Transpalpebral intraocular pressure measurement by Diaton compared to Goldman applanation tonometer in myopic eyes before and after transepithelial photorefractive keratectomy in Saudi Arabia

被引:0
作者
Alzuhairy, Sultan [1 ]
机构
[1] Qassim Univ, Dept Ophthalmol, Coll Med, Buraydah 51452, Saudi Arabia
关键词
intraocular pressure; Goldman applanation tonometry; trans palpebral tonometry; myopia; transepithelial photorefractive keratectomy; CENTRAL CORNEAL THICKNESS; STANDARDS;
D O I
10.18240/ijo.2023.03.07
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
AIM: To compare intraocular pressure (IOP) measured via the trans palpebral IOP (tpIOP) method using a Diaton or with a Goldman applanation tonometer (GAT) and study the determinants of IOP difference in eyes undergoing transepithelial photorefractive keratectomy (TPRK) for myopia. METHODS: This cross-sectional validation study was held in 2020 in an eye hospital in Saudi Arabia. IOP was measured by Diaton and GAT before treatment, after one week, and one month of TPRK. It is considered if IOP difference by Diaton and GAT was less than +/- 2 mm Hg acceptable. The IOP difference was tested if correlated to spherical equivalent (SE), central corneal thickness (CCT), age, gender, or tpIOP. RESULTS: Totolly 200 myopic eyes of 100 patients were included in the study. The mean difference of IOP measured by two methods before TPRK, 1wk, and 1mo after TPRK were 0.790, -0.790, and -0.920 mm Hg, respectively (P<0.001). Diaton could measure IOP effectively 89.5% eyes before TPRK and 82% and 84% at 1wk and 1mo after TPTK, respectively. At week 1, IOP differences were significantly correlated to baseline CCT (P=0.02) and tpIOP at week 1 (P<0.001). One month after TPRK, only tpIOP was significantly correlated to the difference in IOP (P<0.001). CONCLUSION: Diaton is a good screening tool for IOP before TPRK. It helps in monitoring IOP after surgery. Although more practical, it is less efficient than GAT. In eyes with high myopia and low tpIOP before surgery, IOP post-TPRK by Diaton and GAT could differ.
引用
收藏
页码:375 / 381
页数:7
相关论文
共 26 条
  • [1] Refractive Surgery: The Never-Ending Task of Improving Vision Correction
    Al-Swailem, Samar A.
    [J]. MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2014, 21 (01) : 1 - 2
  • [2] Refractive errors among Saudi college students and associated risk factors
    Alsaif, Bayan A.
    Aljindan, Mohanna Y.
    Alrammah, Hanan M.
    Almulla, Malak O.
    Alshahrani, Saad S.
    [J]. CLINICAL OPHTHALMOLOGY, 2019, 13 : 437 - 443
  • [3] American Academy of Ophthalmology, 2017, Refractive Errors & Refractive Surgery Preferred Practice Pattern
  • [4] Chakraborty A K., Taiwan Journal of Ophthalmology, DOI DOI 10.1016/J.TJO.2014.03.002
  • [5] Intraocular pressure measurement: A review *
    Da Silva, Filipe
    Lira, Madalena
    [J]. SURVEY OF OPHTHALMOLOGY, 2022, 67 (05) : 1319 - 1331
  • [6] Das MR, 2022, INDIAN J OPHTHALMOL, V70, P1246, DOI DOI 10.4103/ijo.IJO_93_22
  • [7] Dean AG, OPENEPI OPEN SOURCE
  • [8] Enikov ET, 2015, American Society of Mechanical Engineers, V57380
  • [9] IMI - Defining and Classifying Myopia: A Proposed Set of Standards for Clinical and Epidemiologic Studies
    Flitcroft, Daniel Ian
    He, Mingguang
    Jonas, Jost B.
    Jong, Monica
    Naidoo, Kovin
    Ohno-Matsui, Kyoko
    Rahi, Jugnoo
    Resnikoff, Serge
    Vitale, Susan
    Yannuzzi, Lawrence
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (03) : M20 - M30
  • [10] Gadelkarim Ahmed H, 2017, Adv Ophthalmol Vis Syst, V6, P00165, DOI DOI 10.15406/AOVS.2017.06.00165