Zygomatic implant placement using a robot-assisted flapless protocol: proof of concept

被引:14
作者
Olivetto, M. [1 ,2 ,4 ]
Bettoni, J. [2 ]
Testelin, S. [2 ]
Lefranc, M. [3 ]
机构
[1] Hosp Ctr Annecy Geneva, Dept Maxillofacial Surg, Epagny Metz Tessy, France
[2] Univ Hosp Amiens Picardy, Univ Med Ctr, Dept Maxillofacial Surg, Amiens, France
[3] Univ Hosp Amiens Picardy, Univ Med Ctr, Dept Neurosurg, Amiens, France
[4] Hosp Ctr Annecy Geneva, Dept Maxillofacial Surg, F-74370 Epagny Metz Tessy, France
关键词
Robot-assisted surgery; Computer-assisted surgery; Zygoma; Dental implantation; Proof of concept study; Intraoperative care; Cone-beam computed tomography; COMPLICATIONS; SURVIVAL; ACCURACY;
D O I
10.1016/j.ijom.2022.12.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Robotic assistance can help in physically guiding the drilling trajectory during zygomatic implant positioning. A new robot-assisted strategy for a flapless zygomatic implant placement protocol is reported here. In this protocol, a preoperative computed tomography scan is used to plan the surgical path. After surface registration, the ROSA robot (Zimmer Biomet Robotics) guides several steps, which are performed with shared control. The surgeon performs the drilling and tapping, guided by the robotic arm, which is positioned according to the planned trajectory. Placement of the zygomatic implant is done manually. Immediate intraoperative 3D verification is performed by cone beam computed tomography (flat-panel detector, Medtronic O-arm II). Four zygomatic implants were placed in the case patient according to the flapless protocol, with a mean vector error of 1.78 mm (range 0.52-4.70 mm). A screw-retained temporary prosthesis was placed on the same day. No significant complications were observed. The application of this robotassisted surgical protocol, which guarantees a very high degree of precision, may reduce inaccuracies in the positioning of zygomatic implants that could deviate from the surgeon's plan. This appears to be a potentially safe flapless surgery technique. Drill slipping on the crest or on the maxillary wall is the main source of error in this procedure, emphasizing the usefulness of the assisted surgical
引用
收藏
页码:710 / 715
页数:6
相关论文
共 14 条
[1]  
Aparicio C., 2012, Zygomatic Implants: The Anatomy Guided Approach
[2]  
Bedrossian E, 2010, INT J ORAL MAX IMPL, V25, P1213
[3]   Pilot study of a surgical robot system for zygomatic implant placement [J].
Cao, Zhenggang ;
Qin, Chunxia ;
Fan, Shengchi ;
Yu, Dedong ;
Wu, Yiqun ;
Qin, Jing ;
Chen, Xiaojun .
MEDICAL ENGINEERING & PHYSICS, 2020, 75 (75) :72-78
[4]   Survival and Complications of Zygomatic Implants: An Updated Systematic Review [J].
Chrcanovic, Bruno Ramos ;
Albrektsson, Tomas ;
Wennerberg, Ann .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2016, 74 (10) :1949-1964
[5]   Image-Guided Interventions: Technology Review and Clinical Applications [J].
Cleary, Kevin ;
Peters, Terry M. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, VOL 12, 2010, 12 :119-142
[6]   The Impact of the Reference Imaging Modality, Registration Method and Intraoperative Flat-Panel Computed Tomography on the Accuracy of the ROSA® Stereotactic Robot [J].
Lefranc, Michel ;
Capel, Cyril ;
Pruvot, Anne Sophie ;
Fichten, Anthony ;
Desenclos, Christine ;
Toussaint, Patrick ;
Le Gars, Daniel ;
Peltier, Johan .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2014, 92 (04) :242-250
[7]   Frameless ROSA® Robot-Assisted Lead Implantation for Deep Brain Stimulation: Technique and Accuracy [J].
Liu, Lannie ;
Mariani, Sarah Giulia ;
De Schlichting, Emmanuel ;
Grand, Sylvie ;
Lefranc, Michel ;
Seigneuret, Eric ;
Chabardes, Stephan .
OPERATIVE NEUROSURGERY, 2020, 19 (01) :57-64
[8]  
Medtronic, O ARM IMAGING SYSTEM
[9]   Surgical complications in zygomatic implants: A systematic review [J].
Molinero-Mourelle, Pedro ;
Baca-Gonzalez, Laura ;
Gao, Baoluo ;
Saez-Alcaide, Luis-Miguel ;
Helm, Alexandra ;
Lopez-Quiles, Juan .
MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2016, 21 (06) :E751-E757
[10]   Dynamic Navigation for Zygomatic Implants: A Case Report about a Protocol with Intraoral Anchored Reference Tool and an Up-To-Date Review of the Available Protocols [J].
Pellegrino, Gerardo ;
Lizio, Giuseppe ;
Basile, Francesco ;
Stefanelli, Luigi Vito ;
Marchetti, Claudio ;
Felice, Pietro .
METHODS AND PROTOCOLS, 2020, 3 (04) :1-10