Systemic observation and measurement of microgap formation in dental implant-abutment connection interface under fatigue

被引:0
作者
Mercan, Cigdem [2 ]
Bulutsuz, Asli Gunay [1 ]
Katiboglu, Ahmet Bulent [3 ]
机构
[1] Yildiz Tech Univ, TR-34349 Istanbul, Turkiye
[2] Istanbul Univ, Dent Fac, Istanbul, Turkiye
[3] Atlas Univ, Istanbul, Turkiye
关键词
microgap; fatigue test; experimental design; dental implant-abutment interface; measurement; IN-VITRO EVALUATION; BACTERIAL-COLONIZATION; STRESS-DISTRIBUTION; MICROBIAL LEAKAGE; 2-STAGE IMPLANTS; MARGINAL FIT; DIAMETER; LOAD; INFLAMMATION; MICROMOTION;
D O I
10.1515/mt-2023-0071
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Today, a variety of implant designs with different attachment types are widely used in dental implant systems to eliminate connection problems. However, there are still design issues that can significantly affect the long-term success of dental implants. One of these problems is the formation of a microgap in the abutment-implant interface, which can lead to abutment screw loosening, screw fracture, and abutment fracture under masticatory loading. In this in vitro study, the effects of gingival height, dental implant diameter, and abutment type (solid, digital, and ball) on microgap formation were observed after repetitive fatigue tests using an experimental design. After the tests, the optimal design factors were determined to minimize microgap formation. According to the results of the experimental design, the abutment type was found to be the most influential parameter on microgap formation. An implant diameter of 4.1 mm, the smallest gingival height of 0.5 mm, and a digital attachment combination were identified as the optimal design parameters based on the statistical approach (R-2 = 85.3 %).
引用
收藏
页码:1679 / 1685
页数:7
相关论文
共 48 条
[1]   The peri-implant hard and soft tissues at different implant systems - A comparative study in the dog [J].
Abrahamsson, I ;
Berglundh, T ;
Wennstrom, J ;
Lindhe, J .
CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (03) :212-219
[2]  
Abrahamsson I, 1998, CLIN ORAL IMPLAN RES, V9, P73
[3]   The mucosal attachment at different abutments - An experimental study in dogs [J].
Abrahamsson, I ;
Berglundh, T ;
Glantz, PO ;
Lindhe, J .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1998, 25 (09) :721-727
[4]   Improving the fatigue life of produced dental implants by the thread-rolling process [J].
Aktas, Serkan ;
Kisioglu, Yasin .
MATERIALS TESTING, 2022, 64 (07) :1012-1025
[5]   Microbial leakage through the implant-abutment interface of morse taper implants in vitro [J].
Aloise, Joao Pedro ;
Curcio, Ricardo ;
Laporta, Marcia Zorello ;
Rossi, Liliane ;
Alvares da Silva, Adriana Madeira ;
Rapoport, Abrao .
CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) :328-335
[6]  
Anitua E, 2010, INT J PERIODONT REST, V30, P89
[7]   Bacterial Leakage in Implants With Different Implant-Abutment Connections: An In Vitro Study [J].
Assenza, Bartolomeo ;
Tripodi, Domenico ;
Scarano, Antonio ;
Perrotti, Vittoria ;
Piattelli, Adriano ;
Iezzi, Giovanna ;
D'Ercole, Simonetta .
JOURNAL OF PERIODONTOLOGY, 2012, 83 (04) :491-497
[8]  
Assuncao Wirley Goncalves, 2004, Implant Dent, V13, P358
[9]   Fatigue induced changes in conical implant-abutment connections [J].
Blum, Kai ;
Wiest, Wolfram ;
Fella, Christian ;
Balles, Andreas ;
Dittmann, Jonas ;
Rack, Alexander ;
Maier, Dominik ;
Thomann, Ralf ;
Spies, Benedikt Christopher ;
Kohal, Ralf Joachim ;
Zabler, Simon ;
Nelson, Katja .
DENTAL MATERIALS, 2015, 31 (11) :1415-1426
[10]   Influence of platform diameter in the reliability and failure mode of extra short dental implants [J].
Bordin, Dimorvan ;
Bergamo, Edmara T. P. ;
Bonfante, Estevam A. ;
Fardin, Vinicius P. ;
Coelho, Paulo G. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 77 :470-474