FINITE ELEMENT ANALYSIS OF CUSTOMIZED KNEE IMPLANTS BY VARYING LOADS DURING FLEXION-EXTENSION MOVEMENT

被引:0
|
作者
Alshewaier, Shady A. A. [1 ]
Alshehri, Majed Ali Ghurmallah [2 ]
Begum, S. Sabarunisha [3 ]
Sikkandar, Mohamed Yacin [2 ]
Almukil, Ali Abdullah A. [2 ]
Alhazmi, Aishah Mohammad [4 ]
机构
[1] Majmaah Univ, Coll Appl Med Sci, Dept Phys Therapy, Al Majmaah 11952, Saudi Arabia
[2] Majmaah Univ, Coll Appl Med Sci, Dept Med Equipment Technol, Al Majmaah 11952, Saudi Arabia
[3] PSR Engn Coll, Dept Biotechnol, Sivakasi 626140, Tamil Nadu, India
[4] Ad Diriyah Hosp, Minist Hlth, Diriyah, Saudi Arabia
关键词
Total knee replacement; knee osteoarthritis; off-the-shelf knee; patient-specific; biomechanical; ARTHROPLASTY; BIOMECHANICS;
D O I
10.1142/S021951942340016X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A customized knee implant numerical modeling using finite element analysis (FEA) during flexion extension has been investigated in this paper with varying loads with an objective of studying its kinematics. Computed tomography (CT) images of 15 osteoarthritis subjects' images were used in this work. Various morphological characteristics were extracted from clinical images using a commercial CAE software and biomechanical properties were studied on applying standard loads on customized implant and off-the-shelf (OTS) implants. Patient-specific knee implants have been designed according to the morphological characteristics and bone dimension of patient with compressive loads (1500, 1700 and 3000 N) during normal gait and were compared with the normal knee. Results showed that the stresses are distributed equally to the spacer and the tibial plate, unlike the standard femoral component where the stresses get concentrated on the cut edges. In compressive load, active stress and strain (similar to 5-20 MPa) are lesser (similar to 5%) and in flexion extension also lesser with a scaling factor of 0.785 and 1.0. The designed implant was found to produce similar biomechanical properties when compared to normal knee joint and thus it can be considered as a valuable implant and could replace the standard OTS knee implants.
引用
收藏
页数:15
相关论文
共 26 条
  • [21] A 3D finite element model of a knee for joint contact stress analysis during sport activities
    Bratianu, C
    Rinderu, P
    Gruionu, L
    ADVANCES IN FRACTURE AND FAILURE PREVENTION, PTS 1 AND 2, 2004, 261-263 : 513 - 517
  • [22] Loading on Attune® fixed-bearing cruciate-substituting total knee implant in knee malalignment during activities of daily living: A finite element analysis
    Shetty, Gautam
    Khairkar, Shruti
    JOURNAL OF ORTHOPAEDICS, 2021, 26 : 36 - 41
  • [23] Effect of vertical bone loss on stress distribution at the bone-implant interface around implants of varying diameters-an in silico 3D finite element analysis
    Gupta, Shipra
    Chopra, Priyanka
    Goyal, Parveen
    Jain, Ashish
    MATERIALS TODAY-PROCEEDINGS, 2021, 45 : 4581 - 4586
  • [24] Sensitivity analysis of paediatric knee kinematics to the graft surgical parameters during anterior cruciate ligament reconstruction: A sequentially linked neuromusculoskeletal-finite element analysis
    Dastgerdi, Ayda Karimi
    Esrafilian, Amir
    Carty, Christopher P.
    Nasseri, Azadeh
    Barzan, Martina
    Korhonen, Rami K.
    Astori, Ivan
    Hall, Wayne
    Saxby, David John
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2024, 248
  • [25] Individuals With an Anterior Cruciate Ligament-Reconstructed Knee Display Atypical Whole Body Movement Strategies but Normal Knee Robustness During Side-Hop Landings: A Finite Helical Axis Analysis
    Markstrom, Jonas L.
    Grip, Helena
    Schelin, Lina
    Hager, Charlotte K.
    AMERICAN JOURNAL OF SPORTS MEDICINE, 2020, 48 (05) : 1117 - 1126
  • [26] Effects of micromovement on the changes in stress distribution of partially stabilized zirconia (PS-ZrO2) dental implants and bridge during clenching: A three-dimensional finite element analysis
    Choi, Andy H.
    Matinlinna, Jukka
    Ben-Nissan, Besim
    ACTA ODONTOLOGICA SCANDINAVICA, 2013, 71 (01) : 72 - 81