and the target, respectively. The mean angular error was 6.44 +/- 1.94 deg with HoloLens and 7.14 +/- 4.20 deg with the 2D monitor. All screws were inserted intrapedicularly in both environments. The navigation system enables free visualization, reflects the human eye's perspective, and retains the advantages of MR-based smart glasses. A navigation system compatible with minimally invasive surgery should be developed in the future.
引用
收藏
页码:3983 / 3993
页数:11
相关论文
共 50 条
  • [31] A Comparison of CT-based Navigation Techniques for Minimally Invasive Lumbar Pedicle Screw Placement
    Wood, Martin
    Mannion, Richard
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2011, 24 (01): : E1 - E5
  • [32] The efficacy of immersive virtual reality surgical simulator training for pedicle screw placement: a randomized double-blind controlled trial
    Xin, Baoquan
    Huang, Xing
    Wan, Wei
    Lv, Kai
    Hu, Yiwei
    Wang, Jing
    Li, Song
    Zou, Weiwei
    Xiao, Jianru
    Liu, Tielong
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (05) : 927 - 934
  • [33] Augmented Reality Navigation System Enhances the Accuracy of Spinal Surgery Pedicle Screw Placement: A Randomized, Multicenter, Parallel-Controlled Clinical Trial
    Ma, Yichao
    Wu, Jiangpeng
    Dong, Yanlong
    Tang, Hongmei
    Ma, Xiaojun
    ORTHOPAEDIC SURGERY, 2025, 17 (02) : 631 - 643
  • [34] Augmented reality-based surgical navigation of pelvic screw placement: an ex-vivo experimental feasibility study
    Sandro-Michael Heining
    Vladislav Raykov
    Oliver Wolff
    Hatem Alkadhi
    Hans-Christoph Pape
    Guido A. Wanner
    Patient Safety in Surgery, 18
  • [35] Augmented reality-based surgical navigation of pelvic screw placement: an ex-vivo experimental feasibility study
    Heining, Sandro-Michael
    Raykov, Vladislav
    Wolff, Oliver
    Alkadhi, Hatem
    Pape, Hans-Christoph
    Wanner, Guido A.
    PATIENT SAFETY IN SURGERY, 2024, 18 (01)
  • [36] Three-dimensional Fluoroscopy-based Navigation for the Pedicle Screw Placement in Patients with Primary Invasive Spinal Tumors
    Jin, Bo
    Su, Yi-Bing
    Zhao, Ji-Zong
    CHINESE MEDICAL JOURNAL, 2016, 129 (21) : 2552 - 2558
  • [37] Radiation exposure for pedicle screw placement with three different navigation system and imaging combinations in a sawbone model
    Beisemann, Nils
    Gierse, Jula
    Mandelka, Eric
    Hassel, Frank
    Gruetzner, Paul A.
    Franke, Jochen
    Vetter, Sven Y.
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [38] Pedicle screw navigation using surface digitization on the Microsoft HoloLens
    Liebmann, Florentin
    Roner, Simon
    von Atzigen, Marco
    Scaramuzza, Davide
    Sutter, Reto
    Snedeker, Jess
    Farshad, Mazda
    Furnstahl, Philipp
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2019, 14 (07) : 1157 - 1165
  • [39] Pedicle guide for thoracic pedicle screw placement
    Abode-Iyamah, Kingsley
    Stemper, Luke
    Rachman, Shane
    Schneider, Kelly
    Sick, Kathryn
    Hitchon, Patrick
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2012, 10 (03) : 211 - 220
  • [40] Pediatric Lumbar Pedicle Screw Placement Using Navigation Templates A Cadaveric Study
    Wang, Xing
    Shi, Jun
    Zhang, Shojie
    Zhang, Zhifeng
    Li, Xiaohe
    Li, Zhijun